Clements SUC89660 Manuale utente

User Manual
CLEMENTS
Duplex Adaptors
SUC89660
SUC89661
Manual No. SUC 89660 081
Issue 10

2
CLEMENTS
Safety
Thank you for purchasing this Clements Duplex Adaptor.
For your safety it is imperative that this unit only be
operated by authorised personnel in accordance with the
instructions as described in this manual. Operated in this
way the Duplex Adaptor will provide years of service.
Due to continual improvements in product design the
Duplex Adaptor may vary in detail from the descriptions in
this manual. In the event of further questions please
contact your local distributor or BMDi TUTA Healthcare
direct.
User Manual
Duplex Adaptor
Manual Number SUC 89660 081 Issue 10
Copyright © 2014 BMDi TUTA Healthcare Pty Ltd
The information in this manual was originated by and is the exclusive
property of BMDi TUTA Healthcare Pty Ltd. It is furnished for customer
information only and is not an authorisation or licence to make this
product or to furnish this information to others.
BMDi TUTA Healt care Pty Ltd
Unit 4B, 128-130 Frances Street
Lidcombe NSW 2141 Australia
P one: +61 2 9466 5300
Website: www.medaust.com
C0086
Priory Analysts Ltd
The Pinnacle, 160 Midsummer Blvd
Milton Keynes, MK9 1FF, United Kin dom
EC REP

3
CLEMENTS
Contents
Contents
Specifications .................................................. 4
Transportation and Storage ............................ 5
Installation and Operation .............................. 6
Spares Parts………………………………. .............. 7
Troubles ooting………………………………………. 8
Cleaning .......................................................... 9
Disassembly .................................................... 10
Warranty .......................................................... 14

4
CLEMENTS
Specifications
Vacuum Range 0 to -100 kPa, 0 to -750 mmHg
Flow Rate 0 to 35 Litres of Free Air Per Minute
Lever No Lever
Height 100 mm 40 mm
Width 75 mm 75 mm
Depth 80 mm 80 mm
Weight 300 g 250 g
Duty Cycle Continuous
Ambient Temperature +5°C to 35° C
Standard Conditions 25°C, Sea Level, 101 kPa
Class Class I (EU Class IIa)
ARTG 174693
GMDN 14325
Specifications
Intended Use
Connect two pieces of medical gas equipment to a single
medical gas terminal unit.

5
CLEMENTS
Transportation and Storage
Environmental conditions for transportation and storage are
shown in the following table.
Parameter Minimum Maximum
Temperature 10ºC 40ºC
Humidity 60% RH 95% RH
Barometric Pressure 700 mBar 1060 mBar
Warning Symbols Legend
The warning symbols marked on the equipment and their
meanings are shown as follows.
Caution, consult accompanying documents
Transportation and Storage
Waste Materials
The contents of the collection jars suction tubing bacteria filter
internal exhaust filter may contain biohazard wastes. Handle using
safe handling procedures which may include the use of rubber
gloves and eye protection and dispose of according to local
protocols for biohazard materials.
Recycling
At the end of their service life the Nozzle should be dismantled if
necessary and returned to a local materials recycling centre.

6
CLEMENTS
Description
The Clements Duplex Adaptor allows the connection of two
pieces of equipment to a single wall terminal unit.
Clements Duplex Adaptors are available to connect to most
types of wall outlet.
Installation and Operation
Carefully examine the Dupledx Adaptor for any visual signs
of damage that might have occurred during shipment.
Attach the adaptor to the wall outlet with the index
handwheel or relevant connector.
1.
Ensure any equipment is turned off before connecting to
the duplex adaptor.
2.
Select the outlet to use by moving lever (lever model only)
fully to corresponding position.
Installation and Operation
NOTE:
Only ONE piece of equipment can be in use
at any given time. Ensure t at equipment
connected to t e ot er outlet is switc ed
off.
Item Description
SUC 89660 Duplex Suction Adaptor without Lever, Ring Index
SUC 89661 Duplex Suction Adaptor with Lever, Ring Index

7
CLEMENTS
Spare Parts
SUC 89771 Ring Index Connector Kit for Wall
Suction Products ( Handwheel, 1/8 BSP
Threaded Nipple, O-Ring)
SUC 89240 019 Puritan Bennett Connector with 1/8 BSP Thread
SUC 89240 074 BS MKIV Connector with 1/8 BSP Thread
SUC 89250 006 Drager Connector with 1/8 BSP Thread
SUC 89660 Duplex Suction Adaptor with Lever
SUC 89661 Duplex Suction Adaptor Without Lever
SUC 89660 081 User Manual for Duplex Adaptors
Select from a range of Medical Suction Tubing Collection
Jars Bung Assemblies and Handpieces
Spare Parts

8
CLEMENTS
Fault Check Rectify
No Vacuum / Flow On / Off Lever Ensure turned on
Wall outlet blocked Refer to engineering
dept.
Body blocked Strip and clean
Troubles ooting
Troubles ooting

9
CLEMENTS
Cleaning
Cleaning
Clean using a pH neutral disinfectant for wiping or immersion
cleaning.
Flush after immersion using distilled water and dry before
reassembly of unit.
Cleaning Procedure
Tools and Materials Required
• Allen key set imperial
• Vacuum grease
• Can of spray Dow Corning Molykote D-321R Sealer
• Hand press and fixture for holding body
Cleaning

10
CLEMENTS
Disassembly Procedure
1. Remove unit from wall.
2. Remove the hand wheel by inserting an allen key into the
brass nipple in the centre of the hand wheel. Make sure
the allen key is pushed home into the brass nipple before
unscrewing.
4. Remove the Sel-Lok pin disc springs and O-ring using a
hand press and fixture.
5. Push the plug out by gently twisting and pushing. Once
the plug is out gently clean the plug with a soft brush
using a degreaser. Wash and blow-dry.
Use the same care when cleaning out the bore in the
nozzle body.
6. Lightly coat the plug with Molykote surface spray and let
dry.
7. Apply a light coat of vacuum grease to the body bore.
8. To assemble - reverse the order
9. To test - lever should be firm and smooth to turn vacuum
when tested should read the same as the vacuum source
flow should have a max of 2-3 lpm loss only.
10. Clean any vacuum grease off the body.
Cleaning Procedure