VII. Features
1.
Special design to take the Human Body Temperature with a
3cm~5cm (1.2~2 in)
distance from forehead.
2.
Reliable and stable measurement, thanks to the advantage
Infrared Detection
System.
3.Audible alarm if temperature is more than 38°C (100.4°F).
4.Memorize the last 32 temperature measurements.
5.
Three color backlits LCD digital display screen.
6.
Temperature unit can be displayed in either Celsius or Fahrenheit.
7.Automatic power-off (
<
30 secs) to conserve energy.
8.Longevity use (100,000 readings).
9.Practical, easy to use.
Additional usage:
JXB-178 can also be used to measure the temperature of a baby-bottle or bath (by
using the Surface Temp Mode), or room temperature (by using the Room Mode).
VIII. Instructions
1.Install battery.
2.For the first use or when inserting new Battery, wait from 10~15 minutes for the
warm-up of the unit. This will allow the unit to become acclimated to the
temperature of the room.
3.Press On/Scan button, aim towards the forehead (see the diagram below for the
JXB-178 positioning), from a distance of 3cm~5cm,
When press the “On/Scan”
button in the standby mode, the measuring
is done when the temperature is
showing in the screen or the beep is anouncing, measuring time is one second.
Tips: Do not move the position of the thermometer before
the testing is done.
4.Before taking the temperature, make sure to remove hair and perspiration from the
forehead.
IX. Setting and function of menu
1.Switch on the device
Press the “On/Scan” button, one second after the screen panel in full display, it will
enter the standby mode with the sign “---°C” or “----°F”.
Then press the “On/Scan” button again, you will get the measuring result in 1
second. But if there is no more operation, it will turn off in 30 seconds
automatically.
2.In the switch on state, Setting the mode
a. Press “MODE” button, and the screen will display: Body…°C
b. Press again “MODE” button and the screen will display: Room…°C
c. Press again “MODE” button and the screen will display:
Surface Temp…°C
Note: The thermometer default is set to BODY mode.
Important!
The surface temperature differs from the internal body temperature. To
obtain the internal temperature, always use the “BODY” mode. Please
make sure to select “SURFACE TEMP”mode for an external area reading.
3.F1:Choosing the temperature unit
In the switch on state, Press “MODE” button for 2 seconds,the screen will display
“F1”, then press “MODE” button to transfer between degree Celsius and
Fahrenheit, Comfirm by pressing “MEM” button.
4.F2:Alarm setup
In the switch on state, Press “MODE” button for 2 seconds,the screen will display
“F1”, then press “MEM” button once, the screen will display “F2”, press “MODE”
button to choose the alarming temperature from 37.3°C to 39.1°C(99.1°F to
102.4°F), Comfirm by pressing “MEM” button.
Note: The alarm threshold default value is 38°C(100.4°F)
5.In the switch on state, Press “MEM” (Memory) button, which will then display the
last temperature, and allows for a view of the last 32 measurements.
In the switch on state,Press "MEM" button and hold for 5 seconds, all data in
memory will be deleted. Then press "MEM" button again, the display will show
"CLr"
6.In the switch on state, press “ ” can open or close the buzzer.
When the screen shows “ON”, the buzzer opened.
When the screen shows “OFF”, the buzzer closed.
7.Recalibration of device via the F4 MENU
When there is the difference between JXB-178 and mercury thermometer, and you
believe mercury thermometer from its temperature but it is not convenient to use.
You can use recalibration function to adjust the JXB-178 to make it the same test
result with mercury thermometer after recalibration.
Besides, when you use JXB-178 for the people with different skin color (For
example: the yellow race, the white race, black people and so on) you can use
recalibration too.
Instructions for recalibration:
In the switch on state, Press “MODE” button for 2 seconds,the screen
will display “F1”, then press “MEM” button twice ,the screen will display
“F4”, press “MODE” button to choose the Offset value from -3°C to 3°C(-5.4°F to
5.4°F),
Comfirm by pressing “MEM” button.
