MEASUREMENT FOUNDATION
9
ROBOT 60
Storage of constant / historical storage
Storage of constant
You can store and recall a frequently used val-
ue (e.g. height of a room). Measure the desired
distance, press and hold the button (6) until
Recalling the constant
Press the button (6) once to recall the constant
and then press the button (1) to enter it into
your calculation.
History storage
Press the button (6) twice and the previous 20
results (measurements or calculated results)
are shown in reverse order. Use button (3) to
navigate through the records. Press the button
(1) to use the result from the summary line for
further calculations.
INFO CAUSE REMEDY
204 Repeat procedure
252 Temperature too high Cool down instrument
253 Temperature too low Warm up instrument
255 Receiver signal too weak Use target plate
256 Received signal too
strong
Use target plate (grey
side)
257 Wrong measurement Use target plate (brown
side)
258 Wrong initialization Switch on – off the
instrument
ERROR CAUSE REMEDY
Error Hardware error Switch on/off the device
several times and check if
the symbol still appears. If
so please call your dealer
for assistance.
MESSAGE CODES