2. Within Run Precision - Precision studies with the STAT-
Site®MHgb method using whole blood samples yielded the
following mean values, standard deviations (SD), and coeffi-
cients of variation (CV%). Values are for testing of 20
replicates on each of four meters (n=80).
Whole Blood Sample Mean SD CV% n
8.8 g/dL 9.1 0.3 3.5 80
14.3 g/dL 14.1 0.7 4.7 80
17.1 g/dL 16.9 0.8 4.9 80
PERFORMANCE CHARACTERISTICS
3. Total Precision –
Total precision studies were conducted with four instruments
over 20 days using 2 levels of a whole blood control solution
per NCCLS EP5-T2 guidelines. A total of 80 tests were run on
each instrument. The resulting total precision estimate ranges
over four meters are presented in the following table.
Sample Mean SD Range CV% Range n
Low Control 9.9 0.28 - 0.41 2.9 - 4.2 80
Normal Control 13.0 0.37 - 0.55 2.9 - 4.2 80
www.CLIAwaived.com
TEL: Toll Free 1-888-882-7739
4. Accuracy -
The STAT-Site M Hgb system was evaluated at three clinical sites
with a patient population consisting of adults, children, and
infants.
The correlation obtained between STAT-Site M Hgb system
results and the reference method for venous samples was:
N = 103; y = 1.00x – 0.84; R = 0.96.
The correlation obtained between STAT-Site M Hgb system
results for capillary samples (transfer tube and directly applied
samples used) and the reference method (venous for adults,
capillary samples for children and infants) was: N = 236; y =
0.98X + 0.16; R = 0.93.
5. Interferences -
Triglycerides (1,005 mg/dL) and bilirubin (20 mg/dL) do not
interfere with the STAT-Site®MHgb Test.
•If the patient is experiencing symptoms which are not consistent
with the hemoglobin results obtainedAND you have eliminated
common procedural errors (described in the STAT-Site
®
M
Hgb
Meter
User’s Guide) as the cause, follow your facility’s policies for
treating the symptoms and confirm the blood hemoglobin results
with another laboratory method.
•Never make significant changes to the patient’s medication program
or ignore physical symptoms without consulting a physician.
INTENDED USE
The STAT-Site®MHgb Test Kit is intended for the
quantitative determination of hemoglobin in whole blood using
the STAT-Site®MHgb Meter. The STAT-Site®MHgb Test
may be used with adults, infants, and children in a physician’s
office or other professional point-of-care setting.
SUMMARY AND PRINCIPLE
Hemoglobin is the oxygen-carrying pigment and main component of red
blood cells. Low hemoglobin levels may indicate anemia, recent
hemorrhage or fluid retention. Elevated hemoglobin levels may indicate
hemoconcentration from polycythemia or dehydration.
The STAT-Site
®
M
Hgb
Test provides a direct reading of hemoglobin
concentration in whole blood between 6 and 21 g/dL. Values below or
above this range will be reported as <Lo> or <Hi> respectively.
The STAT-Site
®
M
Hgb
Test consists of a plastic card with reagent pads*
for determining the concentration of hemoglobin. When a drop of
whole blood is applied to the top of the
STAT-Site®MHgb
Test Card,
hemolysis occurs, with release of hemoglobin. Sodium nitrite converts
the hemoglobin to methemoglobin. Sodium azide then reacts with
methemoglobin to form azide-methemoglobin, which is brown in color
and is detected at 565 nm with a small portable reflectance analyzer.
The amount of the color produced due to azide-methemoglobin is
proportional to the concentration of hemoglobin in the sample.
1
REAGENTS
Reagents used in STAT-Site®MHgb Test Cards contain the following
ingredients:
Sodium azide 0.6% w/w
Sodium nitrite 0.8% w/w
Inactive ingredients: 98.6% w/w
PRECAUTIONS
•For in vitro diagnostic use.
•Do not use with serum or plasma.
•Fetal, newborn, or variant hemoglobin samples have not been
evaluated with STAT-Site
®
M
Hgb
.
Children as young as 3 weeks
old were tested and included in the study.
•As with all chemical reagents, contact with the skin should be
avoided with the reactive areas of the Card.
•Handle blood specimens as potentially infectious samples and
follow the guidelines established by the Centers for Disease Control
(CDC) Atlanta, GA, for blood collection and handling (Document
20 CFR 1910.1030).
STORAGE AND STABILITY
The container of Test Cards can be stored at or below room tempera-
ture (28°C/82°F) until the expiration date. This product can be stored in
the refrigerator. If stored refrigerated, it is important to bring the
package to room temperature before opening and removing Test Cards
for testing.
The desiccant included with the Test Cards is not part of the test. It is
included only to keep the Test Cards dry. To ensure the remaining Test
Cards in the container are kept dry, keep the desiccant inside the
container and reseal immediately after removing the needed Test
Card.
