Garner TS-1 Series Scheda tecnica

TS-1 Operation and Installation Manual
TS-1 Series
Continuous Degaussing System
Operation and Installation Manual
www.GarnerProducts.com
(800) 624-1903
10620 Industrial Ave. Suite 100
Roseville, CA 95678
Manufactured in the U.S.A
PN 4075-364E

TABLE OF CONTENTS
WARNINGS2
GENERALINFORMATION3
INSPECTION3
INTRODUCTION3
MAGNETICFIELDS3
MAGNETIC FIELDS MEASURED AROUND PERIMETER 3
PACE MAKER USERS: WARNING 3
SPECIFICATIONS4
INSTALLATION5
POWER REQUIREMENTS 5
LINE VOLTAGE SELECTION REFERENCE 5
INPUT VOLTAGE CONFIGURATION WARNING 6
POWER PLUG 7
FUSE BLOCK 7
LOCATION 7
COOLING 7
OPERATION7
CONTROLS AND INDICATORS OVERVIEW: 7
DEGAUSSING INSTRUCTIONS: 8
DEGAUSSING INSTRUCTIONS FOR DLT MEDIA: 8
DISPLAY MESSAGES: 9
ERROR CODES 10
Charge Error10
Drawer Open Error10
Gauss Error10
Overheat Error10
Change Batteries11
PARTSLISTMODELTS‐112
DECLARATIONOFCONFORMITY13
WARRANTY14

TS-1 Operation and Installation Manual Page 2
WARNINGS
WARNING HIGH VOLTAGE! NEVER Power up this equipment with covers or safety features
removed. Even when the equipment is powered down, high voltage (up to 1,000 Volts) can be
present inside the chassis.
WARNING! NEVER remove the ground pin from the power cord nor use a plug adaptor that
doesn’t provide grounding continuity.
WARNING! NEVER disable safety interlocks. Doing so can allow unpredictable operation that is
dangerous to the equipment operator and bystanders.
CAUTION! NEVER change the internal settings of this equipment. This may reduce the
equipment performance and/or cause damage to the equipment. In particular, incorrect settings
for power mains. The equipment warranty does not cover any problems caused by
unauthorized internal adjustment.
CAUTION! The equipment weighs approx. 66 kg (146 lb). Always have two people available to
carry, lift or turn over the equipment.
CAUTION! Be sure the table, desk, or bench this equipment will be operated or stored on is
stable and can support the weight of this equipment. If this equipment falls on someone, it can
cause serious injuries or death.
CAUTION! The line-voltage receptacle must be near the equipment and must be easily
accessible. Avoid extension cords.
NEVER let metal, wire or other objects protrude into the equipment. This may result in an
electric shock, electrocution or damage to the equipment.
NEVER block any cooling vents on the top, rear or bottom of the equipment.
NEVER expose the equipment to rain, snow or any liquids.
NEVER operate the equipment dusty or dirty environments. Doing so can cause damage to the
equipment and void the warranty.
NEVER install the equipment in a place without adequate ventilation. Heat dissipation may be
reduced, and the equipment may be damaged.
NEVER operate or touch the equipment with wet hands. This may result in an electric shock or
damage to the equipment.
AVOID using or storing the equipment in areas with temperatures below 0°C (+32°F) or above
+50°C (+122°F).
AVOID placing the equipment in excessively dusty environments or in direct sunlight.
AVOID placing the equipment against walls or putting anything on top of the equipment. Always
place unit in a secure place to avoid inadvertent use by unauthorized personnel.

TS-1 Operation and Installation Manual Page 3
GENERAL INFORMATION
INSPECTION
Inspect the degausser for shipping damage as soon as it is unpacked. If the degausser
or any accessories are damaged or fail to operate correctly, notify carrier and Garner
Products, Inc. immediately. For assistance from Garner Products, Inc. call (916) 784-
0200 or fax (916) 784-1425. For warranty information, please refer to the Warranty
section inside this manual.
INTRODUCTION
The Garner Products, Inc. Model TS-1 is a continuous duty capacitive discharge
degausser. The TS-1 is designed to erase hard drives and tape cartridges that fit inside
the drawer. The convenient drawer loading design accommodates many different types
and sizes of media, without the need for adapters.
MAGNETIC FIELDS
Magnetic fields measured around perimeter
The TS-1 produces a single focused magnetic pulse lasting less than 1/20 of a second.
The focused intensity allows the TS-1 to erase today’s highest coercivity media while
also limiting the magnetic field exposure to 1 second for every 20 pieces of media
degaussed.
Distance from
degausser Magnetic Field
(Front Panel) Magnetic Field
(Side) Duration of Field
0cm (0 inches) 1320 gauss 68 gauss < 0.05 Second
10cm (4 inches) 440 gauss 22 gauss < 0.05 Second
20cm (8 inches) 145 gauss 8 gauss < 0.05 Second
30cm (12 inches) 49 gauss 3 gauss < 0.05 Second
Note: The testing room measured 0.4 gauss before degausser was plugged in.
Average Operational Distance (AOD) from user to the TS-1 is 12-18 inches (30-45cm).
According to the American Conference of Governmental Industrial Hygienists (ACGIH)
Threshold Limit Values (2005) - Static Magnetic Fields, whole body exposure limits
recommended for an 8 hour shift are 600 gauss. Because of the short duration of each
pulse from the TS-1 (1/20th* of a second) and the extremely small magnetic field outside
of the unit (<49 gauss, AOD) the total 8 hour whole body exposure is 1/800th of the
recommended limit.
*Pulse duration was rounded to ½ second per discharge for calculations to cover any build up of
magnetic fields generated before or after discharge.
PACE MAKER USERS: WARNING
Due to stray magnetic fields, consult your physician before operating.

