ENSIM SENSORS ECAS 101 Manuale utente

OPERATING
MANUEL
CAPASITIVE LEVEL SWITCH
ECAS
Model :
Model : 82-2021-008
Information in this manual is reviewed and completely reliable. Responsibility is not assumed due to any typing error.
Products in this manual are available only for information purpose and they may be changed without notice.
ECAS 101 / 102 / 103 / 107
ECAS 202 / 203 / 204 / 205 / 20S
ECAS 301 / 304 / 305 / 30D / 30S
ECAS 408A / 408B / 408T / 408Tp / 408Tm
Models :
1
Important Notes:
Used Symbols :
: Caution : Note : Disposal
Please read this manual carefully before installation of the capasitive level switch. User is responsible for accidents and losses arising
from failure to comply with the warnings in this manual.
In the event that capasitive level switch is broken, take measures in order to prevent accidents and losses which can occur in its system.
There is not any fuse and circuit breaker on the instrument; they should have been added to the system by the user.
This manual should be stored in an easily accessible place for subsequent use.
The manufacturer's liability cannot exceed the purchase price of the device according to the law.
Do not make any modification on the instrument and do not try to repair it. Reparation should be made by authorized service staff.
Do not operate the system before making assembly in compliance with the assembly chart related to the instrument.
Products which do not contain label and serial number are considered to be excluded from the warranty scope.
The instrument's useful life, determined and announced by the ministry, is 10 years.
Table of Contents :
General Information...........................................................................................................................................................................2
Installation .........................................................................................................................................................................................5
Failure Detection..............................................................................................................................................................................15
Disassembly of Instrument ............................................................................................................................................................15
Service .............................................................................................................................................................................................15
Recalibration....................................................................................................................................................................................15
Repair................................................................................................................................................................................................15
Disposal ...........................................................................................................................................................................................15
Terms of Warranty ..........................................................................................................................................................................15
Terms of Return ................................................15
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.

2
1. General Information :
1.1. Material Acceptance
Check that there is no damage on the packages during the transportation immediately after the material acceptance. If packages are
damaged, open the packages immediately and check whether products are affected or not, if there is any damage, send your
complaint report to the transporter company and its photocopy to the address of our company.
1.2. Information about Areas of Use
ECAS level switch is a capacitive level sensor for level measurement of conductive liquid, nonconductive liquid, granulated materials with
solid particles, adhesive and acid / base liquids. When a material comes between electrode rod and tank wall, a capacitance change
occurs and when this change exceed adjustment threshold, contact output is delivered.
Designed for difficult process conditions. Models with cooling apparatus can be manufactured for high temperature and pressure conditions.
Calibrations of triggering point and relay operation range can be performed by the user under workplace conditions.
It can be connected horizontally or vertically.
Application Areas
Liquid tanks, food machines, cooling liquid tanks, shipping, glycol tanks, brine, waste water tanks.
Oil tanks, CO2 liquid tanks, high temperature tanks, low-conductive liquids.
Grain silo, cement silo, sand feed, dough, milk powder, organic and plastic granule.
Sticky hot and high viscosity liquid, acid and chemical liquids.
Ambient Conditions: Relative Humidity: 0-98 %RH Ambient temperature: 60C (It is not used under -20 C)
Capacitance definition, assuming two parallel conductive plates are used;
1.3. Working Principle
d
C
R
Isolation
Metal Tank
S
Impedance
Distance
Capacity
Distance
However, there are scarcely any sensor type which this definition
can be pratically utilized. Above Formula can no longer
is reliable especially when residual areas increase due to large
distance (d) (which is usually the case). Thus, measuring
impedance for distance measurements gives more accurate
results than capacitance measurement.
C: capacity , Farad
S: Surface Area , m²
d: Distance , m
eo.er.S
d
C=
e
Excitation applied between 10 KHz250 KHz based on length
for all our models. ( =2 x p x f) Linearity error that may be caused
by conductivity component (R) effect is prevented by electronic
circuit design and mechanical design. Reduced to a level lower than
1ppm, acceptable as zero.
e
Impedance definition Z = R + jL + ( jC )-1
e
R is defined as real component and represent ambient conductivity.
Advantages :
Optionally high temperature-resistant
design.
Easy assembly and sensitivity
adjustment.
No need to clean.
Not affected by foam, liquid splash and
probe coating.
Can be mounted upside down (reverse)