Write the date opened on the container label where indi-
cated. Once you open the container, Test Cards must be used
within 90 days.
Reseal the container immediately after removing a Test
Card. Test Cards should remain in the resealed con-
tainer, with the dessicant, until being removed for use.
Avoid contact with the reagent pads on either side of the Test Card at
all times.
Each box of STAT-Site
®
M
Hgb
Test Cards comes with one CODE Key
that must be inserted into the STAT-Site
®
M
Hgb
Meter before the test can
be run. The CODE Key and Test Cards are matched for product type
and CODE number and are intended to be used with the Test Cards from
the same box . The CODE Key contains electronic information . Handle
with care and keep clean.
Dispose of the CODE Key after using the last Test Card from the
kit.
SPECIMEN COLLECTION AND PREPARATION
To perform a blood hemoglobin test with
STAT-Site®MHgb
Test Cards
on the STAT-Site
®
M
Hgb
Meter you will need a drop (approximately 12
µL) of whole blood. Follow NCCLS Guideline H4A4 for obtaining a
capillary blood sample.
Capillary blood can be obtained from a skin puncture. The puncture site
should be cleaned and dried before pricking the site. Wipe away the
first drop with a gauze pad. Allow a large drop to form at the puncture
site. Avoid “milking” the finger to improve blood flow.
If using venous whole blood, collection tubes containing EDTAor
Heparin as anticoagulants are recommended. Do not use blood
collection tubes containing Sodium Fluoride or Oxalate/Fluoride.
Refrigerated blood should be allowed to reach room temperature
before testing.
BIBLIOGRAPHY
1. Vanzetti Giulio. An azide-methemoglobin method for hemoglobin determination in
blood. J Lab Clin Med 1966; 67: 116 - 26.
2. Fandek N, Moreau D, Newell K C, Ofner A, eds. Clinical Laboratory Tests - Values
and Implications. 2nd ed. Springhouse: Springhouse Corporation, 1995: 328pp.
3. DeMott Wayne R, Tilzer Lowell L, Hematology. In: Jacobs DS, DeMott WR,
Finley PR, Horvat RT, Kasten Jr BL, Tilzer LL, eds. Laboratory Test Handbook.
Hudson: Lexi-Comp, 1992: 517-626.
4. Wallach J. eds. Interpretation of Diagnostic Tests - A Synopsis of Laboratory
Medicine, 4th ed. Boston/Toronto: Little Brown and Co. 1986: 6pp.
5. Painter Pennell C, Cope June Y, Smith Jane L, Appendix. In: Burtis CA,
Ashwood ER, eds. Tietz Textbook of Clinical Chemistry. Philadelphia: WB Saunders,
1994: 2161-2217.
6. Tietz N, ed. Clinical Guide To Laboratory Tests, WB Saunders, 1983: 258-259.
For technical assistance with this product, in the U.S.,
please call our Technical Service Department at (800)
531-5535. Outside of the U.S., please call (830) 249-
0772 or FAX +(830) 249-0851.
TECHNICAL SERVICE
Reference No. Product Description
901025 STAT-Site®MHgb Test Cards (100 Test Cards,
(4 x 25/Container), and CODE Key)
503000 STAT-Site®Hemoglobin Controls
6 x 1.5 mL (3 Low and 3 High level)
900905 STAT-Site®MHgb Battery Pack
5 Batteries (approximately 5,000 tests)
900900 STAT-Site®MHgb Meter (Battery Included)
202012 12 µL Transfer Tubes (100 tubes/bag)
To order call (858) 481-5031
AVAILABILITY
LP DISCLAIMS ALL EXPRESS AND IMPLIED
WARRANTIES OF THE MERCHANTABILITY AND FITNESS PERTAINING
TO THIS PRODUCT WHICH ARE NOT EXPRESSLY DETAILED IN THIS
PACKAGING INFORMATION OR A WRITTEN AGREEMENT BETWEEN
THE BUYER AND SELLER OF THIS PRODUCT.
LP MAINTAINS THAT THIS PRODUCT
CONFORMS TO THE INFORMATION CONTAINED IN THIS INSERT.
PURCHASER MUST DETERMINE THE SUITABILITY OF THE PRODUCT
FOR ITS PARTICULAR USE. USE ONLY IN ACCORDANCE WITH
LABELING INSTRUCTIONS.
*
U.S. Patent 5,104,619
pg 4
1. Linearity - Linearity was established over the range 5.6 to
20.6 g/dL with venous blood using NCCLS EP6-P guidelines.
The linearity regression line is:
y = 1.03 + 0.08, R = 0.9968.
Stat-Site M HemoSite Linearity
5
7
9
11
13
15
17
19
21
23
5 7 9 11131517192123
Expected conc (g/dL)
Observed conc (g/dL)
Distributed by :