OPERATING INSTRUCTIONS
636
INFRARED THERMOMETER
WITH LASER SIGHTING
INTRODUCTION
This instrument is a portable easy to use 3½digit,
compact-sized digital infrared thermometer with laser
sighting designed for simple one hand operation. Meter
comeswithBacklitLCDdisplay, Auto-Holdfunctionand
auto power down (10 seconds approx.) after releasing
Trigger to extend battery life.
SAFETY INFORMATION
It is recommended that you read the safety and operation
instructions before using the infrared thermometer.
CAUTION
• Do not use the unit near any device which generates
strong electromagnetic radiation or near a static
electrical charge, as these may cause errors.
• Do not use the unit where it may be exposed to
corrosive or explosive gases. The unit may be dam-
aged, or explosion may occur.
• Do not keep or use this unit in an environment where
itwillbedirectlyilluminatedbysunshine,orwhereit
willbeexposedtohightemperatures,highhumidityor
condensation. If you do, it may be deformed, its
insulation may be damaged, or it may no longer
function according to specification.
• Do not point the lens at the sun or at any other source
ofstronglight.Ifyou do,thesensormaybedamaged.
• Do not contact the lens against the object whose
temperature is to be measured, or get it dirty, allow it
to be scratched, or allow any foreign material to
adhere to it. Doing so may cause errors.
• Do not touch or hold by the front case. Temperature
reading can be affected by heat from hand.
• Donotplacethemeteronoraroundhotobjects(70°C/
158°F). It may cause damage to the case.
• If the meter is exposed to significant changes in
ambienttemperature(hottocoldorcoldtohot).Allow
20 minutes for temperature stabilization, before tak-
ing measurement.
• Condensation may form on the lens when going from
acoldto hot environment-wait 10 minutes for conden
sation to dissipate before taking measurements.
• This unit is not constructed to be water proof or
dustproof,sodonotuseitinaverydustyenvironment
or in one where it will get wet.
SPECIFICATIONS
GENERAL
Display:
3½ digit liquid crystal display (LCD) with maximum
reading of 1999
Low battery indication: the " " is displayed when
the battery voltage drops below the operating level
Measurement rate: 0.25 second, nominal.
Operating Environment: 32°F to 122°F (0°C to 50°C)
at < 70% R.H.
Storage Temperature:
-4°F to 140°F (-20°C to 60°C) , 0 to 80% R.H. with
battery removed from meter
Auto power off: 10 seconds.
Standby consuming current: <5µA
Battery: Standard 9V battery (NEDA 1604, IEC 6F22
006P)
Battery Life: 9 hours (continuity) typical
(contain Laser and Backlit)
Dimensions(HxWxD): 5.8x4.1x1.65"(148 x 105 x
42mm).
Weight: approx. 157g (including battery.)
Laser Specifications
Laser safety classification of Class 2
Wave Length: Red (630~670nm).
Power out: <1mW, class 2 laser product.
DANGER
Pressing the Trigger turns the laser beam on and off.
Exercise extreme care and do not allow the laser beam
toenter your eye or those of any other person or animal.
• Donotlookdirectlyintothelaserlightfromtheoptical
system.
• When measuring the temperature of an object which
has a mirror finish, be careful not to allow the laser
lightbeamtobereflectedoffthesurfaceintoyoureyes
or those of another person.
• Donotallow thelaserlight beam toimpingeupon any
gas which can explode.
EMC/RFI
Readings may be affected if the unit is operated within a
radio frequency electromagnetic field strength of approxi-
mately 9 volts per meter, but the performance of the
instrument will not be permanently affected.
®
ELECTRICAL
Temperature Range: -22°F to 1022°F / -30°C to 550°C
Display Resolution: 1°F, 0.5/1°C (Auto)
Accuracy:
±(4°F/2°C) for -22°F to 212°F, -30°C to 100°C
±(2% reading) for 213°F to 1022°F, 101°C to 550°C
TemperatureCoefficient: ±0.2%ofreadingor±0.36°F/
0.2°C, whichever is greater, change in accuracy per °F/
°Cchangeinambientoperatingtemperatureabove82.4°F/
28°C or below 64.4°F/18°C.
Response Time: 0.25 second
Spectral Response: 6 to 14µm nominal
Adjustable emissivity (εε
εε
ε): 0.1 to 1.0
Detection Element: Thermopile
Optical Lens: Fresnel Lens
Sighting: 1-beam laser marker <1mW (class 2)
Field of View: 100mmØ at 1000mm (3.9"Ø at 39.0")
Spot size increases with distance from the probe tip as shown
(Spot Diameter measured at 90 % Energy)
OPERATING INSTRUCTIONS
Trigger
Pull the trigger to turn on the meter for measuring
temperature. Releasing trigger to stop measuring tem-
perature and automatically hold the display
reading, the meter power down automatically after 10
seconds.
Button function:
1. Set button
SET MODE & Numeric input key
"SET" annunciator appears when a numerical value can be
set ( during setting of ε, ALM Hi and ALM Lo).
key: The numerical value is increased.
key: The numerical value is reduced.
If either of these numerical value keys is held down, the
numerical value changes rapidly in the appropriate
direction.
The data will store in nonvolatile storage region while
after completingsettings.Inthismodethe automatic power
-off feature is disable.
2. " " button
Press " " button to on the " " annunciator. If " "
annunciator is on, press trigger and the laser beam will turn
onand " "annunciatorwillblink.Releasingtriggertoturn
off the laser beam.
3. " " button
Use " " button to select turn on or off the Back-Light
function.
4. °C/°F button
Readings are displayed in either degrees Celsius(°C) or
degrees Fahrenheit(°F). When the thermometer is turned
on. To change the temperature scale by pressing °C/°F
button .
5. APO button
It will auto power off for about 10 seconds.
Press "APO" button to disable . Auto Power-Off function
that HOLD indication disappears and press again to enable
APO function.
Press SET button switches the mode around the cycle ε!
ALM Hi !ALM Lo !MAX !MIN !HOLD.
HOLD :Releasing the trigger to stop measurement of
temperature, the HOLD indication appears,
and the measured temperature is held.
ε:The thermal emissivity of the object set using
the and keys. (refer to Table 1)
ALM Hi :The upper limit alarm temperature is set using
the and keys. When the measured
temperature is exceeded the Hi setpoint, the
beeper emits a discontinuous pulse tone and
"ALM Hi" is displayed.
ALM Lo :The lower limit alarm temperature is set using
the and keys. When the measured
temperatureisbelowtheLosetpoint,the beeper
emits a continuous pulse tone and "ALM Lo"
is displayed.
MAX :The maximum temperature during measure-
ment is displayed.
MIN :The minimum temperature during measure-
ment is displayed.