FIP FLS M9.05 Manuale utente

1
FLS M9.05
CONDUCTIVITY MONITOR &
TRANSMITTER
SAFETY INSTRUCTIONS
PACKING LIST
General Statements
• Do not install and service the product without following the Instruction
Manual.
• This item is designed to be connected to other instruments which can be
hazardous if used improperly. Read and follow all associated instrument
manuals before using with it.
• Product installation and wiring connections should only be performed by
quali ed staff.
• Do not modify product construction.
Installation and Commissioning Statements
• Remove power to the instrument before wiring input and output connections.
• Do not exceed maximum speci cations using the instrument.
• To clean the unit, use only chemical compatible products.
Please verify that the product is complete and without any damage.
The following items must be included:
• M9.05 conductivity Monitor & Transmitter
• Instruction Manual for M9.05 conductivity Monitor & Transmitter

2
DESCRIPTION
CONNECTIONS TO INSTRUMENTS
The new FLS M9.05 is a powerful conductivity monitor studied to t a broad
range of applications included ultrapure water process. A 4” wide full graphic
display shows measured values clearly together with many other useful
information. Moreover, due to the multicolor bright backlight, measurement
status can be determined easily also from very long distance. A tutorial
software guarantees a mistake-proof and fast set up of every parameter.
Measured values can be showed as resistivity or TDS in according with
customer needs. A cell constant freely xable allows to use all types of 2-cell
conductivity probe. Two 4-20mA output grants to remote values of conductivity
and temperature to external devices. A proper combination of digital outputs
allows customized setups for any process to be controlled.
F3.00 F3.20 F6.30 F3.10 F3.05 F6.60 F6.61 F111
M9.05--------
ULF F3.80 pH/
ORP200
pH/
ORP400
pH/
ORP600
C150/
200
C100/
C300 C6.30
M9.05 - - - - - X X -
TECHNICAL DATA
General
• Associated sensors: FLS conductivity sensors and FLS temperature sensors
• Materials:
- case: ABS
- display window: PC
- panel & wall gasket: silicone rubber
- keypad: 5-button silicone rubber
• Display:
- LC full graphic display
- backlight version: 3-colours
- backlight activation: User adjustable with 5 levels of timing
- update rate: 1 second
- enclosure: IP65 front
• Conductivity input range: 0,055÷200000μS (Cell Costant : 0.01 - 10.0)
• Conductivity measurement accuracy: ± 2.0 % of reading value
• Temperature input range: -50÷150°C (-58÷302°F) (with Pt100-Pt1000)
• Temperature measurement resolution: 0,5°C/°F (Pt100-Pt1000)

3
Electrical
• Supply Voltage: 12 to 24 VDC ± 10% regulated
• 2 x Current output:
- 4-20 mA, isolated, fully adjustable and reversible
- max loop impedance: 800 Ω @ 24 VDC - 250 Ω @ 12 VDC
• 2 x Solid State Relay output:
- user selectable as ON-OFF, Proportional frequency, Proportional Pulse,
Timed Pulse, Off
- optically isolated, 50 mA MAX sink, 24 VDC MAX pull-up voltage
- max pulse/min: 300
- hysteresis: User selectable
• 2 x Relay output:
- user selectable as ON-OFF, Proportional frequency, Proportional Pulse,
Timed Pulse, Off
- mechanical SPDT contact
- expected mechanical life (min. operations): 107
- expected electrical life (min. operations): 105N.O./N.C.switching capacity
5A/240VAC
- max pulse/min: 60t
- hysteresis: User selectable
Environmental
• Operating temperature: -20 to +70°C (-4 to 158°F)
• Storage temperature: -30 to +80°C (-22 to 176°F)
• Relative humidity: 0 to 95% not condensing
Standards & Approvals
• Manufactured under ISO 9001
• Manufactured under ISO 14001
• CE
• RoHS Compliant
• GOST R

