Sensorex TX3100 Manuale utente

TX3100
Microprocessor
pH/ORP
Transmitter
Operation
Manual
TX3100

2
CONTENTS
Precautions for Installation
Brief Instructions ................................................................................................................................. 4
1. Specifications .................................................................................................................................... 7
2. Assembly and Installation
2.1 Transmitter Installation ................................................................................................................ 8
2.2 Panel Mounting Illustration ......................................................................................................... 8
2.3 Wall Mounting and Pipe Mounting Illustration ........................................................................... 9
2.4 Electrode and Housing Assembly .............................................................................................. 10
2.5 Round Junction Box Illustration and Description ...................................................................... 13
2.6 Rectangular Junction Box Illustration and Description ............................................................. 16
3. Overview of pH/ORP Transmitter TX3100
3.1 Rear Panel Illustration ............................................................................................................... 19
3.2 Terminal Function Illustration ................................................................................................... 19
3.3 Terminal Function Description .................................................................................................. 20
3.4 Accessorial Transmitter PH-300T Installation (Optional) ......................................................... 21
3.5 Transmitter and Accessorial Transmitter PH-300T Connection ................................................ 22
3.6 Wiring Illustration ...................................................................................................................... 23
3.7 Electrical Connection Illustration .............................................................................................. 24
3.8 Online pH/ORP Measurement System (Optional) ..................................................................... 25
4. Configuration
4.1 Front Panel Illustration .............................................................................................................. 26
4.2 Keypad ....................................................................................................................................... 26
4.3 LED Indicators ........................................................................................................................... 26
4.4 Display. ...................................................................................................................................... 27
5. Operation
5.1 Measurement Mode ................................................................................................................... 28
5.2 Settings Menu ............................................................................................................................ 28
5.3 Calibration Menu ....................................................................................................................... 28
5.4 Shortcuts .................................................................................................................................... 28
5.5 Default Values ............................................................................................................................ 28
6. Settings
Settings Block Diagram ................................................................................................................... 30
6.1 Settings Menu ............................................................................................................................ 32
6.2 Settings Security Code (Code) ................................................................................................... 33
6.3 Language .................................................................................................................................... 34
6.4 Measurement Parameter (Mode) ............................................................................................... 35

3
6.5 Multi-Point Calibration (Multi-Cal) .......................................................................................... 36
6.6 Temperature ............................................................................................................................... 37
6.7 Relay 1 ....................................................................................................................................... 38
6.8 Relay 2 ....................................................................................................................................... 39
6.9 Clean .......................................................................................................................................... 40
6.10 Analog Output 1 (pH/ORP) ..................................................................................................... 41
6.11 Analog Output 2 (Temperature) ............................................................................................... 42
6.12 Date/Time (Clock) ................................................................................................................... 43
6.13 Sample Measurement Average (Digital Filter) ........................................................................ 44
6.14 Back Light ................................................................................................................................ 45
6.15 Contrast .................................................................................................................................... 46
6.16 Automatic Return (Return) ...................................................................................................... 47
7. Calibration
Calibration Block Diagram .............................................................................................................. 48
7.1 Calibration Menu ....................................................................................................................... 49
7.2 Calibration Security Code (Code) .............................................................................................. 50
7.3 pH Calibration ............................................................................................................................ 51
7.4 ORP Calibration ......................................................................................................................... 55
7.5 Automatic Return (Return) ........................................................................................................ 56
8. Error Messages (Error Codes) ...................................................................................................... 57
9. Maintenance ................................................................................................................................... 58
Appendix ............................................................................................................................................. 59

4
Brief Instructions
◎Settings Menu Description (See Chapter 6 for Details)
Press and simultaneously to display settings information overview. Press_____to enter
settings menu. Press keypad according to the index bar at the bottom of the screen.
Keypad Index
Keypad Index Bar Description
Return to previous level or action
Left or left page
Increase digit
Right or right page
Decrease digit
Confirm and proceed to next step
Settings Items
Function Icon Description
Mode Measurement mode, select pH or ORP measurement
Multi-Cal. Multi-point calibration, adjust number of points
(PC-3310 provides up to three point calibration)
Temperature
Temperature measurement and compensation,
MTC - Manual temperature compensation
PTC/NTC - Auto temperature compensation
Relay 1 First relay settings, select Hi/Lo alarm or OFF
Relay 2 Second relay settings, select Hi/Lo alarm or OFF
Clean Automatic wash time settings
Set cleaning relay ON/OFF duration

5
Analog 1 Current output corresponding to pH or ORP range
Analog 2 Current output corresponding to temperature range
Clock Time and date settings
Digital Filter Digital filter settings, adjust number of samples to be
averaged for each reading
Back Light Backlight settings, set backlight Auto/ON/OFF,
brightness, and sensitivity
Contrast Screen contrast settings
Return Measurement mode return settings
Code
Settings passcode setup, the settings passcode is
precedential to the calibration passcode and can be used to
bypass calibration lock
Language Available in English, Traditional Chinese, and Simplified
Chinese

