8.1. SmartPour Calibration - Items Required
a) 1 x Empty bottle
b) 1 x Measuring Scale (if measuring by weight/density)
c) 1 x Measuring Cylinder (if measuring by volume)
d) 0.8L Water or Actual Spirit
8.2. SmartPour Calibration - Calibrate Button 1
a) Fill up 1 empty bottle with at least 0.8L of water/spirit. Place the filled bottle under the SmartPour. Ensure
that the inlet nozzle is submerged in the water.
b) Turn on the SmartPour by touching the “Power” button.
c) Ensure that the prime step is completed before proceeding. Refer to Section 7, Step 3.
d) Obtain a clean and empty bottle and tare the empty bottle on a weighing scale.
e) Place the empty bottle under the dispense nozzle.
f) Touch “Button 3” and measure the weight of the dispensed liquid. This is the Measured Weight (MW). The
Target Weight (TW) is Desired Dispense Volume ± 2%.
g) Calculate the difference between TW and MW. If the difference is positive, it means the reference point
needs to be increased. If the difference is a negative value, it means the reference point needs to be
decreased.
h) Touch 8 times until a beep is heard. The screen interface will change indicating that the SmartPour is
now in Calibration Mode. The dispensed volume linked to Button 3 will display and blink repeatedly.
i) The default reference point for “Button 3” will be reflected on the display. E.g.: 125 or 135.
Note: The number may vary because the unit is pre-calibrated at the factory.
j) Touch “Button 4” to calibrate the Button 1 refill. Touch “Button 1” to increase the offset and touch “Button
2” to decrease the offset
k) Every 3-reference point will increase/decrease by approximately 1 ml. Touch “Button 1” to increase the
reference point and touch “Button 2” to decrease the reference point. Based on the calculated difference,
adjust the reference point accordingly.
l) Once done, tap on to save and exit calibration mode.
m) Touch “Button 3” and measure the dispensed volume. If the newly adjusted volume is still not accurate,
repeat steps d-l again.