3
2.p.eo.er
In (b/a)
Impedance is calculated by this definition
C=
jLw second component is defined as inductive reactance. This component is present even if we perform capacitive
measurement. However we neglect this. Since we evaluate results based on electrostatic properties of the environment,
no error will occur. Resulting impedance definition is Z= R + (jC )-1.
e
Measurement is made by charge transfer in our capacitive sensors. Total impedance is defined as Z = V / I.
I (current) I = Q/t
Q (Coulomb)
T (sec)
Capacitive reactance we desire to measure is (jC )-1. Meaning that charge and impedance have the same phase.
To summarize, charge transferred to medium is directly proportional with capacitive reactance.
For sensors manufactured as coaxial;
a: Central electrode radius
b: Outer screen radius
L: length
e
e
Excitation applied between 10KHz250KHz based on length for all our models. ( =2xpxf)
Linearity error that may be caused by conductivity component (R) effect is prevented by electronic
circuit design and mechanical design. Reduced to a level lower than 1ppm, acceptable as zero.
-20 40 80 2000
16
20
100
-40 150
Pprocess (bar)
T process (°C)
Process Pressure / Temperature Chart
With Cooling Apparatus
40
80
0-40 100 200150
T ambiance (°C) Ambiance Pressure / Temperature Chart
With Cooling Apparatus
T process (°C)

4
1.4. Technical Specifications and Material Knowledge
1.5. Label Information :
1 NA / NK x5 A / 250 VAC Relay
Opt. Peek, CeramicPTFE, PFA
PBT (Std.)
Ops. Aluminum Injection, Stainless Steel
er
24 VDC
(-) 1 bar...100 bar (Depending on the model.)
304 St.St. Opt. 316 St.St.
1,6
With LED-Power and Contact LED
Max. 1 W
Terminal
EMC, Low Voltage
285 g. for ECAS 101
(-) 20 °C...(+) 60 °C
PBT
- IP 66 , Alüminyum , Pas. Çelik IP 65
Max. 40 Nm
Max. 500 V
(-)40/(+)150°C (Depending on the model)
200°C with cooling apparatus
230°C with PEEK isolation
400°C with ceramic isolation
Supply
Signal Output
Working Pressure
Min.Di-Electric Constant
Measurable Metarial
Connection Material
Display
Power Consumption
Protection Class(EN60529)
Test
Max. Tensile Force
Weight
Isolation Material
Housing Material
Working Temperature
Ambient Temperature
Electrical Connection
Isolation
Conductive liquids, refrigerants
Non-conductive liquids
Solids particulate materials
Adhesive and acid/basic liquids
LEVEL SWITCH
Type
Output
Supply
S.N.
:
:
:
:
ECAS101-Ø10-300mm-0-0006-11- 5239-066-002-00/0
1XNO/NC
24 VDC ± %10
IE20060002-0001
Model
Manufacturer
Website
Serial
Number
www.ensim.com.tr
MADE IN TURKIYE
Warning
Symbols
Warning
Symbols
Approval SignManufacturer Brand Product Name
Product Label :
Protection Case:
Material : 304 Stainless Steel
Welded manufacturing
Opens - Closes Hinged
To Protect Against external conditions.

5
1.6. Target Group
This operating manual has been prepared for qualified technical personnel.
1.7. Security Notes
Please consider below notes for avoiding hazardous cases around operator and its environment.
Only users that authorized in work safety and whom also have read the manual should setup this equipment.
Work safety , accident avoid regulations and national/local setup standards must be meet.
Equipment should only be used in stated specifications.
Equipment must only be mounted while there is no pressure.
1.8. Package and package contents
Please check whether you have taken delivery of below listed content completely or not and check its conformity with criterions in your order:
*Capasitive Level Switch
*This operating manual
2. Installation :
2.1. General Notes :
The device installation is in 2014 / 34 / EU criteria to ensure the safety of atmosphere and people from explosions, must only be done by staff who
knows the safeguards.
Do not apply force to the instrument during the installation!
Do not use the Capacitive Level Switch with a greater pressure than recommended pressure.
Do not forget that instrument is precise, carry it carefully and prevent not to be damaged.
It should be guaranteed that there are not any magnetic particles.
The Max. working pesure should not be exceeded.
2.2. General Installation Stages
*Remove Capacitive Level Switch from the box carefully
*Check whether gasket is appropriate for fluid or not. If is not appropriate, contact with the producer.
*Then, apply below mentioned explanations according to structure of the design.
2.3. Special Notes
*Please ensure that there is no mechanical stress on the shaft following installation. Such case will cause slipping in thecharacteristic curve.
*Capacitive Level Switch must be placed upright or horizontal.
*Allocate valve certainly in the process connection while instrument is used.
*Allocate blowdown valve under bottom flange for blowdown.
*If instrument is mounted outside and if there is any danger of lightning or
excessive pressure, take preventive measures by taking necessary measures.
*In the operating conditions, Capacitive Level Switch may be hot according
to situation of fluid, in this case, do not touch the indicator, otherwise your skin is damaged.
*The grounding product must be done properly. (can be done outside or in housing)
2.4. Installation For Mechanical Connections
*Use appropriate O-Ring or gasket for tightness.
*Ensure that its surface is clean and smooth.
*Assemble the instrument manually.
*Connect the contacts as shown in the figure.
(For stainless steel models max. 30Nm)

