tagarno FHD ZAP Manuale utente

VERSION: 1.9.1 | FIRMWARE 5.8 | 2021-06-15
1
CONTENTS
1. INTENDED USE 2
2. WARNINGS 2
3. TIPS 2
4. LASER POINTER WARNING 2
5. YOU HAVE RECEIVED 3
6. ASSEMBLING - FLEX ARM 4
7. ASSEMBLING - OTHER MOUNTING
OPTIONS 6
8. ASSEMBLING - LASER BEAM
(US ONLY) 7
MANUAL | TAGARNO FHD ZAP
9. OPERATION 10
10. TECHNICAL SPECIFICATIONS 12
11. PC REQUIREMENTS 14
12. MAINTENANCE 15
13. WARRANTY 15
14. ERGONOMICAL RECOMMENDATIONS 15
13. DECLARATION OF CONFORMITY 16
14. COMPLIANCE STATEMENTS 17
Go to tagarno.com/productmanuals to nd manuals and tutorial videos.

2
1. Read the manual before you use the product
2. Use the product only as specied, or the protection supplied by the product can
be compromised
3. Do not position the equipment so that it is difcult to operate the disconnecting
device (appliance inlet of external power supply, equipment input connector)
4. If fluids are spilled on the product, turn the system off immediately by pulling the power
supply out of the electrical outlet
5. In case of re close to the microscope, please turn off and disconnect the system
6. Avoid subjecting the lens to sharp or hard objects
7. Please do not connect the microscope, if visible damages appear
8. Do not dismantle any parts of the microscope, except where noted in the manual
9. Never disassemble or clean internal optical surfaces
10. Use only the power supply provided from TAGARNO
11. Always turn off the system before unplugging, when possible
WARNINGS
The product is a digital magnifying system consisting of a camera unit, PCBs, mechanical
parts and a power supply. The product is intended for marketing worldwide and is designed for
manual visual inspection.
INTENDED USE
Read all safety information before you use the product.
Please pay attention when you see a warning label on the product.
This product is for indoor use only.
You must not discard this electrical/electronic product in domestic household waste.
Please dispose at your local recycling centre.
3. TIPS
1. INTENDED USE
2. WARNINGS
1. When applicable, use both hands to move the XY Table
2. When applicable, use both hands to adjust the height of the product
3. Avoid touching the lens
4. LASER POINTER WARNING
This product is equipped with a red laser pointer to enable easy alignment of the camera and areas of interest during the inspection process.
This product is a Class 2 laser product that complies with IEC60825-1 international standard for lasers.
These labels appear visible on the product:CAUTION
Follow these safety instructions when using the product.
• Never look directly into the laser aperture
• Do not point towards anyone deliberately
• Leave the laser on only when necessary
• Always turn off power during service and maintenance
• Service may only be performed by trained personnel
appointed by TAGARNO
This label is placed near the laser aperture
The label below is placed on the camera head

3
SMART GUIDE
Smart guide Camera head with lens XKEY Control box Handles
5. YOU HAVE RECEIVED
Flex arm advanced
OPTIONS
Flex arm inline
Power supply

4
6. ASSEMBLING - FLEX ARM
ASSEMBLING | FLEXARM (1/2)
Press the cables into place in the slot on the flex arm and secure
them by screwing the cable holder into place on top.
Mount the table mount using the table clam. Assemble the bottom
of the flex arm and the table mount.
NB: For maximum stability at all magncation levels, make sure to
mount the microscope on a sturdy table.
Assemble the flex arm by slotting the two arms together.
Mount the microscope to the top of the flex arm using the vesa
bracket, screws and handles.
TIP!
Find the “How to Install” tutorial video here:
www.tagarno.com/installzap

5
1
2
3
To increase stability, adjust the flex arm weight with the included
allen key. Turn it clockwise to decrease stability or counter
clockwise to increase stability.
For extra stability in locked positions, remove the rubber covering
on the side of the flex arm.
ASSEMBLING | FLEX ARM (2/2)
For further stability, tighten screws in numbered areas.Then, use the included nger screw to lock vertical movement.

