SPECIFICATIONS
MODEL 865 SPECIFICATIONS
TEMPERATURE SENSOR TYPE: Thermistor (2252 Ohm @ 77°F)
*ACCURACY: Includes repeatability, temperature coefficient (65° to 82°F), time sta-
bility (1 year), and linearization conformity errors. Excludes probe errors; however
probe errors around 32°F may be compensated by an internal adjustment.
REPEATABILITY: 0.2°F typical for 1 week at constant ambient temperature.
TEMPERATURE COEFFICIENT: 65° to 82°F; included in accuracy specification.
From 14° to 65° and 82° to 122°F; less than ±(0.06% rdg + 0.01°F)/°F.
SENSOR VOLTAGE: 40mV rms maximum.
MODEL 866 SPECIFICATIONS
TEMPERATURE SENSOR TYPE: Thermistor (2252 Ohm @ 25°C)
*ACCURACY: Includes repeatability, temperature coefficient (18° to 28°C), time sta-
bility (1 year), and linearization conformity errors. Excludes probe errors; however
probe errors around 0°C may be compensated by an internal adjustment.
REPEATABILITY: 0.1°C typical for 1 week at constant ambient temperature.
TEMPERATURE COEFFICIENT: 18° to 28°C; included in accuracy specification.
From –10° to 18° and 28° to 50°C; less than ±(0.06% rdg + 0.01°C)/°C.
SENSOR VOLTAGE: 40mV rms maximum.
MODEL 865 and 866 GENERAL SPECIFICATIONS (unless specified):
DISPLAY: 3-1/2-digit LCD, 13mm (0.5") height. Polarity and decimal point indication.
CONVERSION RATE: 1.5 readings per second.
OPEN SENSOR INDICATION: Display reads between –100° and –199°.
OVERRANGE INDICATION: 3 least significant digits blanked.
TEMPERATURE ACCURACY*
RANGE SPAN RESOLUTION (65° to 82°F; 1 year)
200°F –40.0° to 199.9° 0.1° ±(0.3% rdg + 0.5°F)
–70.0° to –40.1° 0.1° ±(0.3% rdg + 1.0°F)
300°F –70° to 300° 1° ±(0.3% rdg + 1°F)
TEMPERATURE ACCURACY*
RANGE SPAN RESOLUTION (18° to 28°C; 1 year)
150°C–40.0° to 150.0° 0.1° ±(0.3% rdg + 0.3°C)
–55.0° to –40.1° 0.1° ±(0.3% rdg + 0.6°C)