Xylem YSI 910 COD Manuale utente

910 COD Colorimeter
INSTRUCTION MANUAL: PAGES 2 - 13
MODE D‘EMPLOI: PAGES 14 - 25
INSTRUCCIONES: PÁGINA 26– 37
OPERATIONS MANUAL

Contact Information
ORDERING & TECHNICAL SUPPORT
Telephone: (800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
Fax: (937) 767-9353 (orders)
(937) 767-1058 (technical support)
Email: [email protected]
Mail: YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA
Internet: www.ysi.com
When placing an order please have the following information available:
•YSIaccountnumber(ifavailable)
•NameandPhoneNumber
•Modelnumberorbriefdescription
•Billingandshippingaddress
•Quantity
•PurchaseOrderorCreditCard
SERVICE INFORMATION
YSI has authorized service centers throughout the United States and Internationally. For the nearest
servicecenterinformation,pleasevisitwww.ysi.comandclick‘Support’orcontactYSITechnical
Support directly at 800-897-4151.
Whenreturningaproductforservice,includetheProductReturnformwithcleaningcertication.
TheformmustbecompletelylledoutforanYSIServiceCentertoaccepttheinstrumentforservice.
TheProductReturnformmaybedownloadedatwww.ysi.comandclickingonthe‘Support”tab.

3
Table of Contents
CAUTION
Theaccuracyoftheinstrumentisonlyvalidiftheinstrumentisusedinan
environmentwithcontrolledelectromagneticdisturbancesaccordingtoDIN61326.
Wirelessdevices,e.g.wirelessphones,mustnotbeusedneartheinstrument.
General Notes ............................................................4
General Notes .............................................................4
Method Notes .............................................................4
DisplayBacklight ...........................................................5
RecallofStoredData .......................................................5
Replacementofbatteries.....................................................5
Methods .................................................................6
COD Reagents ............................................................6
Preparing the Sample .......................................................6
Guidelinesforphotometricmeasurements .......................................6
Measurement..............................................................7
Menu options .............................................................8
Menu selections............................................................8
Recallofstoreddata ........................................................8
Transmitting stored data .....................................................9
Setting date and time........................................................9
Calibration Mode .........................................................10
Usercalibration ...........................................................10
Factorycalibrationreset ....................................................11
Technical data ...........................................................12
Operating messages .......................................................13
Error codes ..............................................................13
Battery Disposal ..........................................................13
ENGLISH

4
General Notes
• Priortomeasurementensurethatthesampleissuitableforanalysis(nomajorinterference)and
doesnotrequireanypreparationi.e.pHadjustment,ltrationetc.
• Reagentsaredesignedforuseinchemicalanalysisonlyandshouldbekeptwelloutofthereach
ofchildren.
• Ensureproperdisposalofreagentsolutions.
• MaterialSafetyDataSheets:availableonrequest.
Chemical Method Notes:
Method:
The sample is heated at 150oCfortwohoursinthepresenceofpotassiumdichromate,whichisa
strong oxidizing agent. Organic compounds in the sample are oxidized, reducing the dichromate ion
(Cr2O72–) to the green chromic ion (Cr3+).
IntheLowRange(LR)Test(0-150mg/L),theamountofremainingCr6+ is measured; In the
MidRange(MR)(0-1,500mg/L)or(HR)(0-15,000mg/L),theamountofCr3+ produced is
measured.Theresultsaremeasuredastheamountofoxygen(O2)consumedperliterof
water sample. The LR and MR test is displayed in mg/L while the HR test shows g/L.
TheCODreagentcontainsbothmercuryandsilver—themercurycontrolschloride
interferencesandthesilverfunctionsasacatalysttospeedthereaction.
Application:
Samplescanbemeasuredifthechloridecontentdoesnotexceed1,000mg/l(LR/MR)or10,000mg/l
(HR).
Inexceptionalcases,compoundscontainedinthewatercannotbeadequatelyoxidized.Thisresults
inlowreading,comparedwiththereferencemethod.
Differentmethodsofsampling,preparationofthesampleitselfandthetimeelapsedbetweentaking
thesampleandanalysiscanallaffecttheobtainedresults.
General Notes

5
Replacement of batteries:
CAUTION:
Toensurethattheinstrumentiswaterproof:
• sealring(E)mustbeinposition
• batterycompartmentcover(B)mustbexedwiththefourscrews
Ifthebatteriesareremovedformorethanoneminutethedateandtimemenustartsautomaticallywhen
the photometer is turned on the next time.
AAA/LR 03
AAA/LR 03
AAA/LR03
AAA/LR03
(D)batteries
(E) seal ring
(C) notch
(F)instrumentback
(B)batterycompartmentcover
(A) screws
ENGLISH
Pressthe[!]keytoturnthedisplaybacklightonoroff.Thebacklightisautomati-
callyturnedoffduringthemeasurement.
Iftheinstrumentisturnedon,pressthe[!]keyformorethan
4 seconds to access the recall menu.
!
!
Recall of stored data
Display backlight