TS-1 Operation and Installation Manual Page 4
SPECIFICATIONS
Power Supply (TS-1): Selectable: 95~105, 105~115, 115~125, 215~225, 225~235 or
235~240. See “Input Voltage Configuration” section of this
manual.
Power Consumption
(TS-1): 100V: Under Standby 0.3A
While charging 5.0A
At the time of ERASE 0.6A
110V: Under Standby 0.25A
While charging 4.5A
At the time of ERASE 0.55A
120V: Under Standby 0.2A
While charging 4.0A
At the time of ERASE 0.5A
220V: Under Standby 0.17A
While charging 2.5A
At the time of ERASE 0.35A
230V: Under Standby 0.16A
While charging 2.25A
At the time of ERASE 0.325A
240V: Under Standby 0.15A
While charging 2.0A
At the time of ERASE 0.3A
Degausser System: Capacitive Discharge
Operating Time Duty Cycle: Continuous
Erasing time: < 15 milliseconds
Cycle time: 45 seconds per cycle.
Magnetic Field: 20,000 Oe measured in center.
Max. Media Size: 108mm (4.25 in.)(W) x 31.5mm (1.25 in.)(H) x 149mm (5.86 in.)(D)
Erasing Area Size: 108mm (4.25 in.)(W) x 31.5mm (1.25 in.)(H) x 149mm (5.86 in.)(D)
Media: 2.5 type HDD, 3.5 type HDD, DLT, LTO, 3490 and more
Temperature: 5°C - 40°C (41°F - 104°F)
Humidity: 10%H - 40%H (without condensation)
Weight:
TS-1: 146 lbs. (65.7 kg)
Shipping: 181 lbs. (83.6 kg)
Dimensions: TS-1: 19 in. H x 17.5 in. W x 21.5 in. L
(47.5cm H x 44cm W x 54cm L)
Shipping: 34 in. H x 26 in. W x 29 in.
(77.5cm x 62.5cm x 82.5cm)
Warranty: 1 Year Factory Warranty Standard, Optional 3 Year Factory
Warranty available.

TS-1 Operation and Installation Manual Page 5
INSTALLATION
All trouble shooting and wiring checks must be performed by a
qualified electrician or authorized plant personnel.
Power Requirements
It is important to ensure correct power is applied to the degausser. Unless otherwise
specified and noted on the power specification label on the back of the TS-1, the TS-1 is
factory configured for 120VAC. Check the power specifications label to ensure the
factory configured TS-1 input voltage is configured to match the voltage from the power
mains the TS-1 will plug into. Call Garner Products, Inc. if the voltage listed is not
compatible with your voltage or if you have any questions. International customers call
011 916 784-0200.
Line Voltage Selection Reference
The Line Voltage Selection Reference is a guide for configuring the voltage jumper to
the correct tap of the transformer. This allows the TS-1 to be used world wide under a
variety of voltages. For detailed instruction please see the “Input Voltage Configuration”
section of this manual.
All trouble shooting and wiring checks must be performed by a
qualified electrician or authorized plant personnel.

TS-1 Operation and Installation Manual Page 6
Input Voltage Configuration WARNING
The TS-1 model is equipped with a voltage selection block allowing the TS-1 to be
configured for use in any country.
To check or change the voltage configuration of the TS-1, follow the steps listed below:
All trouble shooting and wiring checks must be performed by a
qualified electrician or authorized plant personnel.
If you are installing the TS-1 for the first time, or if the TS-1 has moved to a new
location, confirm the input power setting of the TS-1 is configured to match the power at
your location**.
**If the power at your location is prone to voltage sag, the TS-1 may/will report a
“CHARGE ERROR” fault. This is an operational feature designed to ensure the
degausser will only operate when an optimal charge condition is present. If the
charge error fault persists when the power is known to be stable, call customer
support.
Changing the Voltage Configuration
Step 1 Un-plug the TS-1 from the power mains.
Step 2 Loosen and remove the Power Selection Panel
from the rear panel. (4) T-20 Torx Screws
Step 3 Move the (brown) power selection wire (located on
the right side of the fuse block) to the appropriate
tap/fuse on the right side of the fuse block. The
picture below shows the brown wire on the
120VAC connector.
Step 4 Replace the Power Selection Panel.

