TES TES-2730 Manuale utente

DIGITAL
MULTIMETER
TES-2730
INSTRUCTION MANUAL
TES ELECTRICAL ELECTRONIC CORP.

CONTENTS
TITLE PAGE
SAFETY INFORMATION ................................................................ 1
I. SPECIFICATIONS ........................................................................ 2
1-1 General Specifications ......................................................... 2
1-2 Electrical Specifications ....................................................... 4
II. NAME OF PARTS AND POSITIONS .......................................... 8
III. PRECAUTIONS AND PREPARATIONS FOR
MEASUREMENT ........................................................................ 10
IV. MEASUREMENTS ..................................................................... 11
4-1 DC Voltage Measurement ................................................... 11
4-2 AC Voltage Measurement ................................................... 12
4-3 DC Current Measurement ................................................... 12
4-4 AC Current Measurement .................................................... 13
4-5 Resistance Measurement .................................................... 13
4-6 Capacitance Measurement .................................................. 14
4-7 Frequency Measurement ..................................................... 14
4-8 Temperature Measurement ................................................. 14
4-9 Diode Tests .......................................................................... 15
4-10 Continuity Measurement .................................................... 15
V. BATTERY & FUSE REPLACEMENT ......................................... 16
5-1 Battery Check-up & Replacement ....................................... 16
5-2 FUSE Replacement ............................................................. 16
VI. OPTIONAL ACCESSORY ......................................................... 19

1
SAFETY INFORMATION
Read the following safety information carefully before attempting to
operate or service the meter.
To avoid damages to the instrument do not apply the signals which
exceed the maximum limits shown in the technical specifications tables.
Never measure current while the test leads are inserted into the input
jacks.
Do not use the meter or test leads if they look damaged. Use extreme
caution when working around bare conductors or bus bars.
Accidental contact with the conductor could result in electric shock.
Use the meter only as specified in this manual ; otherwise, the protection
provided by the meter may be impaired.
Read the operating instructions before use and follow all safety
information.
Caution when working with voltages above 60V DC or 30 V AC RMS.
Such voltages pose a shock hazard.
Before taking resistance measurements or testing acoustic continuity,
disconnect circuit from main power supply and all loads from circuit.
Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or
reinforced insulation.
When servicing, use only specified replacement parts.
Comply with EN-61010-1

2
1. SPECIFICATIONS
1-1 General Information
Environment conditions :
Installation Categories
II
Pollution Degree 2
Altitude up to 2000 meters
Indoor use only
Relatively humidity 80% max
Operation Ambient 0〜40℃
maintenance & Clearing :
Repairs or servicing not covered in this
manual should only be performed by
qualified personal.
Periodically wipe the case with a dry cloth
and detergent.
Do not use abrasives or solvents on this
instruments.
Operating Principle :Dual slope integration.
Numerical Display :3 1/2 digit liquid crystal display (LCD)
7mm height. maximum reading 1999 .
Unit and Sign Display :
Fig-1

3
DC Direct current or voltage
AC Alternating current or voltage
Negative polarity
■Decimal point
V Volts
mV Millivolts ( 1×10
-3
volts )
A Ampere ( amps ). Current
uA Microamperes ( 1×10
-6
amps )
uF Microfarads ( 1×10
-6
Farads ), Capacitance.
nF Nanofarads ( 1×10
-9
Farads ), Capacitance.
pF Picofarads ( 1×10
-12
Farads ), Capacitance.
ΩOhms. Resistance
KΩKilohms ( 1×10
3
ohms ). Resistance
MΩMegohm ( 1×10
6
ohms ). Resistance
MHz Megahertz ( 1×10
6
ohms ). Frequency
KHz Kilohertz ( 1×10
3
cycles / sec ). Frequency
℃Celsius temperature scale
℉Fahrenheit temperature scale
Low Battery
Data Hold
Continuity Beeper
Diode
Range Selection :All ranges are measured by single Range
switch operation.
Over Range Indication :“
OL
“ appears on the display.
Low Battery Indication :The is displayed when the battery
voltage drops below the operating voltage.
Sampling Rate :2.5 time per second except in Freq. Mode.
1 time per second in Freq. Mode.
Power Requirements :9-Volt battery (NEDA or JIS 006p IEC6f22)
Battery Life ( typical ) :Approx. 60hours. (Alkaline Battery)
Operating Temperature :0℃to 40℃( 32℉to 104℉)

