Brightwin Technology BRT ETX-1815 Manuale utente

BRT ETX-1815 Current Voltage Calibrator
Users Manual
Shenzhen Brightwin Technology Co., Ltd.

Contents
1 Basic Introduction........................................................................................................ 1
1.1 Function............................................................................................................. 1
1.2 Summary of Source and Measure Functions...............................................1
1.3 Interface (terminal) Description...................................................................... 2
1.4 Key Description.................................................................................................3
1.5 Display Screen..................................................................................................4
2 Basic Operation............................................................................................................5
2.1 Measure and Source........................................................................................5
2.2 Shut Down Mode.............................................................................................. 7
2.3 Backlight Brightness Adjustment....................................................................8
3 Function Usage............................................................................................................ 9
3.1 DC V and DC mV Measurement....................................................................9
3.2 DC mA measurement.................................................................................... 10
3.3 Current Measurement with Loop Power..................................................... 11
3.4 Frequency measurement.............................................................................. 11
3.5 DC V Source................................................................................................... 12
3.6 DC mV Source................................................................................................12
3.7 DC mA Source (active)............................................................................12
3.8 Simulating a 4- to 20-mA Transmitter......................................................... 13
3.9 Frequency Output...........................................................................................14
4 Advanced Application................................................................................................15
4.1 Setting 0 % and 100 % output parameters................................................ 15
4.2 Automatic Ramp the Output......................................................................... 16
4.3 Factory Reset..................................................................................................16
5 Power...........................................................................................................................17
5.1 Charge............................................................................................................. 17
6 Specifications............................................................................................................. 18
6.1 DC Voltage Measurement.............................................................................18
6.2 DC voltage Source.........................................................................................18

6.3 DC mA Measurement.................................................................................... 19
6.4 DC mA Source................................................................................................19
6.5 Frequency Measurement.............................................................................. 19
6.6 Frequency Source..........................................................................................20
7 Product Accessories..................................................................................................21
7.1 Standard Accessories....................................................................................21
7.2 Optional accessories..................................................................................... 21
8 Warning....................................................................................................................... 22

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 1
1 Basic Introduction
1.1 Function
Source of voltage, current and frequency.
Measure of voltage, current, loop current and frequency.
Manual stepping and automatic stepping and ramping.
Support for PC communication
1.2 Summary of Source and Measure
Functions
Function
Measure
Source
DC V
0~30 V
0~10 V
DC mV
0~100 mV
0~100 mV
DC mA
0~24 mA
0~24 mA
Frequency
1.000Hz~99.999kHz
0.00Hz~20.000kHz
Others
24V power supply, Step, Ramp.

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 2
1.3 Interface (terminal) Description
Figure 1.3-1
No.
Name
Description
①
Communication and
charging connector
Connect power adaptor to charge batteries
or connect the calibrator to the computer.
②
Current terminal
terminal for current measure and source.
③
COM terminal
All measure and source public terminal.
④
V, mV, Hz terminal
terminal for voltage, milli volt, frequency
measure and source.
⑤
24V terminal
24V power supply terminal, used in LOOP
mode

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 3
1.4 Key Description
Figure 1.4-1
No.
Description
①
Turns the power on or off
②
Selects the measurement mode
③
Selects output and analog transmitter mode.
④
Enables backlight switch display during start, enters backlight
brightness control mode.
⑤
Selects current or loop function

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 4
⑥
Sets and saves calibrator parameters setting
⑦
Recovers factory default setting
⑧
Sets manual output
⑨
Cycles through:
slow repeating 0%-100%-0% ramp
Fast repeating 0 % - 100 % - 0 % ramp
Repeating 0 % - 100 % - 0 % ramp in 25 % steps
⑩
Set output by 0% of span, Press and hold to store the source
value as the 0 % value
⑾
Decrements output by 25% of span.
⑿
Increments output by 25% of span.
⒀
Sets output by 100% of span, Press and hold to store the
source value as the 100 % value.
⒁
Selects DC V or DC mV function
⒂
Selects frequency function
1.5 Display Screen
Figure 1.5-1

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 5
2 Basic Operation
2.1 Measure and Source
This section acquaints you with some basic operations of ETX-2015/
ETX-1815.
Proceed as follows:
1. The connection of the calibrator as shown in Figure 2.1-1.
Figure 2.1-1

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 6
2. Press more than 2 seconds to turn on the calibrator. The calibrator
checks itself, including check on internal circuit and LCD, during which,
LCD displays all contents for 1s as shown in Figure 2.1-2:
Figure 2.1-2
3. Then the product model and automatic shutdown time(30 min)will be
displayed for 2 seconds as shown in Figure 2.1-3.
Figure 2.1-3
4. Press switch to the voltage mode.
5. Press to enter into output mode selection as shown in Figure
2.1-4.
Figure 2.1-4
6. Press and to increase or decrease 1 of the horizontal line
position (the number automatic carry but the position of the line have no

BRIGHTWIN BRT ETX-1815 CURRENT VOLTAGE CALIBRATOR
www.brightwinelectronics.com 7
change); press or to select a digit to change.
7. Press to select 1 V for the output value, and then press and
hold until the buzzer works to enter 1V as the 0% value.
8. Press to select 5 V for the output value, and then press and hold
until the buzzer work to enter 5V as the 100% value.
9. Press or to step between 0 and 100% in 25% step
increments. The screen will display as shown in Figure 2.1-5.
Figure 2.1-5
2.2 Shut Down Mode
The calibrator comes with the shutdown mode enabled for a time duration set to
30 minutes (displayed for about 2 seconds when the calibrator isinitially turned
on). When the shutdown mode is enabled, the calibrator will automatically
shutdown after the time duration has elapsed from the time the last key was
pressed. To disable the shutdown mode, press and simultaneously. To
enable the mode, press and simultaneously. To adjust the time duration,
press and simultaneously, the screen will display as shown in Figure
2.2-1,then press and/or to adjust the time between 1 and 30 minutes
and then press to store the new time duration (Without pressing any key
for 5 seconds, the calibrator will quit from the adjustment automatically).
Indice
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione