YSI ProSample Series Manuale utente

ProSample User Manual
PORTABLE AUTOMATIC SAMPLERS
Item# 630123REF
Revision A
For Sales & Service Contact
2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242
1-800-833-7958
www.geotechenv.com

2
This page left intentionally blank

3
Table of Contents
Safety Information 5
Precautionary Symbols in the Manual 5
Precautionary Symbols on the Instrument 5
Intended Use and General Overview 6
Keypad Overview 7
LED Functionality 8
Package Contents 8
Unpacking 9
Charging the Battery 10
Initial Charge 10
Battery Installation 11
Subsequent Charges 12
No Mains (AC) Power at Sampling Site 12
Mains (AC) Power Available at Sampling Site 12
Installing at a Sampling Site 14
Suction Hose Connection, Positioning, and Length 14
Instrument Setup and Calibration 16
Installing the Peristaltic Pump Tube 16
Turning the ProSample On 17
Pump Calibration 17
Calibration Containers 17
Calibration Procedure 17
Calibration Check 18
After Calibration and/or Calibration Check 18
Installing Sample Bottle(s) 19
Remove the Top of the Housing 19
Prepare Sample Bottles and Install 19
Reattach the Top of the Housing 20
Installing a Distributor Arm 21
Configure Key Instrument Settings 22

4
Wiring and Connecting Analog, Digital I/O, and SDI-12 Devices 24
Flow Signal Cable Wiring 24
Calibrating the Analog Signal 25
SDI-12 Cable Wiring 26
SDI-12 with Flying Leads 26
SDI-12 with EXO3 Connector 26
Connection of the Flow Signal Cable and/or SDI-12 Cable 26
Programming 27
Sampling Scenario List and Descriptions 27
Stand-Alone Sampling 28
Analog Input Trigger 32
Analog Input Volume Modulation 38
Digital Pulse Input Trigger 42
Data Pulse Input Event 45
SDI-12 Parameter-Driven Sampling 48
Maintenance and Cleaning 54
Draining the Bottom Sampler Housing 54
Desiccant Replacement 54
Peristaltic Pump Tube Replacement 55
Cleaning 55
Changing the Fuse(s) 56
Long-Term Instrument Storage 56
Factory Reset 57
YSIConnect PC Software 58
Replacement Parts and Accessories 59
Warranty and Liability 66
Technical Support 66
Service Information 66
Appendix A: ProSample Menu Tree 67

5
Reference relevant section of this manual for operation and/or safety
information
Appears on the outside of the peristaltic pump cover. This cover
should not be open while the instrument is on, as the user can be
seriously injured if they are exposed to moving parts in the peristaltic
pump assembly.
Safety Information
Please read this entire manual before unpacking, setting up or operating this equipment. Pay attention to all
precautionary statements. Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this equipment in any
manner other than that specified in this manual.
A
A
NOTICE: The manufacturer is not responsible for any damages due to misapplication or misuse of this
product including, without limitation, direct, incidental and consequential damages, and disclaims
such damages to the full extent permitted under applicable law. The user is solely responsible to
identify critical application risks and install appropriate mechanisms to protect processes during a
possible equipment malfunction.
NOTICE: Indicates a situation which, if not avoided, may cause damage to the instrument
Precautionary Symbols in the Manual
NOTE: Information that requires special emphasis
WARNING: Indicates a potentially hazardous situation which could result in injury
The information contained in this manual is subject to change without notice.
Effort has been made to make the information in this manual complete, accurate, and current.
The manufacturer shall not be held responsible for errors or omissions in this manual.
Consult YSI.com for the most up-to-date version of this manual.
When viewing this document as an AdobeTM PDF, hovering
your cursor over certain phrases will bring up the finger-point
icon. Clicking elements of the Table of Contents, website URLs,
or references to certain sections will take you automatically
to those locations.
THIS IS AN INTERACTIVE DOCUMENT
Precautionary Symbols on the Instrument

6
Thank you for purchasing a YSI portable automatic sampler. Samplers in the ProSample series are designed to collect
field samples and to provide temporary storage before they are analyzed in a laboratory.
The ProSample portable automatic samplers are designed for use in surface water, stormwater, and wastewater
applications. These samplers can be used in a variety of scenarios, most of which are described in this manual. If
you have a question regarding the use of a sampler in your project, please contact YSI (YSI.com/contact-us) or an
authorized YSI distributor.
There are four samplers in the ProSample series. Content in this user manual applies to all ProSample models unless
otherwise noted. ProSample models include:
• ProSample P: full-size portable sampler
• ProSample P-12: full-size portable sampler with SDI-12 connectivity
• ProSample PM: mini portable sampler
• ProSample PM-12: mini portable sampler with SDI-12 connectivity
Composite sampling can be completed with all ProSample models. In this type of sampling, a single container is used
to collect multiple samples, resulting in a mixture of all samples collected.
Discrete sampling can be completed with the ProSample P and ProSample P-12. In this type of sampling, samples are
placed in different bottles. There are several bottle configurations available for the purpose of discrete sampling.
All ProSample models have the following features:
• Robust PE, double-walled, insulated housing for temperature control.
• Easy tube replacement for minimal downtime.
• Spring-loaded roller bearings in the peristaltic pump providing long tubing life.
• Highly accurate sample volume using two capacitive sensors in the peristaltic pump for volume control.
• Simple programming and calibration.
• Long battery life.
• Non-volatile data memory for up to 5 years, so you never have to worry about losing your data if power is lost.
Intended Use and General Overview
Intended Use and General Overview

7
Keypad Overview
ProSample Keypad
Select the desired menu.
Confirmation of entered values and selections. A confirmed
selection is indicated with .
Enter Key
Move within the menu.
Move to the next menu.
Entry/change of values.
Arrow Keys
Selection within the menu.
Scrolling within the data memory or bottle memory.
Entry/change of values.
Arrow Keys
Return to higher menu level.
The activity is canceled.
Back Key
Terminate sleep mode.
Back Key
Press for at least 5
sec
Intended Use and General Overview

8
LED Functionality
There is an LED on the front of each ProSample that functions as follows:
• OFF: Device power is off, battery is dead, or no program is running
• Solid Green: Program is running and there are no errors
• Blinking Green: Program is running, but the battery is getting low
• Solid Red: An error has occurred. The display should be viewed for more information regarding the cause
of the error
• Blinking Red: Battery capacity is critically low
Included with each ProSample:
• *5 meter suction hose with screw connection and sinker weight
• 10L PE container (only the ProSample PM and the ProSample PM-12)
• Battery
• USB cable
• Two peristaltic pump tubes
• Printed copy of the ProSample Getting Started Guide
*Up to a 30 meter suction hose can be used with any ProSample model.
Please note the following items are not included with any ProSample. These items must be purchased separately.
• Bottle sets and distributor arms for the ProSample P and ProSample P-12
• SDI-12 cable for the ProSample P-12 and ProSample PM-12
• Flow signal cable
• Battery charger
• Other accessories such as strainers
Please see the Replacement Parts and Accessories section on page 60 for more details and item numbers, including
longer suction hose lengths.
The ProSample User Manual, installation file for YSIConnect, and the ProSample Ordering Guide can be found at
YSI.com/ProSample.
Package Contents
Intended Use and General Overview / Package Contents

9
Carefully unpack the unit and accessories, and inspect for shipping damages. Compare received parts with materials
listed in the Package Contents section. Notify YSI immediately of any damage or missing parts. Save all packing
materials until satisfactory operation is confirmed.
Unpacking
Removing the top tray (1) and the ProSample (2) from the packaging
Unpacking

10
Charging the Battery
Initial Charge
The battery must be completely charged prior to first use. This should take approximately 14 to 16 hours.
NOTICE: A total discharge of the battery can result in damage to the battery. A protective mechanism built-in
which automatically switches off the device when the voltage is too low (9V or less). However, when
storing the battery long-term, regularly connect the battery to the charger to top off the charge, as this
will ensure the battery does not completely discharge.
NOTE: Please include the battery if the ProSample is returned to YSI for service — see the Technical Support
section on page 66 and the Service Information section on page 66.
To charge, connect a ProSample battery charger to the battery. Once the red light on the power supply turns to green,
the battery is fully charged and can be installed in the ProSample unit.
NOTE: There are three different battery chargers available for the ProSample. Please see the Replacement Parts
and Accessories section on page 60 for a list of available chargers.
Connection of the ProSample battery to the battery charger
Charging the Battery
Questo manuale è adatto per i seguenti modelli
4
Indice
Altri manuali YSI Attrezzature da laboratorio
Manuali Attrezzature da laboratorio popolari di altre marche

Agilent Technologies
Agilent Technologies 5800 ICP-OES Manuale utente

Endress+Hauser
Endress+Hauser Cleanfit CPA875 Manuale utente

NI
NI PXI-5422 Manuale

Collomix
Collomix Aqix Manuale utente

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series Manuale utente

Ocean Insight
Ocean Insight FLAME-NIR+ Manuale utente