MECHLINE HYGENIKX+ Istruzioni per l'installazione

INSTALLATION & OPERATING MANUAL REVISION: 09/2020 | page 1 of 16
CONTENTS:
Chapter Description Page
1. Introduction .......................................................... 1
2. Ratings and Specifications .................................... 2
3. Product Safety ....................................................... 3
4. Structure ............................................................... 5
5. Location Advice ..................................................... 5
6. Operating Instructions ........................................
10
7. BioCeptor Unit Sizing Guide ................................ 15
8. Model Information ............................................. 11
9. Troubleshooting .................................................. 11
10. On-Going Machine Requirements ..................... 15
11. After Sales and Service Support ......................... 15
Chapter 1. INTRODUCTION
Thank you for purchasing Mechline’s HyGenikxTM fan assisted (AF) air & surface steriliser system. When correctly
installed, and operated in accordance with this Operator Manual, it will provide many years of service.
HyGenikx is an advanced and compact, wall or ceiling mounted steriliser system, proven to eradicate viruses and
bacteria throughout large occupied areas where air movements are minimal.
Utilising a combination of the most effective air and surface sterilisation technologies available - Germicidal Oxidation,
Germicidal Irradiation, Dual waveband Plasma Quatro - HyGenikx provides round-the-clock hygiene and safety
protection, controls odours and has also significantly been proven to prolong the life of fresh perishable foods.
Mechline’s HyGenikx system has been through rigorous testing, providing customers with the utmost confidence in its
effectiveness and quality.
LEAVE THIS MANUAL WITH THE OPERATOR AFTER INSTALLATION
INSTALLATION INSTRUCTIONS
6. Unit ........................................................................ 6
7. Lamp Replacement ............................................... 8
INSTALLATION &
OPERATING MANUAL

Air & Surface Steriliser System
REVISION: 09/2020 | page 2 of 16 INSTALLATION & OPERATING MANUAL
Chapter 2. RATINGS and SPECIFICATIONS
Region UK EUROPE
Model numbers
HGX- AF-25 / HGX- AF-50
HGX- AF-75 / HGX- AF-100
HGX- AF-25-E / HGX-AF-50-E
HGX- AF-75-E / HGX-AF-100-E
Air purifying technology Nano-titanium dioxide photocatalytic oxidation + UV-C
UV-C lamp tube 1 x 18w PLL composite tube
UV-C intensity inside chamber 15000 µW/cm2
UV-C lamp tube life Approx. 8000 hours
Air residence time inside chamber > 0.5 seconds
Number of photocatalytic
baseplates 3 plates
Noise level @ 1 metre <39dB
Air flow rate 20 m3/hr
Suggested coverage 25 - 30 m2
Installation Horizontal or Vertical Wall or Ceiling mount
Housing material All Aluminium Alloy Extrusion Structure
Colour Silver Aluminium
UV-C lamp tube model 9.5/38/50/70 mg
ELECTRICAL SPECIFICATIONS
Power consumption 33W
Voltage 220-230V
Frequency 50/60Hz
Input supply requirements To be connected to suitably isolated plug socket or switched fused spur
(with dedicated Earth) using Cloverleaf power lead supplied
Cordset
3 Core .75mm2 polarity specific
(L+N+E), 1.8m total length
BS1363/A moulded 3 pin plug with
earth enclosure
3 Core .75mm2 polarity specific
(L+N+E), 1.8m total length
BS1363/A moulded Schuko 2 pin
plug with earth enclosure
Input protection 3A Fuse
DIMENSIONS & WEIGHTS
Dimensions 430mm(L) X 130mm(W) X 100mm(D)
Weight 2.05 kg