In the cases of seasonal or environmental changes a verifica tion and adjustment
should be carried out.
8.CHANGING THE BATTERIES
Display: when the LCD screen displays the flashed symbol
" ", the battery is used.
Operation: Open the lid and change the batteries, taking great
care with the
correct positioning. A mistake with this could cause
damage to the apparatus
and compromise the guarantee of your
JXB-178. Never use rechargeable batteries. Use only batteries
for single usage.
X. Technical specifications
1.Normal using condition
Ambient temperature: 10°C ~ 40°C (50°F ~ 104°F)
Relative humidity: ≤85%
2.Batteries: DC 3V (2 pcs AA batteries)
3.Unit size: 155 x 100 x 40 mm (L x W x H)
4.Unit weight (without battery): 105g
5.Temperature display resolution: 0.1°C (0.1°F)
6.Measuring range:
In body mode: 32.0°C ~ 43.0°C (89.6°F ~ 109.4°F)
Under body mode, there is three color backlits:
Green color backlit: ≤37.3°C (99.1°F), means normal temperature.
Orange color backlit: 37.4°C ~ 37.9°C (99.3°F ~ 100.2°F), means low fever.
Red color backlit: ≥38°C (100.4°F), means high fever.
In surface temp mode: 0°C ~ 60°C (32°F ~ 140°F)
In room mode: 0°C ~ 40°C (32°F ~ 104°F)
7.Precision:
32.0°C ~ 34.9°C (89.6°F ~ 94.8°F) ±0.3°C(±0.6°F)
35.0°C ~ 42.0°C (95°F ~ 107.6°F) ±0.2°C(±0.4°F)
42.1°C ~ 43.0°C (107.8°F ~ 109.4°F) ±0.3°C(±0.6°F)
8.Consumption: ≤300mW
9.Accuracy: ± 0.3°C (0.6°F)
10.Measuring distance: 3cm ~ 5cm (1.2in ~ 2in)
11.Automatic power-off: <30 secs
12.Memory:32 sets
※ Note: The Non-contact Infrared Thermometer Model JXB-178 can take
temperature readings below 32.0°C or above 43.0°C (89.6°F to 109.4°F) but
precision is not guaranteed outside of this range.
Longevity of the product
The JXB-178 was conceived for an intense and professional use, its
longevity is guaranteed for 100,000 takings.
XI. Maintenance of the product
- The protective glass over the lens is the most important and fragile part of the
thermometer, please take great care of it.
- Clean the glass with cotton fabric, wet with 95° alcohol.
- Do not use other batteries than mentioned batteries, do not recharge non
rechargeable batteries, do not throw in fire.
- Remove the batteries when thermometer is not used for an extended period of
time.
- Do not expose the thermometer to sunlight or water.
- An impact will damage the product.
XII. Accessories
User Manual in English 1 pc
XIII. Guidelines
This device complies with the EU Directive 93/42/EEC concerning
medical
products, the ISO 80601-2-56 and the European Standard
EN60601-1-2 and
is subject to particular precautions with regard to electromagnetic compatibility.
XIV. Troubleshooting
If you have problems while using your thermometer, please refer to this
guide to help resolve the problem. If the problem persists, please contact
our customer service.
THE SCREEN DISPLAYS TEMPERATURE HIGHER THAN 43.0°C
(109.4°F):
The temperature is in Fahrenheit. Change the measurement to Celsius.
THE SCREEN DISPLAYS TEMPERATURE LOWER THAN 32°C
(89.6°F):
To take the surface temperature, press the “MODE” button and set to the
reading called “Body” , If the device is in Surface Temp Mode, the 32°C
(89.6°F) temperature displayed is showing the external temperature of your
body, rather than the internal.
THE SCREEN DISPLAYS THE MESSAGE HI
When using the JXB-178 Thermometer, the
message “HI” can show on the screen.
In this case,the temperature is above the
measurement range selected, either
above 43.0 °C (109.4°F) in Body Mode.