4
and Humidity RH below 80%
Storage Temperature :-10℃to 60℃( 14℉to 140℉)
and Humidity RH below 70%
Dimensions :180 (L) ×82 (W) ×38(H) mm
7.09”(L)×3.2”(W)×1.5”(H)
Weight :365g
Accessories :Test leads (pair), Operating Instruction,
Spare fuse ( 0.5A / 250V ), Software, Battery,
RS-232C Cable, 9pin to 25pin Gender
Changer.
Optional Accessories :K type thermocouple.
1-2 Electrical Specifications
Accuracies are ±(...% of reading + ...digits) at 23℃± 5℃,below 80% RH.
DC Voltage
Range Resolution Accuracy Input
Impedance
Overload
Protection
200m
V 0.1mV 600Vrms
2V 1mV
20V 10mV
200V 100mV
0.5%+1
600V 1V 0.5%+2
10MΩ
750Vrms
AC Voltage ( 50Hz to 500Hz )
Range Resolution Accuracy Input
Impedance
Overload
Protection
200mV 0.1mV 600Vrms
2V 1mV
20V 10mV
1%+5 10MΩ
750Vrms

5
200V 100mV
600V 1V 1.2%+5
DC Current
Range Resolutio
n Accuracy Burden
Voltage Overload Protection
200uA 0.1uA
2mA 1uA
20mA 10uA
200mA 100uA
1%+1 0.35V 0.5A / 250V Fast
Blow Fuse & Diode
20A 10mA 1.2%+3 0.8V
20A / 380V Fast Blow
Fuse
20A for 30sec. max.
10A continuity.
AC Current ( 50Hz to 500Hz )
Range Resolutio
n Accuracy Burden
Voltage Overload Protection
200uA 0.1uA
2mA 1uA
20mA 10uA
200mA 100uA
1.2%+3 0.35V 0.5A / 250V Fast
Blow Fuse & Diode
20A 10mA 1.5%+5 0.8V
20A / 380V Fast Blow
Fuse
20A for 30sec. max.
10A continuity.

6
Resistance ( Ω)
Range Resolution Accuracy Max. Open
Circuit Voltage
Overload
Protection
200Ω0.1Ω3.2V
2KΩ1
Ω
20KΩ10Ω
200KΩ100Ω
2MΩ1KΩ
0.8%+2
20MΩ10KΩ1.5%+3
0.5V 600Vrms
Diode ( )
Range Resolution Accuracy Max. Open
Circuit Voltage
Max. Test
Current
Overload
Protection
1mV 2%+ 2 3.2V 1.0mA 600Vrms
Continuity ( )
Range Operation Resolution Max. Open
Circuit Voltage
Overload
Protection
Continuity audible tone for
tested resistance below 30Ω3.2V 600Vrms

7
Frequency ( Hz ) ( Test Range 10Hz - 10MHz Auto Range )
Range Resolution Accuracy Sensitivity
Overload
Protection
2KHz 1Hz
20KHz 10Hz
200KH
z 100Hz
2MHz 1KHz
1.5Vrms
10MHz 10KHz
0.5%+2
3Vrms
600Vrms
Temperature Probe used : K (CA) type sensor
Range Resolution Accuracy
℃0.1/1℃
0℃〜200℃±(0.5%+1.5℃)
-50℃〜0℃±( 1% + 2℃)
200℃〜1300℃±( 1% + 2℃)
℉0.1/1℉
0℉〜200℉±( 0.5%+3℉)
-58℉〜0℉±( 1% + 5℉)
200℉〜1999℉±( 1% + 4℉)
Capacitance (F)
Range Resolution Accuracy Test Frequency
2000pF 1pF
200nF 100pF
2uF 1nF
20uF 10nF
3%+10 400Hz

8
II.
NAME OF PARTS AND POSITIONS
Fig-2
Indice
Altri manuali TES Multimetro
Manuali Multimetro popolari di altre marche

Gossen MetraWatt
Gossen MetraWatt METRAmax 6 Manuale utente

PeakTech
PeakTech 4000 Manuale per l'uso e la cura

YOKOGAWA
YOKOGAWA 90050B Manuale utente

Gossen MetraWatt
Gossen MetraWatt METRALINE DMM16 Manuale utente

Fluke
Fluke 8846A Manuale operativo e di manutenzione

Tempo Communications
Tempo Communications MM200 Manuale utente