Seiler Microlux IV Manuale utente

Microlux IV
LED
Operations Manual

3433 Tree Court Industrial Blvd.
St. Louis, MO 63122
(800) 489-2282
(314) 968-2282
Email: [email protected]
www.seilerinst.com

Contents
1. Parts Name 1
2. Specication 2
3. Installation 3
4. Operation 4
5. Installation & Operation for Attachments 5
6. Maintenance 5
7. Troubleshooting 10
8.Outts11

A3. Eyepiece
B2. Diopter
adjuster
A9. Notch for
attachment
A7. Clamp
B3. Aperture
diaphragm cover
B4. Condenser
Adjusting screw
(left & right)
A8. Condenser
A10. Filter
B5. Field diaphragm
revolver
B6. Condenser
thumb screw
B7. B8. Stage moving knobs
B9. Potentiometer
A11. Conden
set holder
A6. Mechanical
stage
A5. Objective
A1. Body
A4. Nosepiece
B1. Head thumb screw
A3. Eyepiece
1. Parts Name
Fig. 1-1
1

A12. Handle
B10. Condenser
focusing screw
A13. Power Input
A14. Fuse
B11. Power Switch
B12. Tension
adjustment ring
B13. Coarse
focusing knobs
B14. Fine
focusing knobs
Fig. 1-2
A1. Body A2. Seidentopf binocular head A3. Eyepiece A4. Nosepiece A5. Objective A6. Mechanical
Stage A7. Clamp A8. Condenser A9. Notch for attachment A10. Filter A11. Condenser holder
A12. Handle A13. Power Input A14. Fuse A15. Lamp Plate
B1. Head thumb screw B2. Diopter adjuster B3. Aperture diaphragm lever B4. Condenser adjusting screw
B5. Field diaphragm revolver B6. Condenser thumb screw B7. B8. Stage moving knobs B9. Potentiometer
B10. Condenser focusing screw B11. Power switch B12. Tension adjustment ring B13. Coarse focusing
knobs B14. Fine focusing knobs B15. Lamp plate thumb screw
2. Specication
2.1 Total magnications
10X
16X
4X
40X
64X
10X
100X
160X
20X
200X
320X
40X
400X
640X
100X
1000X
1600X
Objectives
Eyepiece
2

2.3 The other specication
2.3.1 Mechanical tube length: 160mm
2.3.2 Conjugate distance: Finite 195mm
Innite
2.3.3 Head: Seidentopf binocular (trinocular) 30º
Interpupillary adjustable distance is 50-75mm
Diopter adjustable range ±5,
Anti-fungal systems
2.3.4 Nosepiece: Quadplex or quinplex nosepiece
2.3.5 Mechanical Stage: Size 190mm X 140mm
X-Y travel 55mm X 78mm
2.3.6 Focusing systems: Coaxial coarse and ne focusing knobs
Coarse stroke 25mm, ne division 2 µ m
Condenser up-down range 25mm
2.3.7 Condenser: Abbe condenser, N.A. 1.25, Adjustable aperture
Aperture center is ajustable
2.3.8 Illumination: Koehler illumination system
2.3.9 Filter: Blue lter is for standard outts, green and amber is optional
2.3.10 Electric components: Input voltage AC 100-265V, 50/60Hz
Output voltage DC1.3.4V
3W LED lamp
Rotation potentiometer
Fuse 2A 5 X 20
2.2 Objectives (with eyepiece 10x)
3

Eyepiece
Binocular Head
1 X CCD Adaptor
Trinocular Head
Notch
obje
Condenser
N.A. 1.25
Filter
Fig. 2
3. Installation
Please install the microscope as follows:
Installation Process: (see Figure 3 on page 5)
1) Installing Lamp: LED is factory installed.
2) Installing binocular head
Loosen B1 binocular head thumb screw, put out A2 head, put it on top of body as in Fig. 1,
tighten the screw.
4