6
Trigger Point
Sensation Zone
24 VDC
NA
NK
Com
+-
Sample Model :
Cooling :
27 HEX
1/2 BSP
PFA
L
68
For Max. 200 °C
2.5. Electrical Installation
Make the electrical connection of the instrument according to details on its label, table and cable figures in this manual
RlyLED: Relay active during normal operation; means operation continues during
calibration. Flashes continuously in normal operation mode if relay is
active and blinks in calibration mode. It is red colored.
PwrLED: Means there is no sensor failure during normal operation, and means
desired measurement values are saved in memory during calibration.
Operates by flashing. If light is continuous, it indicates failure. It is gren colored.
CAL - S Button: Used to acquire High Level-span-value during calibration.
CAL - Z Button: Used to acquire Low Level-zero-value during calibration.
TST Button: During normal operation, functions as Relay Test; during calibration,
performs saving to nonvolatile memory of Zero-Span, the values previously acquired
by S and Z button, function.
LVL - C Pot: Adjusts relay triggering point between Zero-Span values.
LVL - F Pot: Performs as fine tuning for triggering point. Adjustment field is equal
to +/- 5% of the point adjusted by C Pot (total 10%).
DIF Pot: Adjusts Release level of the relay activated by C/F Pot.
Highest adjustable value is equal to half (50%) of the operation region specified by
Z and S. Meaning that, when DIF Pot is at 100% and relay is pulled, the level to
release it shall be reduced as half of the total scale.
*
*
*
*
*
*
*
*
Indification and Calibtarion :
Electronic Unit with Cable:
Electronic unit and sensor component can be separated by a cable that protected against
exterior conditions for easy calibration on site. Thanks to the properties of cable, easy assembly
for user is possible without affecting capacitive measurement.
Relay working range
adjustable. (50%)
L

2.6. Mechanical Dimensions
7
CONDUCTIVE LIQUIDS
Sample Models
27 HEX
PFA
L
1/2BSP
(-) 1 bar...(+) 100 bar
(-) 40 °C...(+) 150 °C
Fully Insulated Probe
Conductive Tank
ECAS 101
L= 250 mm. (Std) Max. 4 m.
200
St. St.
St. St.
2BSP
70 HEX
Fully Insulated Rope
Conductive Tank
L= 1 m. (Std) Max. 32 m.
ECAS 107
Ø40
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
PFA
L
Plastic
ECAS 102
Fully Insulated Coaxial Probe
Insulated Tank
L= 250 mm. (Std) Max. 4 m.
70 HEX
L
2BSP
PFA
(-) 1 bar...(+) 100 bar
(-) 40 °C...(+) 150 °C
St. St.
Plastic
PFA
3/4BSP
32 HEX
L
L= 250 mm. (Std) Max. 1 m.
Fully Insulated Coaxial Probe
Insulated Tank
ECAS 103
(-) 1 bar...(+) 100 bar
(-) 40 °C...(+) 150 °C
St. St.
Plastic
200
St. St.
Ø40
St. St.
1 1/2BSP
60 HEX
PFA
Fully Insulated Rope
Conductive Tank
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
L= 1 m. (Std) Max. 16 m.
ECAS 107
L
Plastic Plastic
80
80
11

8
ECAS 202
LOW CONDUCTIVE LIQUIDS
70 HEX
L
2BSP
St.St.
32 HEX
L
3/4BSP
ECAS 203
St.St.
St.St.
1 1/2BSP
60 HEX
200
Ø40
St.St.
ECAS 204
St.St.
L
ECAS 204
2BSP
70 HEX
200
Ø40
St.St.
St.St.
St.St.
L
41 HEX
L
1BSP
Ø12
PTFE
ECAS 205
L= 250 mm. (Std) Max. 6 m.
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
St.St.
ECAS 20S
60 HEX
L
1 1/2BSP
St.St.
Alüminyum
290
Ø38
Plastic Plastic Plastic
L= 250 mm. (Std) Max. 4 m.
(-) 1 bar...(+) 100 bar
(-) 40 °C...(+) 150 °C
L= 250 mm. (Std) Max 1 m.
(-) 1 bar...(+) 100 bar
(-) 40 °C...(+) 150 °C
L= 250 mm. (Std) Max. 4 m.
(-) 1 bar...(+) 25 bar
(-) 40 °C...(+) 400 °C
L= 1 m. (Std) Max. 16 m.
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
L= 1 m. (Std) Max. 32 m.
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
Partly Insulated Coaxial Probe
Conductive / Insulating Tank
Partly Insulated Coaxial Probe
Conductive / Insulating Tank
Partly Insulated Probe
Conductive Tank
Partly Insulated Rope
Conductive Tank
Partly Insulated Rope
Conductive Tank
Partly Insulated Coaxial Probe
Conductive / Insulating Tank
Plastic Plastic Plastic

