ServAPure INTEGRAL E-POD ZRXSP0D01 Manuale utente

INTEGRAL RANGE - E-POD® UNIT
I N S TA L L AT I O N O N E L I X A D VA N TA G E
SYSTEMS
Cat. #2RXSP0D01
Installation Procedure Kit for 2 or 3 E-POD Units
Where Used • Elix Advantage 3-5-10-15 Systems
In the • 1 E-POD Unit (base, mast, arm and dispenser)
Shipping Box , i RC-Link (sheath containing 2 pieces of tubing and a PS/2 cable)
• 1 coupling straight fitting and 2 elbow guides
• 1 Allen wrench (5 mm)
• E-POD/Q-POD stickers
O t h e r To o l s • To r x 1 0 s c r e w d r i v e r
Needed , Tubing Cutters
Installation • The purpose of this installation procedure sheet is to explain how to
Procedure connect a Second or Third E-POD Unit.
• This procedure also contains information regarding use of multiple E-
POD Units.
• Please refer to the System User Manual regarding the assembly,
installation and use of the primary E-POD Unit.
T h e i n f o r m a t i o n i n t h i s d o c u m e n t i s m e a n t t o b e u s e d b y a Q u a l i fi e d S e r v i c e R e p r e s e n t a t i v e
P F 1 1 1 5 7 R E V 2 0 2 / 2 0 1 3 L A B W A T E R I N S E R T 1 / 5

Installation of a Second E-POD Unit
Preparation of The actions to perform on the Elix Advantage system cabinet are
the Elix described below:
Advantage
step Action
1Place the Elix Advantage system into Standby Mode
2Depressurise the system by pushing the E-POD Plunger
(on E-POD 1) all the way down and releasing it
3Wait until no more water flows out from the E-POD Unit
4Push the E-POD Plunger all the way down and release it
5Power off the Elix Advantage system by disconnecting the
power cord
Over Moulded The position of the over moulded ferrite on the PS2 cables is as shown
Ferrite below.
Location
Be tw ee n Sy st em a nd E -P OD 1 Between E-POD 1 and E-POD 2
T h e o v e r m o u l d e d f e r r i t e i s c l o s e
to the cabinet.
\^E^D r--—3
The over moulded ferrite is close
to E-POD 1.
Preparation of Below are the actions to perform on the second E-POD Unit'
the 2"'* E-POD
Step Action
1Remove the E-POD Unit from the shipping box. Locate all
items mentioned above ("In the Shipping Box")
2Assemble the E-POD Unit. See the User Manual for
Instructions
Continued on next page
THE INFORMATION IN THIS DOCUMENT IS MEANT TO BE USED BY A QUALIFIED SERVICE REPRESENTATIVE
P F 1 1 1 5 7 _ R E V 2 _ 0 2 / 2 0 1 3 L A B W A T E R i n s e r t 2 / 5

Installation of a Second E-POD Unit, Continued
Preparation of
the 2nd E-POD
Unit
(continued)
Step Action
3Connect the tubing, PS/2 cable and termination plug on the
2"" E-POD base
Preparation of Below are the actions to perform on the first E-POD Unit,
the E-POD
Step Action
1Plug the PS/2 c
and E-POD 2) ir
t h e o v e r m o u l d e
PS/2 Cable
between
E-POD 1 —
and system
able (the one connected between E-POD 1
ito the E-POD 1 Unit. Make sure to position
d f e r r i t e a s s h o w n b e l o w.
E-POD 1
PS/2 cable
y between E-POD 1
/ and E-POD 2
2The E-POD 1 Unit has now two PS/2 cables plugged into it.
Continued on next page
T h e i n f o r m a t i o n i n t h i s d o c u m e n t i s m e a n t t o b e u s e d b y a Q u a l i fi e d S e r v i c e R e p r e s e n t a t i v e
P F 1 1 1 5 7 R E V 2 0 2 / 2 0 1 3 L A B W A T E R I N S E R T 3 / 5

Installation of a Second E-POD Unit, Continued
Tubing
Connections
between E-
POD Units
Below are the actions to perform for the hydraulic connections between
the 2 E-POD Units.
Step Action
1Remove the tubing from one port on the E-POD 1 unit, it
does not matter which port is used.
2Using the coupling connect the loose end of tubing from the
E-POD 1 unit to the RC-Link used with the E-POD 2 unit.
3The RC-Link for the E-POD 2 unit now has 1 loose piece of
tubing. Connect this loose end of tubing to the E-POD 1
unit.
4Check that all tubing and electrical cables are connected as
below and are well connected to the E-POD units.
The information in this document is meant to be used by a Qualified Service Representative
P F 1 1 1 5 7 _ R E V 2 _ 0 2 / 2 0 1 3 l a b w / a t e r i n s e r t 4 / 5

Important information
Follow the steps below to reset the POD's.
Step Action
1 Plug in the power cord to the system.
2 Check that the E-POD 1 Unit displays "Pod 1" and E-POD
2 Unit displays "Pod 2". If not, proceed to the next step.
^Pressandrelease®and©onthePODkeypadatthe
same time to reset the E-POD 2 Unit.
If... Then ...
"Pod 2" is not displayed on Reset again by pressing
the E-POD 2 display
E-POD 2 unit is stuck on
"Pod 2"
t h e s a m e b u t t o n s o n t h e E -
POD Unit.
Restart the system by
unplugging and plugging
back the power cord.
Using Multiple E-POD Units
Introduction There are some important things to know when multiple E-POD Units
are used and these are listed below.
Important • it is not possible to dispense from multiple E-POD Units at the same
Information time.
• Do not use an E-POD plunger when another E-POD Unit is
dispensing water.
• The E-POD display indicates if the E-POD Unit is not available for
dispensing water. See the User Manual or the Quick Reference
Guide for more information about how the E-POD Unit indicates that it
is not available for dispensing water.
POD Pak • The E-POD Units may or may not have different POD Paks on them.
For example, you may have a Millipak® Filter on the E-POD 1 Unit
a n d a B i o P a k F i l t e r o n t h e E - P O D 2 U n i t .
• Make sure the POD Pak is registered on each E-POD Unit. See User
Manual for more information about registering the type of POD Pak
on an E-POD Unit.
T h e i n f o r m a t i o n i n t h i s d o c u m e n t i s m e a n t t o b e u s e d b y a Q u a l i fi e d S e r v i c e R e p r e s e n t a t i v e
P F 1 1 1 5 7 R E V 2 0 2 / 2 0 1 3 L A B W A T E R I N S E R T 5 / 5
Indice