Air & Surface Steriliser System
INSTALLATION & OPERATING MANUAL REVISION: 09/2020 | page 3 of 16
Chapter 3. PRODUCT SAFETY
3.1 GENERAL ADVICE
Electrical – HyGenikx AF models require an electrical connection via an existing isolated plug socket or switched fused spur
(with dedicated Earth). Power must always be disconnected prior to carrying out any work. If you are unsure how to
disconnect the power, or if there is any damage to sockets, cables, connections, or the unit itself please stop work
immediately and request advice from Mechline or a qualified electrician.
Ultraviolet light – WARNING! The HyGenikx unit should never be turned on if the cover has been removed. Be aware
direct exposure to Ultraviolet light can be harmful to skin and eyes causing irritation or even temporary blindness. If there
is any damage to the unit please disconnect power and contact Mechline.
Ozone – Whilst HyGenikx is not an ozone generator, a very small volume of ozone is released as a by-product by the
combination of technologies used. At the low levels produced there is no potential health risk during correct operation.
Our HGX-W-15-R/HGX-T-15-R refuse model is our only unit designed for use in unoccupied areas. At higher levels ozone
can cause irritation to throat, cough, headaches or shortness of breath, these symptoms should only be short term and be
resolved when the individual is removed to fresh air. If there are any concerns about the ozone produced please disconnect
power to the unit and contact Mechline. More about ozone on page 14.
Injury – HyGenikx units should always be securely mounted in a safe position, ensuring cables are not a potential trip
hazard and that every unit can be easily reached for servicing, with any relevant electrical outlets easily accessible. In the
event of a broken lamp, care must be taken to avoid cuts to hands (the complete unit can be returned to Mechline for this
work to be carried out if required) and appropriate protection should also be worn.
3.2 SYMBOLS
MEANING/DESCRIPTION SYMBOL
Dangerous Voltage
To indicate hazards arising from dangerous voltages.
Protective Earth (Ground)
To identify any terminal which is intended for connection to an
external conductor for protection against electric shock in case of
a fault or the terminal of a protective earth (ground) electrode.
Warning/Caution
An appropriate safety instruction should be followed or caution
to a potential hazard exists.
Refer to Manual
Refer to the relevant instructions detailed within the product
manual.
Wear Gloves
To indicate that appropriate gloves should be worn.
Before operating the unit, read these instructions carefully. Correct installation and operation will provide optimum
performance and a longer life for this high-quality air & surface steriliser system.

Air & Surface Steriliser System
REVISION: 09/2020 | page 4 of 16 INSTALLATION & OPERATING MANUAL
3.2 SAFETY INSTRUCTIONS
3.3 PRODUCT CONFORMITY
3.3.1 Mechline’s HyGenikx AF air & surface steriliser system conforms to the essential safety requirements of the
Low Voltage Directive (LVD) 2014/35/EU, Electromagnetic Compatibility Directive (EMC) 2014/30/EU
and is manufactured in accordance with following standards: EN55014-1:2017, EN61000-3-2:2014,
EN610000-3-3:2013, EN55014-2:2015, EN60335-2-65:2003+A1:2008+A11:2012, EN60335-
1:2012+A11:2014.
Mechline’s HyGenikx AF air & surface steriliser system is supplied to the marketplace having been
produced under the control of our comprehensive BSI ISO 9001:2015 Quality Management System
(certificate FS 616985).
3.2.1 Ensure the unit is installed, operated and maintained by trained and authorised personnel.
3.2.2 Children should not be allowed to play with the unit.
3.2.3 The unit should be used as provided and not be altered or tampered with as the safety of the unit may
be compromised.
3.2.4 Only connect the unit to a single-phase socket or a suitably fused and isolated spur with mains voltage, as
specified on the “Ratings and Specifications” section.
3.2.5 Before attempting servicing or maintenance tasks, the unit MUST be isolated from the power source
and disconnected.
3.2.6 Should a fault occur with the unit, immediately isolate and disconnect from the power source.
3.2.7 Ensure safe storage/positioning of the electrical cabling when not in use. Should the electrical cable
become damaged at any time, do NOT operate the unit.
3.2.8 Do not use the unit outdoors where it can be exposed to moisture.
3.2.9 Do not handle the plug or the unit with wet hands.
3.2.10 Do not operate the unit near water. If the unit comes into contact with liquid, isolate the power
immediately.
3.2.11 Any modifications to the electrical supply or connections to the unit must only be carried out by a
qualified electrician in accordance with the appropriate Codes of Practice.
3.2.12 The mains socket or fused outlet must be installed in an accessible location close to the unit to enable
disconnection.
3.2.13 Mechline recommends periodic inspection and verification of the integrity of the electrical connections in
accordance with recommendations of BS 7671, IEE Wiring Regs.
3.2.14 Keep foreign objects away from the unit’s air inlet and outlet areas – this may cause damage to the unit or
electric shock. Turn off the unit if there is any damage or if the unit is making excessive noise.
3.2.15 In public areas we would recommend the unit be hard wired into a fused spur.
3.2.16 Keep the unit and its power cable away from hot surfaces.
3.2.17 Do not use the unit where combustible gases or vapours may be present.
3.2.18 Use only original Mechline UV lamps.
3.2.19 Do not touch the glass portion of the lamp or the aluminium air diffuser plates with bare hands, oil from
skin can damage these parts. Always use protective gloves or a cloth when handling.
3.2.20 Do not look directly at the UV light. Exposure to UV light could burn the skin and can cause eye damage.
Reflected blue light showing through the outlet, inlet or edge of unit openings and front window is safe
to look at.
3.2.21 Be aware that there may be an initial odour caused by electrical components, this will clear after the first
48 hours of use.

