AWS LB-SERIES Manuale utente

LB-501 (500g x 0.01g)
LB-1000 (1000g x 0.1g)
LB-3000 (3000g x 0.1g)
User Manual
LB-Series
Copyright © 2010 American Weigh Scales, Inc.
All rights reserved. Rev. 5
English
Français
Español

LB-Series Manual
Thank you for purchasing the American Weigh
LB-Series compact bench scale. Please read all operating
instructions carefully before use. Scales are precision
measuring devices and should always be handled with
proper care. To ensure years of reliable service, keep these
simple tips in mind:
• Do not exceed the scale’s maximum capacity.
Overloading your scale can permanently damage it!
• Avoid exposure to extreme heat or cold. Scales
perform best at normal room temperature. If
temperatures have changed dramatically, calibration
and adjustment may be necessary.
• Allow your scale to warm up for 30-60 seconds before
performing calibration and adjustment(if available).
• Store your scale in a clean, dry location. Dust, dirt,
and moisture can accumulate on the weighing
sensors and electronics causing inaccuracy or
malfunction.
• Avoid using your scale in close proximity to strong
radio wave emitting devices such as computers, cash
registers, and cordless phones.
• Always weigh on a at and level surface, free from
vibrations and drafts. The corner of a room is usually
the most stable.
• Gently apply all items to be weighed. Do not drop
items onto the weighing platform.
• Avoid dropping your scale. The warranty does not
cover damage due to rough treatment or overload.
• Check the batteries rst if you are having any trouble
with your scale. This simple step can remedy most
scale issues.
• Do not disassemble your scale. This product contains
no user serviceable parts.
- English

ON
MEM
OFF
CAL
MODE
TARE
PRINT
I. Part List
1. Scale
2. Expansion Bowl
3. 200g Calibration Weight (x2) *LB-1000 and LB-501 ONLY*
4. AA Batteries (x4)
5. 6V Power Supply
II. Key Description
III. Battery Installation
1. Before using the scale for the rst time, check
the battery compartment and remove any plastic
insulation strips or packaging that may be in place to
prevent battery drainage during shipping/storage.
2. If you are replacing the batteries, make sure the
positive and negative contacts are properly aligned.
OFF Key
ON/MEM Key
CAL / MODE Key
Backlight ON/OFF
TARE / PRINT Key
+
-
+
-
+
-
+
-

LB-Series Manual
IV. Operation
1. Place the scale on a at, level surface and press the
ON/MEM key to turn the scale on.
2. Once the start-up process completes and the scale
goes to zero you are ready to weigh.
3. Place the quantity to be weighed onto the weighing
platform.
4. Wait a couple seconds for the display to stabilize
before taking a weight reading.
V. Changing the Unit of Measure
1. Press the MODE/CAL key to change the scales
displayed unit of measure.
2. If the scale is turned off then back on, it will default
to the last unit used.
• LB-501 Units: g (gram) oz (ounce) ct (carat)
gn (grain)
• LB-1000 & LB-3000 Units: g (gram) oz (ounce)
ozt (troy ounce) dwt (pennyweight)
VI. Using the Tare Feature
You can use this scale’s tare feature to subtract the weight
of an empty container or vessel for net weight determination.
NOTE: The weight of the container will subtract from the
available capacity.
1. Place the empty container or vessel on the weighing
platform.
2. Press the TARE/PRINT key once. The display will
then reset to zero. If the container is removed, it’s
weight will be displayed as a negative number.
3. You may now ll the container with the quantity to
be weighed. The display will show the net weight of
- English

the contents.
4. To return to normal weight display (gross weight),
remove any items from the weighing platform and
press the TARE/PRINT key again to re-zero the
scale.
VII. Counting Feature
1. To enter parts counting mode, press CAL/MODE to
scroll through the weighing modes until “C” appears
on the bottom left of the display. The minimum
sample quantity is 5 and will be displayed as “S=5”
2. Continue to press the CAL/MODE key to select the
desired parts sample quantity: 5, 10, 20, 50 pieces.
After “50” is displayed the scale will cycle back to
g (grams).
3. With the sample quantity value still displayed on
the LCD, place the sample quantity on the platform.
Allow the sample to stabilize for 3~5 seconds then
press TARE/PRINT. The sample quantity is now set.
4. You can now add any number of the item on the
scale and a count will be displayed.
VIII. Memory Function
1. With the balance on, press the TARE/PRINT key to
ensure the balance is zeroed.
2. Place an item on the scale and press ON/MEM
when the display is stable. This action adds the
displayed weight into the memory buffer which is
normally set to zero. The display will show “0.0”
again.
3. Repeat step 2 above as many times as you need,
until you are ready to see the cummulative result.