9
PTFE
Ø21
80
L
8050
L
ECAS 301
L= 250 mm. (Std) Max. 1 m.
(-) 1 bar...(+) 25 bar
(-) 40 °C...(+) 150 °C
SOLID PARTICULATE MATERIALS
41 HEX
1BSP 41 HEX
ECAS 305
St.St.
St.St.
Antistatic
For Materials
PTFE
Ø40
50 80
200
L
60 HEX
1 1/2BSP
PTFE
St.St.
St.St.
ECAS 304
L= 1000 mm. (Std) Max. 16 m.
(-) 1 bar...(+) 25 bar
(-) 40 °C...(+) 150 °C
L= 380 mm. (Std) Max. 1 m.
ECAS 30D ECAS 30S
Seramic Insulated Probe
Insulating Tank
41 HEX
Ø21
1BSP
St.St.
St.St.
Ø12
80508050
L
Aluminium
51 HEX
215
High Temperature Version
Seramic
60 HEX
Ø12
L
180
St.St.
L= 300 mm. (Std) Max. 4 m.
Plastic Aluminium
L= 250 mm. (Std) Max. 6 m.
(-) 1 bar...(+) 25 bar
(-) 40 °C...(+) 150 °C
(-) 1 bar...(+) 25 bar
(-) 40 °C...(+) 200 °C
(-) 1 bar...(+) 25 bar
(-) 40 °C...(+) 400 °C
Compled Insulated Probe
Conductive Tank
Partly Insulated Probe
Conductive Tank
Double Partly Insulated Probe
Insulating Tank
Partly Insulated Rope
Conductive Tank
1BSP
1 1/2BSP

10
ADHESIVE AND ACID / BASIC LIQUIDS
70 Hex
2 BSP
L
Aluminium
PFA
St.St.
L
70 Hex
Aluminium
St.St.
St.St.
PTFE
L
70 Hex
Delrin
Plastic
St.St.
St.St.
PFAPFA
2 BSP
Plastic
PTFE
70 Hex
PVC Coaxial
Plastic
PVC
70 Hex
PFAPFA
1 BSP
Plastic
PVDF
46 Hex
L
L
L
L= 250 mm. (Std) Max. 4 m.
(-) 1 bar...(+) 100 bar
(-) 40 °C...(+) 150 °C
L= 250 mm. (Std) Max. 6 m.
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
L= 250 mm. (Std) Max. 6 m.
(-) 1 bar...(+) 25 bar
(-) 20 °C...(+) 80 °C
L= 250 mm. (Std) Max. 1 m.
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
L= 250 mm. (Std) Max. 1 m.
(-) 1 bar...(+) 60 bar
(-) 40 °C...(+) 150 °C
L= 250 mm. (Std) Max. 1 m.
(-) 1 bar...(+) 6 bar
(-) 40 °C...(+) 60 °C
2 BSP
ECAS 408A
Double Probe (Single Fully Insulated)
Conductive / Insulating Tank
ECAS 408B
Double Partly Insulated Probe
Conductive / Insulating Tank
ECAS 408B
Double Partly Insulated Probe
Conductive / Insulating Tank
ECAS 408T
Double Partly Insulated Probe
Conductive / Insulating Tank
ECAS 408Tp
Double Partly Insulated
PVC Coaxial Probe
Conductive / Insulating Tank
ECAS 408Tm
Double Partly Insulated Probe
Conductive / Insulating Tank
2 BSP
Questo manuale è adatto per i seguenti modelli
17
Altri manuali ENSIM SENSORS Interruttore

ENSIM SENSORS
ENSIM SENSORS DX-ELS Manuale utente

ENSIM SENSORS
ENSIM SENSORS DX-ELB 11 Manuale utente

ENSIM SENSORS
ENSIM SENSORS ELP Manuale utente

ENSIM SENSORS
ENSIM SENSORS DX-ELS Series Manuale utente

ENSIM SENSORS
ENSIM SENSORS DX-ELF 101 Manuale utente

ENSIM SENSORS
ENSIM SENSORS EFS Manuale utente

ENSIM SENSORS
ENSIM SENSORS ISS 04 Manuale utente

ENSIM SENSORS
ENSIM SENSORS DX-ECAS Series Manuale utente

ENSIM SENSORS
ENSIM SENSORS DX-ELT Manuale utente

ENSIM SENSORS
ENSIM SENSORS ELM Series Manuale utente