TS-1 Operation and Installation Manual Page 7
Power Plug
Plug the EIA connector attached to the power cord supplied into the receptacle on
the back of the TS-1.
Fuse Block
Located behind the removable panel on the front side of the TS-1, there are six
separate fuses, one each for 100, 110, 120, 220, 230 & 240 VAC power. Only one
fuse is used at a time. Therefore, for any voltage range, the two unused fuses of
the same value can be used as spares.
Location
The degausser must be placed on a flat, hard surface. Keep at least a 6 inch
(15cm) distance from any object that may interfere with the cooling fan exhaust
located on the backside of the degausser. As a general rule, media not intending
to be erased should be kept about 60cm (2 feet) away.
Cooling
Forced air-cooling is used to maintain safe operating temperatures within the
degausser. The air intake and exhaust are located on the back panel. To ensure
proper cooling, do not block the air-intake or exhaust and maintain at least a 15cm
(6 inches) distance from the back of the degausser to any obstacle.
OPERATION
Controls and Indicators Overview:
LCD Display The LCD display gives status, automatic count, total
count, software version, field strength, user
instructions and fault messages.
Media Drawer The media drawer holds the media to be degaussed
and activates and deactivates the degaussing cycle
when in the closed and open positions.
Control Panel The Control Panel consists of two active push buttons
and three indicator lamps.
POWER: The Power button controls power to the TS-1. When
in the ON position, the Power button illuminates
Yellow.
STATUS: The Status button illuminates Red when the unit is
charging and flashes during precharge.
ERASE: The Erase button initiates an erase cycle when
pressed. The Erase button illuminates Green when
the TS-1 is ready to erase media, blinks when the
drawer is open and flashes during precharge.

TS-1 Operation and Installation Manual Page 8
Degaussing Instructions:
Step 1 Press power button ON.
Step 2 Lift up on the drawer to unlock it and slide the drawer out
until it stops
Step 3 Place the media into the drawer and push the drawer to the
closed position. Orientation of the media is not important as
long as the drawer closes freely and completely.
Step 4 Press the “Erase” button.
Note: After approximately 45 seconds, the LCD will read
“Erase Successful Remove Media” indicating that the media
has been degaussed and is ready for removal from the
media drawer.
Step 5 Remove media by lifting up and sliding out the drawer.
Step 6 Insert another hard drive or turn the power off and close the
drawer.
Degaussing Instructions for DLT Media:
Remove the plastic protective cases. Simply follow the degaussing instructions above
for complete erasure.
Note: For DLT media that is to be re-used.
Although the data is completely degaussed in one operation, a second pass rotated at
90 degrees is required to smooth out any magnetic fluctuation. Note: If the second
pass is not completed, the tape drive may eject the tape from the tape drive and
indicate that the tape needs cleaning. The second pass at 90 degrees will correct this
condition.
If you have any questions, please call Garner Products, Inc. at 1-800-624-1903 or 916-
784-0200.
First Pass
Second Pass

TS-1 Operation and Installation Manual Page 9
Display Messages:
The LCD display gives status, automatic count, total count, software version, field
strength, user instructions and fault messages.
The following information appears on the LCD Display from power-up through one
automatic degaussing cycle:
Action LCD Display Description
Step 1:
Press POWER
button ON
GARNER PRODUCTS
0 −− 00
Total Erase Cycles: The left counter
indicates the total number of times the
right counter reached 50,000. The right
counter indicates the total number of
erase cycles on the TS-1 up to 49,999.
TS-1
Firmware Ver 1.2 Displays degausser model and
firmware version.
Ready
Session = 0
Indicates TS-1 is ready to have media
loaded into the drawer.
Session is the number of erase cycles
since the last power up.
Step 2:
Open Drawer
Insert Media
Close Drawer
Drawer Open
Indicates media drawer is open – this
is normal when inserting media to be
erased. Note: The READY light blinks
when drawer is open.
Ready
Session = 0 TS-1 is ready to perform an erase
cycle.
Step 3:
Press ERASE
Button
Precharge
The TS-1 is managing the current to
evenly charge the capacitors.
Charging 30
>>>>>>>>>>>>>>>>
Displays capacitor charging status as it
counts down from 60 seconds.
Although 60 seconds is the maximum
time allowed for charging, full charge is
normally completed between 25 and
35 seconds.
ERASE
Displays as the fully charged
capacitors are discharged into the
degaussing coil.
FS(Tesla)– 2.0 +
----------==|===
Displays the measured Field Strength
in Tesla during the erase cycle. The
bar graph easily identifies the field
strength. (1 Tesla = 10,000 gauss)
Erase Successful
Remove Media
Indicates the onboard diagnostics
confirmed the magnetic field was
sufficient to fully erase the media and it
is ok to remove the media.
Altri manuali per TS-1 Series
1
Indice
Manuali Attrezzature da laboratorio popolari di altre marche

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manuale utente

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manuale utente

NI
NI PXI-5422 Manuale

Collomix
Collomix Aqix Manuale utente

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manuale utente

Ocean Insight
Ocean Insight FLAME-NIR+ Manuale utente