Montageanleitung
Notice de montage

English Page 3 024................................................
Deutsch Seite 25 046..............................................
Français Page 47 068..............................................

3
PW15AHi
A2526-4.0 en/de/fr HBM
Contents Page
English
Safety instructions 4..............................................
1 Application 6..................................................
2 Special features 7..............................................
3 Marking and variants 7.........................................
4 Mechanical construction and mounting instructions 8............
4.1 Design of the PW15AHi 8..................................
4.2 Before start‐up 8..........................................
4.3 Mounting instructions 9.....................................
5 Electrical configuration and function 10..........................
5.1 Function 10................................................
5.2 Signal conditioning 11.......................................
5.3 Trigger functions 12.........................................
5.4 Limit value function 13......................................
5.5 Extreme value function 13...................................
5.6 Dosing and filling control 13..................................
5.7 Diagnostic function 14.......................................
6 Electrical connection 15.........................................
6.1 PW15AHi pin assignments 15................................
6.2 Supply voltage 16..........................................
7 Interfaces 18...................................................
7.1 RS‐485 4-wire interfaces (UART) 18..........................
7.2 CANopen interface 19.......................................
7.3 DeviceNet interface 20......................................
7.4 Digital control input 22.......................................
8 Specifications 23...............................................

PW15AHi
4
A2526-4.0 en/de/frHBM
Safety instructions
The device must not be modified from the design or safety engineering point of
view except with our express agreement. Any modification shall exclude all
liability on our part for any damage resulting therefrom.
Repair is specifically forbidden. Repairs may only be carried out by HBM.
All the factory defaults are stored at the factory so that they are safe from
power failure and cannot be deleted or overwritten. They can be reset at any
time by using the command TDD0.
The production number set at the factory cannot be changed.
PW15AHi load cells are exclusively designed for weighing technology
measurement tasks and directly associated control and regulatory tasks.
Use for any purpose other than the above is deemed to be non-designated
use.
There are not normally any hazards associated with this product, provided
the notes and instructions for project planning, installation, appropriate
operation and maintenance are observed.
It is essential to comply with the safety and accident prevention regulations
specific to the particular application.
Installation and start‐up must only be carried out by suitably qualified
personnel.
During installation and when connecting the cables, take action to prevent
electrostatic discharge, as this may damage the electronics.
The required power supply is a (12 - 30 V) extra‐low voltage with safety
separation from the mains (see Specifications).
When connecting additional devices, comply with the safety requirements.
All the interconnecting cables must be shielded cables. The shield must be
connected extensively to ground on both sides.
For additional information, please refer to the Online Help on 1‐FIT‐AED‐DOC
(system CD).

5
PW15AHi
A2526-4.0 en/de/fr HBM
Residual dangers are indicated in these mounting instructions by the following
symbols:
Symbol: ATTENTION
Meaning: Possibly dangerous situation
Warns of a potentially dangerous situation in which failure to comply with
safety requirements could result in damage to property or some form of phys
ical injury.
Symbols for application instructions and useful information:
Symbol: NOTE
Means that important information about the product or its handling is being
provided.
Symbol:
Meaning: CE mark
The CE mark enables the manufacturer to guarantee that the product complies
with the requirements of the relevant EC directives (the Declaration of
Conformity can be found at http://www.hbm.com/HBMdoc).
Symbol:
Meaning: Statutory waste disposal mark
In accordance with national and local environmental protection and material
recovery and recycling regulations, old devices that can no longer be used
must be disposed of separately and not with normal household garbage.
If you need more information about waste disposal, please contact your local
authorities or the dealer from whom you purchased the product.

PW15AHi
6
A2526-4.0 en/de/frHBM
1 Application
PW15AHi load cells belong to the digital load cell and measuring chain family
specially developed by HBM for rapid dynamic weighing processes. They
acquire measurement signals from strain gages, condition them digitally, output
them and can network them for buses on request.
They supply a fully‐filtered, scaled and digitized output signal for direct
connection to a bus system or a PC. They operate at high data rates of up to
1200 measurements per second and can be easily and rapidly adapted to the
actual weighing system by means of adjustable parameters.
The integrated trigger function allows event‐driven determination of weight,
which greatly reduces the external software requirements for checkweigher
applications, for example.
The PW15AHi includes dosing control. As this load cell does not have any
digital outputs, the external Master control must make digital I/Os available.
The PW15AHi generates the information to control these digital outputs in the
respective measurement statuses (MSV measurement status). The Master
control then only needs to forward these control bits to its digital outputs to
drive the dosing control valves.
The load cell is supplied as required with an RS485, CANopen or DeviceNet
interface.
PW15AHi load cells are designed in stainless steel and have a high degree of
protection (IP 68).
The AED_Panel 32 PC software is available to facilitate parameter settings, to
display dynamic measurement signals and for comprehensive frequency
analysis of the dynamic system.
This part of the operating manual describes the hardware and the functions of
PW15AHi digital load cells. The commands for serial communication are
described in the online documentation (= AD103C commands).