4
DIMENSIONS
Mechanical installation
The conductivity monitor & transmitter M9.05 is available just in one packaging
for panel or wall installation. The panel version is installed using the panel
mounting kit (M9.SN1), while the wall mounting version is got xing the panel
mounting version on the wall mounting kit (F9.KWX). The mounting kits can
be ordered directly connected to the monitor or separately and then simply
installed on it.
PANEL MOUNTING
WALL MOUNTING

5
Panel installation
Fix instrument on the panel rotating by hand the xing snails (M9.SN1).
Wall installation
Use the panel mounting kit (M9.SN1) to x the M9.05 on the dedicated frontal
cutout of the wall mounting kit (F9.KWX).
Tighten front screws of box and waterproof connectors of cables, internally
mount caps on screw sites to get a IP65 watertight installation.

6
WIRING
REAR TERMINAL VIEW
General recommendation
Always ensure the power supply is switched off before working on the device.
Make wiring connections according to wiring diagrams.
• Terminals accept 26 to 12 AWG (0.08 to 2.5 mm2)
• Strip around 10 mm (0.4”) of insulation from the wire tips and tin bare ends to
avoid fraying.
• Ferrules are suggested when connecting more than one wire to a single
terminal.
• Remove the upper part of the terminals for an easy cabling.
• Insert wire tip or ferrule completely into the terminal and x with the screw
until nger tight.
• Do not route the sensor, DC power, or 4-20mA cables in conduit containing
AC power wiring. Electrical noise may interfere with sensor signal.
• Routing the sensor cable in grounded metal conduit can help prevent
electrical noise and mechanical damage.
Wall Installation
Pull the electrical cables through liquid tight connectors.
Use electrical cables with the proper external diameter for the liquid tight
connector.
PG11/PG9: external diameter between 2-7 mm (0.079-0.276”)

7
POWER/LOOP WIRING DIAGRAM
Refer to dedicated sensor manual for its wiring. In general conductivity
sensor wires can be connected independently without caring about +IN/REF
connectors.
In case a temperature sensor (Pt100-Pt1000) is not available, place a brigde
connection between 28 - 29 and between 29 - 30.
Stand-alone application,
no current loop used
Connection to a PLC/Instrument with a separate power supply
or
Connection to a PLC with built-in
power supply
Power Supply
12 - 24 VDC
12 - 24 VDC
+ VDC
- VDC
2
1
Power Supply
12 - 24 VDC
4 - 20 mA
Input 2
PLC Terminals
4 - 20 mA
Input 1
+ VDC
- VDC
2
1
- LOOP 2
+ LOOP 1
+ LOOP 2
- LOOP 1
21
22
23
24
Power Supply
12 - 24 VDC
12 - 24 VDC
PLC
4 - 20 mA
Input 2
4 - 20 mA
Input 1
+ VDC
- VDC
2
1
- LOOP 2
+ LOOP 1
+ LOOP 2
- LOOP 1
21
22
23
24
Power Supply
12 - 24 VDC
12 - 24 VDC
PLC
4 - 20 mA
Input 2
4 - 20 mA
Input 1
+ VDC
- VDC
2
1
- LOOP 2
+ LOOP 1
+ LOOP 2
- LOOP 1
21
22
23
24
Power Supply
12 - 24 VDC
12 - 24 VDC
PLC
4 - 20 mA
Input 2
4 - 20 mA
Input 1
+ VDC
- VDC
2
1
- LOOP 2
+ LOOP 1
+ LOOP 2
- LOOP 1
21
22
23
24
Power Supply
12 - 24 VDC
12 - 24 VDC
PLC
4 - 20 mA
Input 2
4 - 20 mA
Input 1
+ VDC
- VDC
2
1
- LOOP 2
+ LOOP 1
+ LOOP 2
- LOOP 1
21
22
23
24
PROBE WIRING DIAGRAM
Conductivity probe connection
In general conductivity sensor wires can be connected independently without
caring about +IN/REF connectors.
25
26
+IN
REF Conductivity Sensor