LB-Series Manual
4. To view the cummulative weight, press ON/MEM.
The total is displayed and memory cleared.
5. Press TARE/PRINT again to zero out the balance
and the scale is ready for normal weighing again.
IX. Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
LB-501 0.03g 400g OIML M2
LB-1000 0.2g 400g OIML M2
LB-3000 0.2g 2000g OIML M2
X. Adjustment
1. Press and hold the MODE/CAL key until the display
shows “CAL”.
2. Press the MODE/CAL key again. The display will
ash “CAL” followed by the calibration weight value
(see chart).
3. Place the calibration weight on the center of the
weighing platform. Make sure you are using the
correct weight for your specic model.
- English

4. Once the weight is accepted, the display will show
“PASS” then return to normal operation. Calibration
is complete!
XI. Troubleshooting
• Problem: Scale will not turn on
• Solution: Check the batteries
• Problem: Displayed weight uctuates randomly
• Solution: Make sure your workspace is stable and
free of any air currents or vibration
• Problem: Displayed weight is inaccurate
• Solution: Perform a calibration adjustment
XII. Error Codes
1. E - Calibration error. Repeat calibration.
2. O-Ld - Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. - Batteries are low. Replace the batteries.
LB-501 LB-1000 LB-3000
Max. Capacity 500g 1000g 3000g
Readability 0.01g 0.1g 0.1g
Power 4 x AA or
6VDC 300mA AC-Adapter
Dimensions 7.3 x 5.5 x 1.5”
Platform Dimensions 5.2” Ø
Warranty 10 Year Limited Warranty

LB-Series Manual
LB-501
500x0.01g
17.64x0.001oz
2500x0.05ct
7716x0.1gn
LB-1000 LB-3000
1000x0.1g
35.27x0.005oz
32.15x0.005ozt
643.1x0.1dwt
3000x0.1g
105.82x0.005oz
96.45x0.005ozt
1929x0.1dwt
- English

Merci d’avoir acheté cette balance numérique de poche
LB-Series d’American Weigh. Veuillez lire le mode d’emploi
attentivement avant utilisation. Les balances sont des appareils de
mesure de précision et doivent toujours être manipulées avec soin.
Pour assurer un fonctionnement able des années durant, veuillez
garder ces simples conseils à l’esprit :
• Ne pas dépasser la capacité maximale de la balance. Cette
balance possède une protection intégrée anti-surcharge,
mais même une protection anti-surcharge a ses limites!
• Éviter d’exposer la balance à des températures extrêmes,
chaudes ou froides. Les balances fonctionnent mieux à
des températures normales. Si la température a nettement
changé, un nouveau calibrage pourrait être nécessaire.
• Laisser la balance chauffer pendant 30-60 secondes avant
d’effectuer le calibrage (si disponible).
• Ranger la balance dans un endroit propre et sec. Poussière,
saleté et humidité peuvent s’accumuler sur les capteurs
de pesage et sur les circuits électroniques occasionnant
imprécisions et dysfonctionnements.
• Éviter d’utiliser la balance à proximité immédiate d’appareils
émettant des ondes radio tels que les téléphones sans l.
• Toujours utiliser la balance sur une surface plane et
horizontale, sans vibrations ni courants d’air. L’angle d’une
pièce est souvent l’endroit le plus stable.
• Poser délicatement les objets devant être pesés. Ne pas
laisser tomber les objets sur la plate-forme de pesage.
• Éviter de laisser tomber la balance. La garantie ne couvre
pas les dégâts causés par un traitement brutal ou une
surcharge.
• Vérier les piles si un problème est détecté sur la balance.
Cette étape simple permet de remédier à la plupart des
problèmes.
• Pour une meilleure précision, le calibrage doit être vérié et
éventuellement ajusté dans le cadre de contrôles de routine.

LB-Series Manual
ON
MEM
OFF
CAL
MODE
TARE
PRINT
XIII. Liste des composants
1. Balance
2. Récipient d’expansion
3. Poids d’étalonnage de 200g (x2) *UNIQUEMENT les modèles
LB-1000 et LB-501*
4. Piles AA (x4)
5. Alimentation électrique 6V
XIV. Description des touches
XV. Installation des piles
1. Avant la première utilisation de votre balance, vériez le
compartiment des piles et enlevez toutes les bandes isolantes
de plastique ou l’emballage qui pourraient s’y trouver an
d’éviter le déchargement des piles pendant le transport/
l’entreposage.
2. Si vous remplacez les piles, assurez-vous de bien installer les
contacts positifs et négatifs.
OFF Key
ON/MEM Key
CAL / MODE Key
Backlight ON/OFF
TARE / PRINT Key
+
-
+
-
+
-
+
-
- Français
Questo manuale è adatto per i seguenti modelli
3
Indice
Lingue:
Altri manuali AWS Scala

AWS
AWS 330CVS Manuale utente

AWS
AWS AMW-810 Manuale utente

AWS
AWS CP3-100 Manuale utente

AWS
AWS CP-300 Manuale utente

AWS
AWS AT Series Manuale utente

AWS
AWS EXTEND-5K Manuale utente

AWS
AWS LB-1000 Manuale utente

AWS
AWS AMW-DISC-55 Manuale utente

AWS
AWS CP3 Series Manuale utente

AWS
AWS CAMEO-5K Manuale utente