10
8 USB Data Transfer
Measurement results stored in memory may be transferred to a personal computer using
the USB port. Pull out the USB plug and use the supplied cable. Software required for
data transfer will be available on our web page at http://agratronix.com.
9 Cleaning and Maintenance
Store the tester in a dry place, protected from direct sunlight. Remove the battery
before long term storage. There are no user serviceable parts inside the device.
Do not remove the screws and open the unit.
MOIST.OFFSET The moisture display may be adjusted to match an oven
test or another reference method using the moisture offset
setting.
Once set, the offset correction will be applied to all moisture
measurements of the same type (e.g. all hay bales). The ad-
justment will not affect results previously stored to memory.
Use the arrow keys u or dto adjust the offset. You may
clear the offset by pressing the arrow keys u and dsimul-
taneously. Conrm your adjustment by pressing the e key
under the displayed symbol.
As a reminder, the offset will be displayed above the mois-
ture result on the moisture measurement screen in small
letters (e.g. OFFSET +0.6%). The offset will also be shown
upon bale selection.
NOTE: Offset correction will only work well fairly close to the
moisture where it was originally set. Therefore, if you apply
offset correction for wet haylage, please clear the offset
before measuring dry hay.
T-COMP. Set bale temperature for use in temperature compensation
of moisture measurement. See chapter 6.2.
+SETTINGS Enter the settings menu. See chapter 5
CLEAR AVG/HI Clears average and high moisture estimate for the currently
selected bale. All measurement results remain in memory.
Therefore, if you later increase averaging length in the set-
tings menu (see chapter 5), previous results will be included
in calculation again. If innite (∞) averaging length has been
selected, this function will not have an effect.
CLEAR BALE Clears the bale type of the currently selected bale and
erases all measurement results stored for this bale.
CLEAR ALL Clears the entire measurement memory. All moisture and
temperature results and bale type selections will be erased.