8
Internal PLC
connection
N.O.
COM
4
3
PLC
Imax = 50 mA
Power sup.
Power sup.
O.C. IN
O.C. IN
SOLID-STATE RELAY WIRING DIAGRAM
(FOR SSR1 AND SSR2)
Connection to a PLC with NPN input Connection to a PLC with PNP input
N.O.
COM
4
3
PLC
Imax = 50 mA
Power sup.
Power sup.
O.C. IN
O.C. IN
Internal PLC
connection
Connection to a PLC / Instrument
digital input with separate Power
Supply
Connection to a PLC / Instrument
digital input for Voltage Free Contacts
(REED)
N.O.
COM
3
4
Power Supply
12 - 24 VDC
12 - 24 VDC
PLC / Instrument
Digital INPUT
Digital INPUT
Imax = 50mA
lmax = 50mA
lmax = 50mA
PLC
DIGITAL INPUT N
DIGITAL INPUT 2
DIGITAL INPUT 1
REF PLC
N.O.
Imax = 50 mA
Imax
COM
3
4
lmax = 50mA
lmax = 50mA
Pt100 - Pt1000
two wires connection
Pt100 - Pt1000
no connection
Pt100 - Pt1000
three wires connection
28
27
29
30
PT100/1000
27
28
29
30
PT100/1000
28
27
29
30
PT100/1000

9
SOLID-STATE RELAY WIRING DIAGRAM
(FOR SSR1 AND SSR2)
RELAY WIRING DIAGRAM
(FOR RELAY 1 & RELAY 2)
HOLD AND REED CONNECTION
The alarm is OFF during normal
operation and goes ON according to
Relay settings
The alarm is ON during normal
operation and goes OFF according to
Relay settings
The alarm is off during normal
operation and goes ON according to
Relay setting.
If Imax > 50 mA use external Relay
Connection to an User
NO
10
11
12
RELAY 2
NC
COM
Alarm
AC or DC
Power NO
10
11
12
RELAY 2
NC
COM
Alarm
AC or DC
Power
N.O.3
4 COM
AC or DC
Power
User
Imax = 50mA
N.O.3
4 COM
AC or DC
Power User
Imax = 50mA
lmax = 50mA
N.O.3
4 COM
AC or DC
Power
User
Imax = 50mA
N.O.3
4 COM
AC or DC
Power User Imax = 50mA
lmax = 50mA
N.O.
N.O.
N.C.
COM
4
3
COM
External Relay
V= 12-24 VAC/VDC
Imax = 50 mA
Imax
Imax
+V
-V lmax = 50mA
- HOLD
+ HOLD
18
17
19
20 - REED
+ REED
12-24
VDC

10
OPERATIONAL
OVERVIEW
The M9.05 conductivity monitor and
transmitter features a full graphic display
and a ve-button keypad for system
set-up, calibration and operation.
Full graphic display has a white
backlight during standard conditions,
a green backlight in case a external
device control is activated (ON/OFF,
PROPORTIONAL FREQUENCY,
PROPORTIONAL PULSE and TIMED
PULSE), a red backlight in case a set
alarm is activated (O.V.A. and O.T.A.,
always with priority).
The ve push buttons of the keypad
are used to navigate display levels and
modify settings.
The function of each button may change
according to display level; please refer
to following table:
VIEW LEVEL
MENU DIRECTORY
↓
↓
conductivity and temperature
item code -
software release
conductivity or temperature -
analog output 1
conductivity or temperature -
analog output 2
conductivity direct access to
calibration*
conductivity
conductivity - last calibration
Settings
Calibration
Outputs
Options
View data
↓
↓
*"Conductivity direct access to calibration" includes the "in-line adjustment" option to align on site the
measurement with a instant reference value.
Indice