clare HAL Series Manuale utente

HAL Series
Electrical Safety Tester
Quick Start Guide
Seaward Electronic Ltds Ltd.
Bracken Hill, South West Industrial Estate,
Peterlee, Co. Durham. S 8 2SW. England.
Tel: +44 (0)191 586 3511 Fax: +44 (0)191 586 0227
www.seaward.co.uk

HAL Series
Quick Start Guide
2
Limited Warranty & Limitation of Liability
Seaward Electronic Limited guarantees this product for a
period of 1 year. The period of warranty will be effective at the
day of delivery.
© Copyright 2013
All rights reserved. Nothing from this edition may be
multiplied, or made public in any form or manner, either
electronically, mechanically, by photocopying, recording, or in
any manner, without prior written consent from CLA E. This
also applies to accompanying drawings and diagrams.
Due to a policy of continuous development CLA E reserves
the right to alter the equipment specification and description
outlined in this publication without prior notice and no part of
this publication shall be deemed to be part of any contract for
the equipment unless specifically referred to as an inclusion
within such contract.

HAL Series
Quick Start Guide
3
1.
Important Info
The aim of this quick start guide is to provide simple set up
options in order to configure and use the HAL. It is not an
exhaustive instructional document, for full details please refer
to the supplied instruction manual.
The information contained in this document assumes that all
relevant risk assessments and safety steps have been
performed prior to operation and that all users of the
equipment have been appropriately trained.
2.
Registering your HAL Tester
Seaward Group guarantees this product for a period of one
year, upon successful registration of the product. The period of
warranty will be effective at the day of delivery.
In order to ensure the continued performance of this product
Seaward recommends that this tester be serviced and
calibrated on an annual basis, by Seaward Group or any of
Seaward’s authorised service facilities using Seaward
approved parts and components.
Any unauthorised modifications, tampering or physical damage
sustained through negligent use or handling will void the
warranty.
3.
HAL Functions
There are 5 variants of the HAL, please refer to the table below
to view the functions available on your model
AC &
DC
Hipot I
Earth
Bond
Load
5kVA &
Leakage
Load
1000VA &
Leakage
ear
Test
Port
Multiple
Channel
Scanner
HAL101 x x
HAL102 x x x
HAL103 x x x x
HAL104 x x x x x
HAL
LED x x x x x

HAL Series
Quick Start Guide
4
4.
What s in the Box
HAL101 HAL102 HAL103
HAL
104
HAL
LED
Test Instrument 1 1 1 1 1
Power Supply Lead 1 1 1 1 1
Hipot Clip & Lead 1 5 Pairs 1 1 1
Hipot Probe & Lead 1 1 1 1
Earth Bond Probe 1 1 1
Test Output Box 1 1 1
Guard Plug 1 1 1 1 1
External Control
Connectors and Pins 2 2 2 2
5.
HAL Tour
All HAL models have the same front end, however the
connections on the ear of the HAL will differ dependant upon
the model. Please see below for your model and the
connections / options available
Front Panel – All Models

HAL Series
Quick Start Guide
5
•Visual Display – Main Interface Screen
•Pass / Fail Indicators – Indicate test status during and after a
test
•On / Off Button – Supplies Mains Power to operate the HAL
•Function Keys –Short cut options, function will differ
dependant upon current screen
•Start Button – Used to confirm selections and Start Tests
•eset Button – Used to abort selections and Stop Tests
•otary Encoder – Used to scroll Menu Options and Alter Test
Parameters
•Hi Tension Port Positive – Used for the Output of Test
Voltages
•Hi Tension Port Negative – Used as a return to measure any
current flow
•Hi Tension Presence Lamp – Used to indicate the HAL is
performing a High Voltage Test
Rear Panel – HAL101
1. Beacon Port – For connection to a safety status beacon (Part
Number H-5027)
2. Communications Port – For communication with the HAL, used
for download of data and remote control
3. Barcode Port – For connection to an S-232 Barcode Scanner to
allow input of Product ID’s and test sequence selection (Part
Number 194A922)
4. Printer Port – For connection to a printer for output of data
results. Can be used with the Desk Test n Tag printer for Pass /
Fail labels (Part Number 312A912)
5. Cooling Fan – Allow airflow to the internal circuitry, please do not
cover.

HAL Series
Quick Start Guide
6
6. External Fuse – User replaceable
7. Guard Switch Socket – Used as a safety interlock, can be used
as a start condition but will inhibit the test if the guard is opened.
Should be a zero volt connection.
8. emote Start Socket – Allows for remote start / stop switches to
be connected to the HAL, These will function as per the Green
and ed buttons on the front panel of the HAL. Should be a zero
volt connection
9. Status Indicator Socket – Outputs a signal to indicate Pass or
Fail condition, which could be utilised by a PLC for example. The
status connections are unpowered and require an external
supply voltage to operate. See manual for full details
10. Mains Supply Socket – For connection to the mains supply to
power the HAL101.
11. Mains Voltage Selector Switch – Sets the Mains operating
voltage for the HAL, switches between 115V and 230V. Please
ensure the selector switch is in the correct position prior to
powering on the HAL.
Rear Panel – HAL102
1. Beacon Port – For connection to a safety status beacon (Part
Number H-5027)
2. Communications Port – For communication with the HAL, used
for download of data and remote control
3. Barcode Port – For connection to an S-232 Barcode Scanner to
allow input of Product ID’s and test sequence selection (Part
Number 194A922)
4. Printer Port – For connection to a printer for output of data
results. Can be used with the Desk Test n Tag printer for Pass /
Fail labels (Part Number 312A912)

HAL Series
Quick Start Guide
7
5. Cooling Fan – Allow airflow to the internal circuitry, please do not
cover.
6. External Fuse – User eplaceable
7. Guard Switch Socket – Used as a safety interlock, can be used
as a start condition but will inhibit the test if the guard is opened.
Should be a zero volt connection.
8. Test Voltage Output Ports – 4 Pairs of Test Output ports, totalling
5 sets of ports including the front ports.
9. Mains Supply Socket – For connection to the mains supply to
power the HAL102.
10. Mains Voltage Selector Switch – Sets the Mains operating
voltage for the HAL, switches between 115V and 230V. Please
ensure the selector switch is in the correct position prior to
powering on the HAL
Rear Panel – HAL103
1. Beacon Port – For connection to a safety status beacon (Part
Number H-5027)
2. Communications Port – For communication with the HAL, used
for download of data and remote control
3. Barcode Port – For connection to an S-232 Barcode Scanner to
allow input of Product ID’s and test sequence selection (Part
Number 194A922)
4. Printer Port – For connection to a printer for output of data
results. Can be used with the Desk Test n Tag printer for Pass /
Fail labels (Part Number 312A912)
5. Cooling Fan – Allow airflow to the internal circuitry, please do not
cover.
6. External Fuse – User eplaceable

HAL Series
Quick Start Guide
8
7. Guard Switch Socket – Used as a safety interlock, can be used
as a start condition but will inhibit the test if the guard is opened.
Should be a zero volt connection.
8. ear Test Output connection – Test outputs for rear port,
connection point for the supplied Output box.
9. emote Start Socket – Allows for remote start / stop switches to
be connected to the HAL, These will function as per the Green
and ed buttons on the front panel of the HAL. Should be a zero
volt connection
10. Status Indicator Socket – Outputs a signal to indicate Pass or
Fail condition, which could be utilised by a PLC for example. The
status connections are unpowered and require an external
supply voltage to operate. See manual for full details
11. Mains Supply Socket – For connection to the mains supply to
power the HAL103.
12. Mains Voltage Selector Switch – Sets the Mains operating
voltage for the HAL, switches between 115V and 230V. Please
ensure the selector switch is in the correct position prior to
powering on the HAL.
Rear Panel – HAL 104 & LED
1. Beacon Port – For connection to a safety status beacon (Part
Number H-5027)
2. Communications Port – For communication with the HAL, used
for download of data and remote control
3. Barcode Port – For connection to an S-232 Barcode Scanner to
allow input of Product ID’s and test sequence selection (Part
Number 194A922)

HAL Series
Quick Start Guide
9
4.
Printer Port – For connection to a printer for output of data
results. Can be used with the Desk Test n Tag printer for Pass /
Fail labels (Part Number 312A912)
5.
Cooling Fan – Allow airflow to the internal circuitry, please do not
cover.
6.
Mains Voltage Selector Switch – Sets the Mains operating
voltage for the HAL, switches between 115V and 230V. Please
ensure the selector switch is in the correct position prior to
powering on the HAL.
7.
External Fuse – User eplaceable
8.
Guard Switch Socket – Used as a safety interlock, can be used
as a start condition but will inhibit the test if the guard is opened.
Should be a zero volt connection.
9.
External Fuse – User eplaceable
10.
ear Test Output connection – Test outputs for rear port,
connection point for the supplied Output box.
11.
Digital Output Socket –Allows for connection to the 8 Digital
Outputs, which can be used for relay control, as an example.
Please refer the full manual for pin out information
12.
Analogue Input Socket - Status Indicator Socket – Outputs a
signal to indicate Pass or Fail condition, which could be utilised
by a PLC for example. The status connections are unpowered
and require an external supply voltage to operate. See manual
for full details
13.
External Fuse – User eplaceable
14.
Mains Supply Socket – For connection to the mains supply to
power the HAL101.
15.
Power Test Voltage Selector Switch – Select between Norm, and
Aux. Norm will use the incoming mains supply powering the HAL
and route it through to the output box to supply power to the
EUT. Aux will route through an external voltage supplied via the
Auxiliary Voltage input ports.
16.
Auxiliary voltage Inlet – 2 x 4mm sockets to allow input of
external Aux voltage supply for the Power & Leakage tests.
Supplied voltage can be 50/60Hz and between 80 –
250
VAC
6.
Navigating the HAL Menu
The HAL Menu System is driven primarily using the otary
Encoder and Green (STA T) and ed ( ESET) buttons.

HAL Series
Quick Start Guide
10
The function of the rotary encoder is to navigate through the
HAL menu system. otating the wheel clockwise will move the
cursor forward or increase Test Parameters, where as rotating
the wheel anti clockwise, will move the cursor backwards or
reduce Test Parameters
The Start or Green Button has the primary function of starting a
Test, however when configuring the HAL it is also used to cycle
through available options, for example when configuring the
system, the memory can be turned off. When this option is
navigated to and highlighted, pressing the green button will
cycle between ON and OFF.
The ed or Abort Button has the primary function of aborting a
Test in progress, but it can also be used to escape from a
particular screen and return back to the previous menu
The Four function keys directly beneath the LCD Display of the
HAL are generally used as short cuts to useful options within a
particular screen. These keys serve as single press options to
save time navigating through the HAL menu structure.
TIP. Where a function key displays OTATE FIELD, selecting
this option will allow the rotary encoder to be used to increase
test parameters. This is particularly useful when there are many
options available, such as selecting a test voltage.
7.
System Configuration
The System Configuration menu is used to define how the HAL
behaves when performing tests. This section is shown using a
HAL104 menu screen. Some of the options shown may not be
available on other models due to their functionality.
System Configuration can be found by navigating from the
MAIN MENU to SETUP > SYSTEM CONFIGU ATION.
Questo manuale è adatto per i seguenti modelli
5
Indice
Altri manuali clare Apparecchiature di prova
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione