JETT BTM-D1D Manuale utente

Table of Content 01
BTM-D1D Wireless Body Temperature
Monitor User Manual 02
Introduction of BTM-D1D – Wireless Body
Temperature Monitor 03
Operating Instructions 06
The Instructions of Indicators 11
Importance Precautions 13
Q & A 14
Technical Specications 16
111
Table of Content

BTM-D1D includes:
• ReceiverX 1
• Armband X 1
• AAA battery X 2
• 3 Volt lithium cell battery (CR2032) X 1
• Arm-belt X 1
• Assisted arm belt X 1
• Quick Guide Card X 1
• Manual X 1
• Warranty CardX 1
General Body Temperature Reference
Ear Temperature 35.8℃∼38.0℃
ArmpitTemperature 34.7℃∼37.3℃
Oral Temperature 35.5℃∼37.5℃
Anus Temperature 36.6℃∼38.0℃
BTM-D1D Wireless Body Temperature Monitor
User Manual
2

(1) The instruction of the keys and functions:
• Power key: To turn on/off the Receiver.
• Temperature setting key: To set up the Hi/Lo temperature alarm.
• Hi setting key: To increase the Hi temperature alarm setting and to switch
the℃/℉.
• Lo setting key: To decrease the Lo temperature alarm setting.
• Power indicator: To indicate the status of power.
Beeper
LCD Screen
LO Setting Key
HI Setting Key &
℃/℉Switch
Temperature
Setting
Power Key
Power Indicator
AAA batteres X 2
Introduction of BTM-D1D
Wireless Body Temperature Monitor
3

(2)The LCD screen of the Receiver illustration:
(3) The armband:
High / Low Temperature
Alarm
Temperature Indicator
Wireless Transmdission
Indicator
Armpit Temperature Mode
Receiver Battery Indicator
Armband Battery
Indicator
℃/ ℉
Power Key
Power Indicator
Temperature Sensor
CR2032 Battery
Front side Back side
Battery Installation:
1.Turn the counter-
clockwise by using a
coin or a screw-driver to
open the battery cover.
2. Insert the cell battery. 3. Turn the clockwise to
close the battery cover.
4

Note:
1. This product is made by FDA Grade Silicone. Clean the surface of armband
with wet tissue. Do not immerse the armband into water.
2. Recommend to use CR2032 battery manufactured within 2 month. If use
poor quality or years-old CR2032 battery, it may easy to become powerless
when power-on which means the battery can not supply power normally.
How to assemble the armband:
How to wear the armband:
1. Wear the armband on the left arm.
2. Adjust the belt at the proper tightness.
3. Move the temperature sensor under the
armpit.
4. Wear the assisted arm belt for fixing the
sensor under the armpit.
5
to attach
(1)
(2)
(3)
(4)

(1) How to use the armpit temperature function:
POWER ON
Step 1: Wear the armband on the left arm and adjust the belt at the proper
tightness. Move the temperature sensor under the armpit.
Step 2: Press the “Power key” on the armband for 2 seconds until the green
light flash and it will flash every 4 seconds to indicate the device is
working.
Step 3: Press the “Power key” on the Receiver for 2 seconds until it beeps
and the LCD screen display the armpit temperature which means the
wireless transmission is connected.
Step 4: Read the temperature on the LCD screen to check the change of the
patient’s temperature. It takes 3-5 minutes to have stable temperature
monitoring of armpit. The temperature sensor will update armpit
temperature every 12 seconds to be displayed on LCD screen.
POWER OFF F
Step 1: To turn the Receiver off, press and hold the “Power key” for 2 seconds
until it beep to turn off the LCD screen of the Receiver.
Step 2: Remove the armband from the patient’s arm and press “Power key”
on the armband until the red light flashes which means it is turned off.
Remark:
1. When power on, the green light only flashes once but not flashes every 4
seconds and disconnected, please change the new battery.
Operating Instructions
6

2. If outside of signal area and then result in disconnection, please remove
the receiver close to the armband within 3 min and then the signal will be
connected.
3. Any abnormal occurred when turn on the armband, please check and
change the battery.
4. Recommend turning on both Receiver and Armband power within 20
seconds for connection, otherwise it will automatically power off.
(2)How to set the temperature alarm:
Before using the body temperature monitor function, the operator may need
to set both high and low temperature alarm figures according to the patient’s
general armpit temperature status. When the temperature is out of the setting
range, the Receiver will start a series of "Beeps" alerting the caregiver to take
appropriate actions. The alarm setting procedures must be done when the
armband is on. Follow the steps below:
Step1: High temperature setting: Press the “Temperature setting key” for 2
seconds, the Receiver will beep twice to get into the temperature
alarm setting mode. The LCD screen flashes “HI” and the original
figures will be displayed.
Step2: Press “Temperature Setting Key” to adjust the alarm temperature
setting. The temperature standard is 0.5℃ a time such as 38.5℃to
39℃then to 39.5℃and so on, then press “Temperature setting key”
to complete the setting.
Step3: Low temperature setting: low temperature setting starts automatically
7

after finishing high temperature setting. Please press “Temperature
Setting Key” when “LO” is flashing, and then adjust the low
temperature setting. The temperature standard is 0.5℃ a time such
as 34℃to 34.5℃then 35℃then 35.5℃and so on and then press
“Temperature setting key” to complete the setting.
Remark:
‧ The range of setting: The default figures of the temperature alarm setting
are: Hi: 38℃(104℉) and Lo:34℃(96℉). The high temperature alarm
range from 36℃to 40℃(96℉-104℉). The low temperature alarm range
from 32℃-35.5℃(89℉-96℉).
‧Faulty-prevention mechanism: If the above steps setting are not
completed, the Receiver will return to the temperature monitoring mode
without changing any setting within 10 seconds.
‧Priority alarm processing: The product is designed to process the alarm
situation prior to the High/Low temperature alarm setting mode, or any
other operations. Without saving any setting if not pressing “Temperature
setting key”, the user is better off to operate the temperature alarm setting
with the full power supply.
‧Please re-set the Hi/Lo temperature alarm figures after replacing the
batteries. The Receiver will take the default as the temperature alarm
setting after battery replacement.
8

(3) Check the Hi/Lo temperature alarm setting:
(4) The Abnormal temperature alarm:
Alarm: The Hi/Lo temperature alarm setting on the LCD screen will be flashing
on the “Hi” or “Lo” if the reading of the patient is out of the temperature
alarm setting range. That is, the “Hi” and its figures are flashing. The
Receiver will beep every 2 seconds to inform the caregiver to take
appropriate action.
Anti-sleepyhead mechanism: The Receiver will check every 10 minutes the
temperature of the patient if it is still in the
temperature alarm setting. The Receiver will
The Hi/Lo setting
9

keep noticing the caregiver to take appropriate
action via the constant beeps if the patient’s
temperature is out of the temperature alarm
setting. Press any key on the Receiver to stop
the beep alarm.
Remark:
‧Please keep checking the patient’s temperature when alerting, even some
action to the patient had been done. For not triggering the alarm every
10 minutes, the caregiver can set the Hi/Lo temperature alarm a little bit
higher/lower, and then result the setting after the patient is getting better.
‧In case that the temperature, when power on, lowers the low temperature
alarm setting then starts the alarm system, this product will trigger the low
temperature alarm system after normal temperature higher than the low
temperature setting by 0.5℃.
‧It may cause low temperature alarm when the armband becomes loose.
Please check immediately the patient and adjust appropriately the arm-
belt.
10