Air & Surface Steriliser System
INSTALLATION & OPERATING MANUAL REVISION: 09/2020 | page 5 of 16
Chapter 4. STRUCTURE
Chapter 5. LOCATION ADVICE
We recommend that the unit is wall mounted, vertically installed, with the logo at the top of the unit and the cable at
the bottom. The bottom of the unit should be at a minimum height of 1.8m AFFL. Alternatively, the unit can be wall
mounted horizontally, ceiling mounted or placed on a high, flat, stable surface using the rubber feet supplied.
• Do not install the unit in a position where the air inlet and outlet could be blocked. This will affect
its efficiency.
• Do not install the unit in a position where a person can be in close proximity for extended periods of time.
• Avoid installing in close proximity to mechanical extraction and open windows.
• For optimum results keep the unit located in the same position and operational 24 hrs a day/7 days a week.
HYGENIKX UNIT
LAMP
A
Power cable socket
I
UV lamp
B
On/Off switch
J
Quartz part of lamp
C
Air inlet vent
K
End terminal
D
Cover
E
Air outlet vent
F
Aluminium air diffuser plates
G
UV lamp base
H
Lamp holding plate
A
B
C
D
E
F
G
H
I
J
K
NOTE:
The cover needs
to be fitted so the
HyGenikx logo is
at the air outlet
vent end (E).

Air & Surface Steriliser System
REVISION: 09/2020 | page 6 of 16 INSTALLATION & OPERATING MANUAL
Chapter 6. INSTALLATION INSTRUCTIONS
6.1 MOUNTING INSTALLATION STEPS
6.1.1 Equipment needed:
6.1.2 The following installation steps are for mounting the HyGenikx unit vertically on a wall (recommended).
Alternative mounting options include wall mounting horizontally or ceiling mounting. If ceiling mounting the
unit this will require four brackets for secure installation. The mounting brackets can be adjusted along the
base plate for secure fixings. Please use correct fixings for wall or ceiling mounting.
NOTE: Air flows out the opposite end to the power cable socket. The air flow must not be restricted.
6.1.3 It is important to first identify the best location for the unit. Ensure you have read Chapter 5 Location Advice.
6.1.4 Wall mounting vertically requires two brackets. Unscrew the rubber feet and mount two of the supplied
mounting brackets at the top of the unit (the opposite end to the power cable socket) as shown below.
x 2
x 2