6
◎Calibration Menu Description (See Chapter 7 for Details)
Press and simultaneously to display calibration information overview. Press_____to
enter calibration menu. Press keypad according to the index bar at the bottom of the screen.
Keypad Index:
Keypad Accordingly item Description
Return to previous level or action
Left or left page
Increase digit
Right or right page
Decrease digit
Confirm and proceed to next step
Calibration Items:
Function Icon Description
TECH Use TECH buffers as standard solution for calibration
NIST Use NIST standard buffers (DIN 19266) as standard
solution for calibration
Any Use any buffer solution by users’ definition for calibration
Return Measurement mode return settings
Code Calibration passcode setup

7
1.Specifications
Model TX3100
Measurement Modes pH/ORP/Temp
Range
pH -2.00~16.00 pH
ORP -1999~1999 mV
TEMP -30.0~130.0˚C
Resolution
pH 0.01 pH
ORP 1 mV
TEMP 0.1˚C
Accuracy
pH ±0.01 pH (±1 Digit)
ORP ±0.1% (±1 Digit)
TEMP ±0.2˚C (±1 Digit)
with temperature error correction
Temperature
Compensation
NTC30K / PT1K Auto temperature compensation
Manual temperature compensation
Calibration Mode TECH, NIST, Any Buffer, up to three point calibration
Ambient Temp. 0~50˚C
Storage Temp. -20~70˚C
Input Impedance > 1012Ω
Display Large LCM with backlight sensor and contrast adjustment;
Text mode
Language Available in English, Traditional Chinese, Simplified Chinese
Analog Output 1 Isolated DC 0/4~20 mA corresponding to main measurement,
Max load 500Ω
Analog Output 2 Isolated DC 0/4~20mA corresponding to temperature,
Max load 500Ω
Settings Contact RELAY ON/OFF contact, 240VAC 0.5A Max (recommended)
Activate Hi/Lo, Hi/Hi, Lo/Lo, selectable two limited programmable, ON/OFF
Wash RELAY contact, ON 0~99 min 59 sec / OFF 0~999 hr 59 min
Voltage Output DC±12V, 1W Max for PH-300T (Optional)
Protection IP65
Power Supply 100V~240VAC±10%, 6W Max, 50/60Hz
Installation Wall / Pipe / Panel Mounting
Dimensions 144 mm × 144 mm × 115 mm (H×W×D)
Cut off Dimensions 138 mm × 138 mm (HW)
Weight 0.8 kg

8
2. Assembly and Installation
2.1 Transmitter Installation
This transmitter can be installed by panel mounting, wall mounting, or 2” pipe mounting.
Panel Mounting:
Prepare a square hole of 138 mm x 138 mm on the panel box, then insert the controller directly
into the hole. Insert the accessorial mounting bracket from the rear, and fix into the pickup groove.
2.2 Panel Mounting Illustration
Min. 200 mm
Min. 200 mm
Hole distances on the panel box
Panel mounting illustration,
fixed with mounting bracket
138 mm
138 mm
Hole size
++
+
+
Mounting bracket
90 mm

10
Put heat-shrinkable
sleeve on coaxial shield
2.4 Electrode and Housing Assembly
2.4.1 Cable Preparation (Cable Stripping Procedure)
a. Strip off approximately 40 mm of the pH/ORP coaxial cable jacket; separate the coaxial inner
wire and coaxial shield wire as seen in the illustration above.
b. Make sure to remove the conductive rubber of the coaxial inner as seen in the following
illustration.
c. Put heat-shrinkable sleeve on the coaxial shield.
d. Attach connection tips on both ends of coaxial inner and coaxial shield.
e. Extend the cable to the transmitter without any joints except at the junction box. Connect the
transparent coaxial inner directly to the GLASS terminal on the back of transmitter, then
connect the metal coaxial shield to the REF terminal.
pH/ORP Signal Cable Preparation Instruction:
Refer to the cable preparation illustration above. Peel 5 mm of the black conductive rubber wrap
on the coaxial inner away from the connection tip.
The length of the separated coaxial inner and coaxial shield wires is suggested to be approx. 40
mm for the convenience of wiring.
6mm 12mm Conductive rubber
Coaxial Inner Stripping Illustration
Transparent coaxial inner
(Peel away conductive rubber)
5mm
40mm
pH/ORP Cable Preparation

19
3. Overview of pH/ORP Transmitter TX3100
3.1 Rear Panel Illustration
3.2 Terminal Function Illustration
POWER
WASH/Cln
2
4/20m
A
+
-
±12V
+
-
REL2REL1
AC
INPUT
POWER
1
4/20m
A
+
-
Temp
(T/PShield)
SG
T/P
GLASS
REF
pH/
ORP
Indice
Altri manuali Sensorex Trasmettitore

Sensorex
Sensorex TX3000 Manuale utente

Sensorex
Sensorex CX100 Manuale utente

Sensorex
Sensorex TX2000 Manuale utente

Sensorex
Sensorex CX105 Manuale utente

Sensorex
Sensorex TX100 Scheda tecnica

Sensorex
Sensorex TX105 Manuale utente

Sensorex
Sensorex TX2000 Manuale utente

Sensorex
Sensorex CT-1000 Manuale utente

Sensorex
Sensorex TX2000 Manuale utente