Table of contents
page
1. Calibration 3
1.1 Zero calibration 3
1.2 Corner calibration 4
2.3 Weight calibration 8
2. Parameter settings 12
2.1 To backup initial parameter settings + calibration 14
2.2 Check the (last) stored calibration memory 14
2.3 To change US setting into EU or EU settings into US, recover calibration settings 14
2.4 To recover initial parameter settings + calibration 14
3. Indicator startup routine - software version 2.18 15
4. Indicator power supply from fork lift truck battery - connection 15
Warning:
Only trained and authorized personnel are allowed to calibrate the scale.
NOTE !!!
Before starting the calibration procedure be sure the jumper JP1 is
connected.
Version 2014 and earlier Version 2015
Deactivate Correction sensor(Tilt sensor) parameter P40=no
After Calibration procedure activate correction sensor P40= CS_FA
Remove JP1 for calibration lock.