TTrig // Quick Guide
4
D04-05Aen201908 Quick GuideTTrig
TriOS Mess- und Datentechnik GmbH · Bürgermeister-Brötje-Str. 25 · D-26180 Rastede · Germany
Phone:
+49
(0)
4402
69670
-
0
·
F
ax:
+49
(0)
4402
69670
-
20
·
[email protected] ·
www.trios.de
Start of automatic measurement
Pressing the start button ("3. START") starts the automatic measurement.This function of the start button is only available again
after a reset or disconnection of the power supply.
The following procedure should be observed after the start of the automatic measurement:
Sequence of the rst measurement after activation by pressing the start button.
START LED WIPER LED SENSOR LED Function
1. Flashes quickly OFF OFF Sensor initialisation
2. Flashes quickly ON OFF Wiper cleans
3. Flashes quickly OFF ON Sensor boots
4.* Flashes every 20 s OFF Flashes 3 x fast Measurement
5. Flashes every 20 s OFF ON Data backup
6. Flashes every 20 s OFF OFF Automatic measurement running
During active automatic measurement, the start LED lights up once every 20 seconds.
* If instead the yellow service LED ashes every 20 s, please refer to the section Troubleshooting/Sensor not
found.
Data storage and retrieval
All measurement data is only stored on the internal 2GB SD card of the sensor.
A notebook can be connected via an RJ-45 socket on theTTrig so that the stored measurement data can be downloaded via the
web interface. Press the "Service" button on theTTrig to supply the sensor with power for the entire period. In service mode the
automatic measurements are interrupted, "Skipped" is entered in the comment line of the next measurement.
Establishing a connection to the sensor
Step 1 Connect a notebook/PC to theTTrig via the RJ-45 socket.
Step 2 Press the "Service" button on theTTrig to activate the service mode.
Step3 Open a web browser. The following URL establishes the connection to your sensor: http://192.168.77.1/ (see also
manual of the respective sensor)
Comments
Each record has a comment function with four comment lines.
They can be found in the web interface under the menu "Measurement" "Settings".
Lines 1 and 2 are freely available to the user.
Lines 3 and 4 are overwritten byTTrig.
The settings of the rotary encoders are queried once. Later adjustment has no eect on the measur-
ing intervals or the delay!
NOTICE