6
Methods
COD Reagents
Selecttheappropriatevialforthedesiredrange*:
LR-Range: 0–150mg/l,partnumber251990Y,25count
MR-Range: 0–1500mg/l,partnumber251991Y,25count
HR-Range: 0–15000mg/l,partnumber251992Y,25count
*Reagents are also availble in 150 count packages.
Preparing the Sample
Openareactionvialwithawhitecapandaddthespeciedvolume.
(Ensurethatappropriatepersonnelsafetyequipmentisused.)
LR-/MR: 2 ml water sample
HR: 0.2 ml water sample
Prepareablank(Note1)byusingdeionizedwater(TOC-free)insteadofthesample
(LR/MR: 2 ml, HR: 0.2 ml).
Replacethecaptightly.Invertthevialgentlyseveraltimestomixthecontents(Thevialwillbecome
hotduringmixing!)anddigestthevialsfor120minutesinthereactoratatemperatureof150°C.
Removethevialsfromthereactorandallowthemtocooldownto60°Corless.Mixthecontentsby
invertingeachvialseveraltimeswhilestillwarm.Thenallowthevialstocooltoambienttemperature
beforemeasuring.
Guidelines for photometric measurements
1. Runsamplesandblankswiththesamebatchofvials.
Theblankisstablewhenstoredinthedarkandcanbeusedforfurthermeasurementswithvials
fromthesamebatch.
2. Don‘tplacehotvialsintheadapter.Allowthevialstocooltoroomtemperaturefor
a minimum 45 minutes. It is recommended to leave the vials to cool over night.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is important to place
thevialscarefullyintheadapter.Theprecipitantatthebottomofthesampleshouldnotbe
suspended.
4. Cleantheoutsideofthevialswithatoweltoremovengerprintsorothermarks
.
5. Avoidspillageofwaterorreagentsolutionintothesamplechamberbecausethiscanleadtoincor-
rect test results.
6. Contaminationofthetransparentcellchambercanresultinincorrectreadings.Checkatregularin-
tervalsand–ifnecessary–cleanthetransparentcellchamberusingamoistclothorcottonswab.
7. Largetemperaturedifferencesbetweentheinstrumentandtheenvironmentcanleadtoerrors–
e.g.duetotheformationofcondensationinthecellchamberoronthevial.
8. Toavoiderrorscausedbystraylightdonotusetheinstrumentinbrightsunlight.

7
ENGLISH
Measurement
Fixtheadapterfor16mmvialsonthesamplechamber.
Turntheunitonusingthe[ON/OFF]key.
Thedisplayshowsthefollowing:
Selecttherequiredtestusingthe[MODE]key.
Scroll Memory (SM)
Toavoidunnecessaryscrollingfortherequiredtestrange,theinstrument
memorizes the last rangeusedbeforebeingturnedoff.Whentheinstrumentis
turned on again, the scroll list comes up with the last used test range.
Thedisplayshowsthefollowing:
Placetheblankintheadapter(Note1–4)makingsurethatthe
marks are aligned.
Blanksarespeciallypreparedforeachindividualtestrange.
Pressthe[ZERO/TEST]key.
The“Range“symbolashesforapprox.8seconds.
Thedisplayshowsthefollowing:
Afterzeroingremovethevialfromtheadapter.
Placethesampleintheadapter(Note2–4)makingsurethatthe
marks are aligned.
Pressthe[ZERO/TEST]key.
The“Range“symbolashesforapprox.3seconds.
The result appears in the display.
LR-/MR-range: in mg/l
HR-range: in g/l
The result is saved automatically.
Tolerance:±3.5 %(fullscale)
Repeating the test:
Pressthe[ZERO/TEST]keyagain.
New zero calibration:
Pressthe[ZERO/TEST]keyfor2seconds.
RANGE
RANGE
On
Off
Zero
Test
Mode
RANGE
0.0.0
RESULT
RANGE
Zero
Test
Zero
Test
Zero
Test

8
Menu Options
Menu selections
Pressthe[MODE]keyandhold.
Turntheunitonusingthe[ON/OFF]key.
Allowthe3decimalpointstobedisplayedbeforereleasingthe[MODE]key.
The[!]keyallowsforselectionofthefollowingmenupoints:
diS recall stored data
Prt printing stored data
setting the date and time
usercalibration
Theselectedmenuisindicatedbyanarrowinthedisplay.
1diS – Recall of stored data
Afterconrmingtheselectionwiththe[MODE]keythephotometershowsthelast
16datasetsinthefollowingformat(automaticallyproceedsevery3secondsuntil
result is displayed):
Number nxx(xx:16...1)
Year YYYY (e.g. 2010)
Date mm.dd (monthmonth:dayday)
Time hh:mm (hourhour:minuteminute)
Range Range
Result x.xx
The[ZERO/TEST]keyrepeatsthecurrentdataset.
The[MODE]keyscrollsthroughallstoreddatasets.
Exitthemenubypressing[!]key.
On
Off
Mode
!
!
Time Cal
Store Date
2
3 4
Mode
Mode
Zero
Test
diS
1
Time Cal
Store Date
1
1
1
23
4

