Page 5
General
capaNCDT 6110 / 6112 / 6120
Proper Environment
Temperature range controller Operation Storage
DT61xx +10 ... +60 °C (+50 ... +140 °F) -10 ... +75 °C (+14 ... +167 °F)
DT61x0/IP -20 ... +60 °C (-4 ... +140 °F) -20 ... +75 °C (-4 ... +167 °F)
Temperature range sensor Operation Storage
DT61xx -50 ... +200 °C (-58 ... +392 °F)
DT61x0/IP type CSE
type CSF -40 ... +100 °C (-40 ... +212 °F)
Temperature range sensor cable Operation Storage
Permanent Up to 10,000 h
DT61xx
CCx, CCx/90 -100 ... +200 °C
(-148 ... +392 °F)
-50 ... +200 °C
(-58 ... +392 °F)
CCmx, CCmx/90
CCgx, CCgx/90 -20 ... +80 °C
(-4 ... +176 °F)
-20 ... +100 °C
(-4 ... +212 °F)
-50 ... +80 °C
(-58 ... +176 °F)
DT61x0/IP
- Protection class
Sensors IP 54 (DT61xx, when plugged in)
IP 40, IP 54 (DT61x0/IP)
Controller IP 40 (DT61xx)
IP 68 (DT61x0/IP)
- Humidity 5 - 95 % (non-condensing)
- Ambient pressure Atmospheric pressure
- The space between the sensor surface and the target must have an unvarying dielectric constant.
- The space between the sensor surface and the target may not be contaminated (e. g. water, rubbed-off parts, dust,