THE SCREEN DISPLAYS THE MESSAGE LO
When using the JXB-178 Thermometer, the
message “LO” can show on the screen.
In this case, the temperature analyzed is
under the measuring range selected, either
less than 32°C (89.6°F) in Body Mode.
This message displays for various reasons. Please find below a list of the main
issues:
4 5 6 7
Body
Body
Reasons for LO
message display Advice
Make sure there is no obstruction or
dampness prior to taking temperature.
Make sure there is no air blowing
in the area of use; this could affect the
infrared reading.
Pause for 3~5 seconds minimum
between readings; a 15 seconds
pause is recommended.
Take measurements at the
recommended distance (app. 3~5 cm;
1.2in~2.0in).
Temperature reading hampered
by hair or perspiration.
Temperature hampered by an air
draft or dramatic change in ambient
temperature.
Temperature readings are too close
together, and the thermometer did not have
time to reboot.
The measuring distance is too far.
Symbol Reference
Trade mark
IEC 60417-5333,Type BF applied part
IEC 60417-5031 Direct current
Protected against access to hazardous parts with a finger and against
vertically falling water drops when enclosure tilted up to 15°
Refer to instruction manual / booklet
DISPOSAL: Do not dispose this product as unsorted municipal waste.
Collection of such waste separately for special treatment is necessary.
This symbol shall be accompanied by the name and the address of
the manufacturer
Specifies serial number
SN
XV. Explanation of symbols
XVI. EMC Declaration
Guidance and manufacturer’s declaration – electromagnetic immunity
The “JXB-178” is intended for use in the electromagnetic environment
specified below. The customer or the user of the “JXB-178” should ensure
that it is used in such an environment.nt.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Electrical fast
transient/burst
IEC 61000-4-4
±6 kV contact
±8 kV air
Not Applicable
Floors should be wood, concrete
or ceramic tile. If floors are covered
withsynthetic material, the relative
humidityshould be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Not Applicable
Mains power quality should
be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not Applicable
Mains power quality should
be that of a typical commercial
or hospital environment. If the
user of the “JXB-178” requires
continued operation during
powermains interruptions, it is
recommended that the “JXB-178”
be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercialor hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
±2 kV for power
supply lines
±1 kV for input/
output lines
Guangzhou Berrcom Medical Device Co., Ltd.
Address: No.38 Huanzhen Xi Road, Dagang Town, Nansha,
511470 Guangzhou, Guangdong, PEOPLE’S REPUBLIC OF CHINA
Tel: +86(20)34938449
Fax: +86(20)34936960
Wellkang Ltd
Address: The Black Church,St. Mary's Place,
Dublin 7, D07 P4AX, Ireland
Tel: +353(1)4433560
Web: www.CEmark.com
Made in China
Guidance and manufacturer’s declaration – electromagnetic immunity
The “JXB-178” is intended for use in the electromagnetic environment specified
below. The customer or the user of the “JXB-178” should ensure that it is used in
such an environment.
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the “JXB-178”, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
d=[ ]√P
d=[ ]√P 80MHz to 800MHz
d=[ ]√P 800MHz to 2.5 GHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz 3 V/m
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site survey,a
should be less than the compliance
level in each frequency
range.b Interference may
occur in the vicinity of
equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength
in the location in which the “JXB-178” is used exceeds the applicable RF
compliance level above, the Medical JXB-178 should be observed to verify
normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the “JXB-178”.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
a
b
Not Applicable
3.5
V1
3.5
E1
7
E1
Recommended separation distances between
portable and mobile RF communications equipment and the Medical JXB-178
The “JXB-178” is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Medical
JXB-178 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the “JXB-178 as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
0,01
0,1
1
10
100
/
/
/
/
/
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80
MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and people.
d=[ ]√P d=[ ]√P
3.5
E1
d=[ ]√P
7
E1
3.5
V1
MODE BUZZER MEM