1. Safety Precautions
1.1 Safety requirements
Centrifuge involveselements of risk all the time as it utilizes a high speed body of rotation. Follow precautions and all the safety
Follow precautions and all the safety requirements described on this user manualto prevent any damage and failureof equipment
and loss of lives.
1.Centrifugeshould beplacedon flatplace to maintainlevel.Bigvibrationis caused by operation of equipment in thecondition
of spindleinclined.
2.Check voltageto be used before centrifugeisconnected to power supply.
Wrong voltagecan damage equipment.
3.Userotors provided by Hanil Science Industrial Co Ltdand use parts and accessories recommended.
HanilScience Industrial Co Ltd is not responsible for damages of equipment and accidentscaused by using partsand accessories
not recommended.
4.Understand property of matter fullyand take needful safety actions in case you perform centrifugation of dangerous materials
such as pathogenic, toxicand radioactive substances.
5.Remove contaminantscompletely andtake needful actionssuch asventilationor isolationof centrifuge in case centrifugeis
contaminated withdangerous materialssuch as pathogenic,toxic and radioactive substances.
6.Do not perform centrifugationof materialscapable of producingvolatileand explosible steam.
7.Do not operate equipment more thanmaximumRPM allowedby rotor installed.
8.Do not decelerate or stop rotor rotatingwithhands.
9.Rotor should be used after balancing of symmetry between weightof samples to be used.
10.Rotor shouldbe fixedintoshaft and itslidshouldbe tightenedcompletely.
Otherwise, the lid canbe taken off during rotation.
11.A trained and qualifiedtechnician should perform repairing directionsnot mentioning on thisuser manual.
12.User shouldremove contaminantscompletely in advancein case user requests repairs toa qualifiedtechnician.
13.Rotor and chamber should be kept dry allthe time before use of centrifugeto make long perioduse of centrifugeand maintain
thebest operatingconditions.
14.Payattentionthat aluminumrotor has a chemical reaction and starts to becorrosive ifaluminumrotor contacts laboratory
cleaning liquid, cesium, silver, and (especially chloride ion) of strong acid, strongly alkalineor strong alkali.
15.According to IEC61010-2-020, 30cmof a safe distance around centrifuge should be assured during operation.
Nothing shouldbe puton thissafe distance in case it is destroyed.