6
7. ASSEMBLING - OTHER MOUNTING OPTIONS
ASSEMBLING | OTHER MOUNTING OPTIONS
Alternatively, mount the microscope on a wall with an angle
bracket or similar solution suitable for your process environment.
These mounting options are not included with your delivery.
To use angle bracket, detach Vesa mount and mount the angle
bracket to the microscope.
For safe operation, mount only the microscope on support
equipment capable of bearing a load of at least 6kg (3 times the
weight of the microscope).

7
8. ASSEMBLING - LASER BEAM (US ONLY)
IMPORTANT! Laser beam attenuator. Unmount lens protection cap
when using this microscope.

8
3. 4. 5.1. 2.
ASSEMBLING & CONNECTING (2/3)
Wire Description Type Connect to Cable length Specication
1Output HDMI Type A Monitor 2700mm HDMI out 720p50, 720p60, 1080p25, 1080p30, 1080p50, 1080p60
2USB 3.0 output USB 3.0 Type A Computer 2700mm USB 3.0 Device, Self-powered
3Control box D-SUB 9-Pin
Female
Control box 3000mm Output: 3.3V 330µA
Use only control boxes supplied by TAGARNO
4Power supply
(DC)
DC Jack Power socket 3000mm Input: 12V 0.5A
Use only the power supply provided by TAGARNO:
Brand/Model: Mean Well/GEM18I12
Protection class II
Input: 100-240V 0.45-0.2A, 50/60Hz
Overvoltage category ll
Output: 12V 1.5A Max
5ESD Ring Terminal
Connector
ESD Grounding
point
3000mm ESD Grounding point
Connect HDMI cable (1), USB 3.0 cable (2), control box (3), power
supply (4) and ESD - Electro Static Discharge (5).
WARNING! Do not use force when inserting the cable plugs.
Connect HDMI cable from the microscope to a monitor.
WARNING! Do not connect the HDMI cable to a PC.

9
Connect the microscope to the XKEY Control box.
ASSEMBLING & CONNECTING (3/3)
Connect the power supply cable to the power supply battery and
then to a electrical outlet.
Connect grounding cable to grounding point, to ensure ESD of the
microscope.
Connect USB 3.0 cable from the microscope to a PC.

10
Short push switches between Iris, Gain and Preset mode.
Increase Iris/Gain or switch between zoom preset 1, 2 or 3.
Decrease Iris/Gain or switch between zoom preset 1, 2 or 3.
To change and save a zoom preset, use zoom buttons to
select a zoom level and simultaneously press left/right
buttons. OSD will show STORED: PRE(X)
Hold for 3 sec. to use auto exposure mode
RESET TO FACTORY SETTINGS
Follow this procedure to change back to factory setting 1080P60.
1. Turn the microscope off
2. Press and hold the center button down while turning the power on
3. Keep holding the center button down for 25 seconds
4. Release the center button and turn the power off
5. Turn the power back on and the microscope is set to factory
setting 1080P60
Turn off/on
Laser on/off (WARNING! Laser radiation when turned on)
Zoom in
Zoom out
Push both plus and minus to change from auto focus to
manual focus. You can now adjust the focus manually by
using the plus/minus buttons
If you wish to focus on something close
If you wish to focus on something further away
9. OPERATION
XKEY Control box for TAGARNO functions.
Do only connect equipment distributed by TAGARNO.
Control buttons for TAGARNO functions.
Do only connect equipment distributed by TAGARNO.
CONTROL OPTION | XKEY CONTROL BOX
OPERATION (1/2)
TIP!
Find the "How to Use" tutorial video here:
www.tagarno.com/usezap
Altri manuali per FHD ZAP
1
Questo manuale è adatto per i seguenti modelli
1
Indice
Altri manuali tagarno Microscopio