A15 Lamp plate
B15 Lamp plate
screw
Fig. 3
3) Installing objectives
Take out the objective from the packing box, and drive them into the holes of nosepiece orderly and
tightly according to times.
4) Installing eyepiece
Take out the eyepiece tube cover, then get the eyepiece from the packing bag, and insert into the tubes.
5) Installing condenser
a. Turn B13 coarse focusing knob, rise A6 mechanical stage to top;
b. Turn B10 condenser focusing screw, and take A11 condenser holder to the lowest position;
c. Loosen B6 condenser thumb screw, put the condenser to the holder, tighten B6 condenser thumb
screw.
d. Turn B10 condenser focusing screw, and raise the condenser to working position.
6) Installing blue lter or the other attachment
If blue lter will be used, the installation should be done before process 5, or do the installing during
process 5-C, then put A10 blue lter or the other attachment to the bottom hole of A8 condenser.
7) Installing the module attachment
The microscope can be with A9 module attachment and is multifunctional. Analyzer and wavelength
plate can be installed in A9 notch. Polarizer is installed over eld diaphragm or below the condenser.
Filter can be installed in the notch if the other attachment is in the bottom hole of condenser.
8) Installing CCD adaptor
Loosen the thumb screw in trinocular tube, put the CCD adaptor into trinocular tube, then tighten the
screw.
5

4. Operation
4.1 Operation process instructions
1) Turn power switch, adjust light
2) Place the sample
3) Put 10X objective into light
path and focus the sample
4) Adjust interpupillary distance
and diopter
5) Adjust condenser aperture
diaphragm
6) Adjust and center eld diaphragm
Put the objective into light path
and focus
Observe
Turn on B11 power switch and turn B9 pentiometer
Put the sample on A6 stage (open the clamp, and
put the sample on it), turn B7, B8 stage knobs to
put the sample into light path
Turn A4 nosepiece, put 10x objective into light path,
turn B13, B14 coarse and ne focus knobs.
Adjust A2 head to make two elds into one;
adjust B2 diopter adjuster to make diopter
for both eyes the same.
Adjust A8 condenser (B10 condenser focusing screw)
and aperture (B3 Aperture diaphragm lever).
Adjust eld diaphragm (B5 eld diaphragm
revolver), adjust center screw B4 to center
eld diaphragm.
4.2 Operation
1) Illumination
1. Conrm the voltage is available
2. Turn on power switch B11
3. Turn B9 potentiometer to adjust light
6

2) Sample placing
Put the sample (cover is up) on stage A6 and t it with clamp A7. Turn B7, B8 stage knobs to put the sample
into the light path (Fig. 4).
uTwo samples can be placed on the stage together
3) Focusing (Fig. 5)
Put 10x objective into light path, turn coarse
focusing knob B13, observe from right eyepeice
with right eye, then turn slowly ne focusing
knob B14 to make the image clear after nding image.
u The tension adjustment ring B12 can adjust the tension of the coarse and ne focusing unit to prevent
the stage from sliding down automatically and to improve the comfort of operation. Rotate clockwise to
decrease tension, rotate counter-clockwise to increase tension.
u Don’t turn left and right coarse and ne focusing knobs in different directions with power at the same
time, if so, the focusing system will be damaged. (Fig. 6)
4) Interpupillary distance and diopter adjustment
a. Interpupillary distance adjustment
Interpupillary distance is different for everyone, so interpupillary
distance should be adjusted before using binocular microscope.
Please take the two eyepiece tubes to turn until the bright rings
observed by two eyes are in superposition entirely. (Fig. 7)
b. Diopter adjustment (Fig. 8)
As focusing for binocular, user should observe right eyepiece with right eye and make the right eyepiece
clear by focusing adjustment, then observe the left eyepiece at the same time, adjust the diopter ring of the
left eyepiece tube to make the image of the left eyepiece clear as same as the right eyepiece.
Fig. 4
Fig. 5
Fig. 6
7
Indice
Altri manuali Seiler Microscopio