Important information:
Please read the instructionsbefore operating the product.
Warning:
Tubing may crack due to wear. This results in the overflow of
fluid from tubing. People and instruments may be injured. You
must check tubing often and change tubing as needed.
Please connect power directly to a wall socket. Avoid using an
extension cord.
If power line or plug have wear or other damage, please unplug
it by holding the plug, not the line.
If the following situation occurs, please turn off the power
and unplug the unit, holding the plug, not the line.
1. Fluid splashes on the pump.
2. You think the pump needsto maintain or amend.
Your power socket must have a ground wire, and have reliable
grounding.

Catalog
PDS II Product Introduction..............................................................- 1 -
PDS II Appearance............................................................................- 2 -
PDS II Controller Interface fracture...................................................- 4 -
Filling Volume Reference Parameter (Media is Water) .................- 14 -
PDS II Technical Specifications.........................................................- 15 -
Product Size Chart:.......................................................................- 16 -
Main Functions and Features of PDS II...........................................- 17 -
External Control Interface Instruction of the Control Unit:.........- 18 -
Maintenance:...................................................................................- 19 -
Warranty and After Sales Service:...................................................- 20 -

-1 -
PDS II Product Introduction
PDS II(Programmable Dispensing
System)Series includes one controller and many
independent filling units. Each filling unit has two working
modes: Independent working mode and logic working mode.
In independent working mode, each filling unit can be
used as one independent peristaltic pump to achieve the
transferring or dispensing of liquid. The filling units work
independently. The tubing size and pump models of
each filling unit can be the same or different.
In logic working mode, many filling units are
operated according to the preset logical relation.The
tubing size and pump models of each filling unit can be the
same or different.
PDSII with 7‟‟ TFT color display touch
interface, intelligent calibration function,and online
micro-adjusting function offers various external controls
for your needs, supportsRS232 and RS485 (two
interfaces), standard MODBUS protocol, and accomplishes
external control in different industry conditions. Suitable
for transferring liquid with no pollution and high accuracy.
The accuracy can be 0.5%-1%.
PDS II

-2 -
PDS II Appearance
A B C
Composition:
A——Pump Heads
B——Filling Unit
C——Controller
Product Composition
PDSII consists of one controller and many independent
filling units, adding or deleting filling units as needed. You can
also set the validity of each filling unit as needed.
Below is a diagram of cascading many filling units:
①② ⑧
PDS II

-3 -
Cascading instruction:
1 - All the filling units (independently) can connect to the
controller. The controller has a separate interface for each
filling unit.
2 - Each controller can control 8 filling units max.
A
BC
Back side of filling unit
Interface Specification:A——Power switch
B——Power socket
C——Interface connecting the controller(G)
Interface Specification of controller
D
E
F
G
Back side of controller
D——Power switch
E——Power socket
F——External input interface
G——Interface connecting the filling unit(C)
PDS II

-4 -
PDS II Controller Interface Hierarchy
系统
主界面
单元
参数设置
有效
单元
常用
模式
全速
界面
时间
日期
选择要查
看的单元
流量校
正
步骤查看
步骤参数
查看
独立运
行常用
模式
逻辑运
行常用
模式
外控设置
界面
Main
Interface
Unit
parameter
Settings Effective
unit Common
mode Interface at
full speed External
control
interface
Review
unit See the
steps Date and
Time
Check the
parameter
s
Commonly
used logic
operation
mode
Run
independentl
y commonly
used mode
The flow
calibration
PDS II controller operation interface introduction
1 - Boot interface:After power on, the system will display
the welcome interface.
2 - Main Interface:
The diagram of main interface:
A - Speed display:The interface will loop display the speed of the
PDS II

-5 -
effective units every 5 seconds.
B - Dynamic display:Display the working state of each unit in
real time, dynamically display the monitoring result. It also has
an alarm function. If a bottle appears red (alarm signal), the
responding unit‟s bottle is absent. Please check the production
line.In the stop condition, you can click the bottle to
enter the parameter setting interface.
a - When the unit is on dispensing mode, you can click the bottle
to enter the parameter check interface, and fine-tune the flow
rate online.
b - When the unit is in transferring mode, you can click the bottle to
enter the parameter setting interface.In this condition,you
can only change the flow rate, speed,and direction.Other
parameters can‟t be modified.
c - If the unit is not the effective unit, it will highlight errors when
clicking the bottle.
C - Choose Looping units:Click to choose the unit displaying
in the “D” area.
D - Parameter display:This area will loop display the system settings
of the effective units every 5 seconds. Click the lock sign in
the upper right corner to lock the current unit and stop loop
display.
E - Date/Time display:Displays the Date/Time in real-time.
Click to set the parameters.
F - Settingsbutton: Click this button to enter the
settingsinterface.You can select the effective unit, common
mode, and the external control settings.
G - Full speed button:Click this button to enter the full speed
interface.
H - Start/stop selection area:This area can stop or start the
working unit.
3 - Settingsinterface
PDS II

-6 -
Common mode settings: Click this button to enter the
common mode interface.
Effective unit: Click this button to enter the effective unit
selection interface.
External control setting interface:Click this button to enter the
communication setup interface.
4 - Loop unit selecting interface
Select the unit
Please choose to view the unit
1234
678
5
ESC
In the main interface, you can click the “FCS” logo to enter
the loop unit selecting interface. After selecting the units, the
speed will display in “A“ area and system parameters in “D”
area will change to display the selected unit’s speed and
parameters.
PDS II

-7 -
5 - Effective unit interface.
The effective unit interface diagram is below:
Click the “effective unit” button in the main interface to
enter the effective unit selection interface. You can set the
validity of the unit here. When finished, you can click the
“OK” button to go back to main interface. If aunit is in
operation, the current working unit can‟t be modified.
Please note: Selecting effective unit is the first step to
set parameters.You can only operate on the effective units.
6 - Independent operating parameters settingsinterface
The diagram of the Independent operating parameters settings
are below:
PDS II
Indice
Manuali Attrezzature da laboratorio popolari di altre marche

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manuale utente

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manuale utente

NI
NI PXI-5422 Manuale

Collomix
Collomix Aqix Manuale utente

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manuale utente

Ocean Insight
Ocean Insight FLAME-NIR+ Manuale utente