9
ENGLISH
E 132
1Prt – Transmitting stored data (to Printer or PC)
Note: To print data, or to transmit to a PC, the optional Series 900 Data Hub(Inf-
raredInterfaceModule)isrequired.
The Series 900 Data Hubandtheconnectedprinter/PCmustbeready.Press
the[MODE]keytostartthetransmitting,theinstrumentdisplays“PrtG”(Printing)
forapprox.1secondfollowedbythenumberoftherstdatasetanditstransmis-
sion.Alldatasetswillbetransmittedoneaftertheother.Afternishingtheinstru-
ment returns to test mode.
Theprintjobcanbecancelledbypressingthe[On/Off]key.Theinstrumentturns
off.
IftheinstrumentisnotabletocommunicatewiththeIRiM,atimeoutoccursafter
approx.2minutes.TheerrorE132isdisplayedforapprox.4seconds.Subse-
quently, the instrument returns to test mode (see also IRiM manual).
2Setting date and time (24-hour-format)
Afterconrmingtheselectionwiththe[MODE]keythevaluetobeeditedwillbe
shownfor2sec.
Thesettingstartswiththeyear(YYYY)followedbytheactualvaluetobeedited.
Thesameappliesformonth(mm),day(dd),hour(hh)andminutes(mm).Setthe
minutesrstinstepsof10,pressthe[!]keytocontinuesettingtheminutesin
stepsof1.
Increasethevaluebypressingthe[MODE]key.
Decreasethevaluebypressing[ZERO/TEST]key.
Proceedtothenextvaluetobeeditedbypressing[!]key.
Aftersettingtheminutesandpressingthe[!]keythedisplaywillshow“ISSET”
and the instrument returns to the measurement mode.
On
Off
PrtG
Prt
1
Store Date
Time Cal
Mode
!
SET
DATE
YYYY
(2 sec.)
Mode
Zero
Test
Time Cal
Store Date
2
3

10
Calibration Mode
4User calibration (cAL)
Afterconrmingtheselectionwiththe[MODE]keytheinstrumentwill
show CAL/LR.
Scrollthroughrangesusingthe[MODE]key.
Placetheblankintheadaptermakingsurethatthemarks are aligned.
Pressthe[ZERO/TEST]key.
The“Range“symbolashesforapprox.8seconds.
Thedisplayshowsthefollowinginalternatingmode:
Performmeasurementwithastandardofknownconcentrationasdescribed.
Pressthe[ZERO/TEST]key.
The“Range“symbolashesforapprox.3seconds.
The result is shown in the display, alternating with CAL.
Ifthereadingcorrespondswiththevalueofthecalibrationstandard(withinthe
speciedtolerance),exitcalibrationmodebypressingthe[ON/OFF]key.
Changing the displayed value:
Pressingthe[MODE]keyonceincreasesthedisplayedvalueby1digit.
Pressingthe[ZERO/TEST]keyoncedecreasesthedisplayedvalueby
1 digit.
Pressthecorrespondingkeyuntilthereadingequalsthevalueofthecalibration
standard.
Bypressingthe[ON/OFF]key,thenewcorrectionfactoriscalculatedandstored
intheusercalibrationsoftware.
Conrmationofcalibration(3seconds).
Note:TheinstrumentcalibrationintherangeMRisautomaticallytakenalsofor
HR.
usercalibration(Displayincalibrationmode)
factorycalibration(Displayincalibrationmode)
CAL
cAL
Time Cal
Store Date
4
Cal
CAL
LR
0.0.0
CAL
Zero
Test
Zero
Test
RANGE
RANGE
CAL
RESULT
+ x
CAL
Mode
On
Off
Zero
Test
: :
RESULT
Indice
Lingue:
Altri manuali Xylem Attrezzature da laboratorio

Xylem
Xylem SI Analytics TitroLine 5000 Manuale utente

Xylem
Xylem WTW Turb 750 IR/T Manuale utente

Xylem
Xylem SI Analytics TO 7280 Manuale utente

Xylem
Xylem SI Analytics TitroLine 6000 Manuale utente

Xylem
Xylem SI Analytics TITRONIC 500 Manuale utente

Xylem
Xylem TitroLine 7800 Manuale utente

Xylem
Xylem WTW OxiTop Box Manuale utente

Xylem
Xylem WTW Turb 750 IR/T Manuale utente
Manuali Attrezzature da laboratorio popolari di altre marche

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manuale utente

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manuale utente

NI
NI PXI-5422 Manuale

Collomix
Collomix Aqix Manuale utente

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manuale utente

Ocean Insight
Ocean Insight FLAME-NIR+ Manuale utente