Air & Surface Steriliser System
INSTALLATION & OPERATING MANUAL REVISION: 09/2020 | page 7 of 16
6.1.5 Measure 2.23 metres AFFL. Then hold the unit in its location and mark the two drill holes through the
mounting bracket holes.
6.1.6 Carefully place the unit in a safe place and drill the two mounting holes.
6.1.7 Insert wall plugs into drilled holes. Use hammer to fit wall plugs into position. Insert screws provided into
wall plugs and screw into position, leaving 5mm to hang the HyGenikx unit.
6.1.8 Align mounting bracket holes with screws and hang the HyGenikx unit into position as shown.
6.1.9 Before the unit is ready to use, the power cable will need to be connected to a suitably isolated socket
or fused spur. Turn the unit ON via the switch located at the bottom of the unit.
2.23m
MARKED MOUNTING
HOLES
MARKED MOUNTING
HOLES
5mm
1
1
1
2
2
2

Air & Surface Steriliser System
REVISION: 09/2020 | page 8 of 16 INSTALLATION & OPERATING MANUAL
Chapter 7. UV-C LAMP REPLACEMENT
7.1 IMPORTANT ADVICE
• Do not touch the anodized aluminium air diffuser plates.
• Always use disposable gloves (or a tissue) when handling the UV lamp.
• Any dirt on the lamp can be cleaned with a dry cloth or soft brush.
• After approximately 8000 hours or 12 months of continuous burning, the
UV-C lamp will lose 20% to 30% of its power and must be replaced.
• The lamp must be replaced every 12 months to ensure optimum performance of the system.
7.2 LAMP REPLACEMENT
7.2.1 Turn OFF using the switch
on the unit. Disconnect
from electricity and remove from placement.
7.2.2 Gently lay the unit flat on its back, unscrew the four cover screws
and remove the cover as shown below.
WARNING: DO NOT PLUG THE UNIT INTO THE MAINS
WITHOUT THE COVER IN PLACE. Direct exposure to the
ultraviolet light can cause eye and skin irritation.
2
1
On/Off Switch

Air & Surface Steriliser System
INSTALLATION & OPERATING MANUAL REVISION: 09/2020 | page 9 of 16
7.2.3 Unscrew and remove the two finger screws from the lamp holding plate – step 1 below.
7.2.4 Carefully slide OLD LAMP from its socket and lift out (always use gloves or a tissue) – steps 2 and 3 below.
7.2.5 Slide the OLD LAMP out from the holding plate – step 4 below.
7.2.6 With a soft brush remove any dust and debris
from the unit – step 5 below.
7.2.7 Take the NEW LAMP and slide it into the holding
plate (always use gloves or a tissue) – step 1 below.
7.2.8 Fit the new lamp into the unit – following steps 2-5.
2
1
3
4
1
5
2

Air & Surface Steriliser System
REVISION: 09/2020 | page 10 of 16 INSTALLATION & OPERATING MANUAL
7.2.9 Fit the cover (with the HyGenikx logo at the air outlet vent end), replace and tighten screws – steps 6 and 7 below.
7.2.10 Place unit back in its placement. Reconnect the power, and turn ON using the switch
located on the unit – step 8 below.
7.3 SERVICING INFORMATION
7.3.1 Inspection of area:
• If you are carrying out a service check on a HyGenikx unit some time after installation please ensure the
area is safe and secure to work in.
Appropriate signage should be visible to warn public that work is being carried out.
• Are there any obvious bad odour sources?
The HyGenikx unit will help to control odours but it is important that bad odour sources on site continue to be
addressed. Overflowing bins, unflushed toilets, leaks etc. should all be rectified. Refer to Section 9.2 for some
useful check points.
7.3.2 Inspection of each unit:
• Is the lamp light visible through the vents of unit?
If not please refer to Section 9.3 Lamp/Unit Failure Checklist.
• Is there an odour present in the environment that you think is coming from the unit?
After installation it is not unusual for the unit to produce a slightly metallic odour within the first 48 hours of use
due to the lamp bedding in. If the unit is working correctly but odour remains after this time please refer to
Section 9.1.3.
• Is the unit operational 24/7?
It is important to confirm with the site that the unit/power to the area is not disconnected at any point (with
units connected to lighting circuits – check turning lights off does not disconnect unit).
For technical assistance e-mail [email protected], quoting the HyGenikx model number. The model number can be
found on both sides of the unit.
3
4
5
8
7
On/Off Switch
6
Air outlet vent
Altri manuali per HYGENIKX+
1
Indice