MATOS Aria 150 Eco Manuale utente

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
INSTRUCTION MANUAL
MATOS ARIA Refrigerators
Applicable Models
MATOS Aria 150 Eco Solid Door
MATOS Aria 150 Eco Glass Door

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Contents
Before First Use ......................................................................................................................................................3
Positioning and safely using your MATOS Aria Medical refrigerator .....................................................................3
Transporting and moving your MATOS Aria Medical refrigerator .........................................................................3
Energy saving tips...................................................................................................................................................4
Product Information ....................................................................................................................................5
Important Information ...........................................................................................................................................5
Product Specifications............................................................................................................................................5
Product details .......................................................................................................................................................6
Controller overview and operation ........................................................................................................................7
Viewing & resetting minimum and maximum temperatures.................................................................................8
Changing the factory set point ...............................................................................................................................9
Changing the temperature scale between Celsius (°C) and Fahrenheit (°F) ..........................................................9
Changing the remote alarm contacts ...................................................................................................................10
Warning Alarms....................................................................................................................................................10
Remote alarm contacts ........................................................................................................................................11
Maintenance and Service of your MATOS Aria Medical Refrigerator .......................................................12
Defrosting.............................................................................................................................................................12
Cleaning................................................................................................................................................................12
Reversing the door ...............................................................................................................................................13
Replacing the light bulb (Glass Door Model ONLY) ..............................................................................................15
Storage .................................................................................................................................................................15
Electrical Connection............................................................................................................................................16
Disposal of your refrigerator ................................................................................................................................17
Troubleshooting Guide ..............................................................................................................................17
Alternative Manual Location......................................................................................................................18
Contact Details...........................................................................................................................................19
Sales & Marketing Enquiries:................................................................................................................................19
Spares and After Sales Service: ............................................................................................................................19
For any other information please visit: ................................................................................................................19

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Thank you for purchasing a MATOS Aria Refrigerator.
Before First Use
Before you first use this pharmacy Refrigerator please carry out the following actions:
•Check that the refrigerator has not been damaged in any way during transportation.
•If any damage is found it must be reported to our customer services department immediately
on 0844-815-3755
•Ensure all packaging has been removed, including cardboard, polystyrene and any tape used to
hold shelves in place for transportation.
•The refrigerator has been cleaned prior to despatch; however, we advise that it should be
cleaned using lukewarm water containing a mild detergent and a soft cloth.
•We recommend that the refrigerator should be left in an upright position for 24 hours prior to
switching on.
•This appliance is intended for professional use in applications such as pharmacies, laboratories,
medical facilities etc.
•This appliance is not intended for use by persons (including children) with reduced physical,
sensory or mental capabilities, or lack of experience and knowledge, unless they have been given
supervision or instruction concerning use of the appliance by a person responsible for their
safety.
•Children should be supervised to ensure that they do not play with the appliance.
Positioning and safely using your MATOS Aria Medical refrigerator
•Make sure the appliance is placed in a dry, well ventilated site, away from heat sources.
•The appliance must be placed on a level surface. If necessary, adjust the feet on the product so
the appliance remains level during operation.
•The appliance should have sufficient clearance around it to provide adequate ventilation –10cm
at each side, 6cm at the rear and 2.5cm at the top.
•Before ANY maintenance activity, pull the plug out of the socket by gripping the plug, DO NOT
pull on the plug cord.
•DO NOT use double-wire extension cords. If an extension cord is necessary make sure it is a cord,
which possess a protection safety certificate.
•If the main power supply cord is damaged, it should be replaced by a qualified service engineer.
•DO NOT attempt any repairs to the appliance yourself (apart from those highlighted in the
Maintenance and Service Section). Repairs carried out by someone without the relevant training
are putting their personal safety at risk and will invalidate your warranty.
•Do not store explosive substances such as aerosol cans with a flammable propellant in this
appliance
•WARNING: this appliance is only to be operated by authorised personnel or personnel who have
correct knowledge on how the appliance operates.
Transporting and moving your MATOS Aria Medical refrigerator
The refrigerator must always be moved in the vertical position. The cabinet must not be tilted anymore
than 40°. In the event that the cabinet is tilted more than 40°, the power supply must not be connected
until the appliance has stood upright for at least 24 hours.

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Energy saving tips
This appliance is cooled by energy-efficient R600a refrigerant. To keep running costs as low as possible
always:
•Position the appliance away from heat sources.
•Make sure the air can circulate freely around the appliance, don’t block ventilation grid.
•Ensure that products being stored in the appliance are below room temperature upon entry.
•Make sure the door is opened as little as possible whilst in use and closed as quickly as
possible to prevent unnecessary temperature fluctuations.

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Product Information
Important Information
The refrigerator contains environment-friendly, non-ozone depleting refrigerant R600a. As R600a is a
flammable gas, it is important to avoid damage to the refrigeration circuit during transport and
installation. If the refrigeration circuit is damaged, avoid using a naked flame in the vicinity of the
refrigerator and connecting power to the refrigerator. Also make sure there is good ventilation in the
room. If you are in doubt please contact your supplier.
WARNING: Keep ventilation openings, in the appliance enclosure or in the built-in structure, clear of
obstruction.
WARNING: Do not use mechanical devices or other means to accelerate the defrosting process, other
than those recommended by the manufacturer.
WARNING: Do not damage the refrigerant circuit
WARNING: Do not use electrical appliances inside the food storage compartments of the appliance,
unless they are of the type recommended by the manufacturer.
Product Specifications
This refrigerator has been designed and built with the intention of the product being used within a
Pharmacy and is designed to store contents between +2°C to +8°C.
Please see below for specific product information:

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Product details
3
4
5
6
1. Lock
2. Electronic Controller
3. Antimicrobial handle
4. Wire Shelf
5. Wire Basket
6. Adjustable foot (2 at front of appliance)
There are also 2 roller castors at the rear of the appliance to ease product movement.
Additional or replacement parts are available from our spare parts department.
See back of booklet for contact details.
1
2

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Controller overview and operation
1. Menu
2. Lock keypad
3. Light (not available on all models)
4. Maximum temperature / Up
5. Power
6. Mute
7. Minimum temperature / Down
8. Set / Enter
1
2
3
4
5
6
8
7

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Viewing & resetting minimum and maximum temperatures

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Changing the factory set point
The set point is the point at which the refrigerator normally holds the temperature at; in some
circumstances such as high/low room temperatures it may be necessary to adjust this slightly.
Changing the temperature scale between Celsius (°C) and Fahrenheit (°F)

Rollex Medical –MATOS ARIA Eco 150L User Manual v.02 24/04/2018
Changing the remote alarm contacts
Warning Alarms
This refrigerator has been designed to provide both an audible and visual alarm should any of the
following situations occur:
•Door is left open for a set period
•Internal air temperature goes outside the set parameters, either too high or too low
•Internal load temperature goes outside the set parameters, either too high or too low
Should one of these alarms occur this can be muted by pressing on the controller.
The load temperature is a simulated temperature which is intended to provide an indication of the
temperature of the refrigerator’s contents (e.g. a 5ml vaccine). The temperature of a refrigerator’s
contents will change more slowly (up or down) than the air temperature within that refrigerator should
there be a fault or the door is left open.