7
PW15AHi
A2526-4.0 en/de/fr HBM
2 Special features
Rapid digital filtering and scaling of the measurement signal
Serial interfaces: RS‐485 4‐wire, CANopen, DeviceNet
All settings made via the serial interface
Power failsafe parameter storage
Indestructible storage of factory settings
Selection of output speed of measured values up to 1200 measured values/s
Four limit value functions
Extreme value functions (MIN/MAX)
Four trigger functions (internal level triggering, external triggering)
Dosing control
Diagnostic functions
Supply voltage 12 V ... 30 VDC
Degree of protection: IP 68
3 Marking and variants
The PW15AHi can be delivered with three different interface types.
The accuracy class is C3 (OIML).
Maximum capacity: 10 kg, 20 kg
Interface: B = RS‐485, C = CANopen, D = DeviceNet
PW15AHi product family (3rd generation)
Electrical connection: 0 = socket
Fig. 3.1 PW15AHi order reference, explanations: RS‐485 = RS‐485 4‐wire bus
The following preferred types are defined (other variants on request):
Interfaces
Max. capacity RS-485 CANopen DeviceNet
10 kg 1‐PW15AHi/3B0/10kg 1‐PW15AHi/3C0/10kg 1‐PW15AHi/3D0/10kg
20 kg 1‐PW15AHi/3B0/20kg 1‐PW15AHi/3C0/20kg 1‐PW15AHi/3D0/20kg
50 kg 1‐PW15AHi/3B0/50kg 1‐PW15AHi/3C0/50kg 1‐PW15AHi/3D0/50kg

PW15AHi
8
A2526-4.0 en/de/frHBM
4 Mechanical construction and mounting instructions
4.1 Design of the PW15AHi
PW15AHi digital load cells are characterized by a compact stainless steel
measuring body and are designed with degree of protection IP68. The
electronics are integrated in the measuring body.
==
==
25 27
6M6
10
12
18
190.2
150
15
12
6
M6x0.5 (1x)
12
12
40
12.5
M6 (5x)
190.2 190.2
20
Fig. 4.1 PW15AHi dimensions
4.2 Before start‐up
Take the following precautions into account during mounting and
operation:
The length of the fastening screws must be selected so that the maximum
screw‐in length of 10 mm is not exceeded. Non‐compliance can cause
damage to the load cell.
Avoid force shunts during setup.

9
PW15AHi
A2526-4.0 en/de/fr HBM
4.3 Mounting instructions
1: PW15AHi mounting at load application: 2 x M6; property class 10.9;
tightening torque 10 Nm; max. screw‐in depth 10 mm.
recommended flatness 0.1 mm on the connecting plate
2: PW15AHi mounting with mounting base: 2 x M6; property class 10.9:
tightening torque 10 Nm; max. screw‐in depth 10 mm
Mounting base
Recommended
flatness: 0.1
Platform center
Load direction
Platform
Load
application
(customer
design)
(1)
(2)
Fig. 4.2 PW15AHi mounting instructions
The PW15AHi load cell must be mounted on a clean surface with a flatness
better than 0.1 mm. Overload protection can be integrated in the form of a
M6x0.5 fine‐thread screw.
It is recommended that the load is applied in the center of the platform, to
minimize off‐center load errors and moments.
Before installing several PW15AHi into an installation with a bus system, take
the following into account:
The printed production number (typeplate) is required for setting up data
communication. If the typeplate can no longer be seen after installation, the
numbers of each PW15AHi should be noted beforehand. This enables different
addresses to be assigned during the initial start‐up.
Alternatively, before connection to the bus system, you can connect each
PW15AHi individually with a PC, in order to set different addresses (see ADR
command, Online Help).

PW15AHi
10
A2526-4.0 en/de/frHBM
5 Electrical configuration and function
The electronics of the PW15AHi digital load cell basically comprise the
following function groups:
Single‐point load cell
Amplifier
Analog/digital converter (A/D)
Evaluation unit (P)
Power failsafe parameter storage (EEPROM)
Serial interface (RS‐485, CANopen or DeviceNet)
Control input IN1 (Trigger/Stop Dosing)
5.1 Function
Interface
Supply voltage
UB / GND
Serial
interface
IN 1
Trigger
Stop Dosing
EEPROM
Linearization
Serial number
Digital filter
Data rate
Sensitivity
Zero setting
Voltage
controller
Trigger
Voltage
controller
Fig. 5.1 Block diagram
The analog transducer signal is first amplified, then filtered and converted to a
digital value in the analog/digital converter. The digitized signal is conditioned
in the microprocessor and forwarded via the serial interface. All the parameters
can be stored power failsafe in the EEPROM.
Indice
Lingue:
Altri manuali HBM Accessori

HBM
HBM U2A Manuale utente

HBM
HBM C2A/M1 Series Manuale utente

HBM
HBM WE2107 Manuale utente

HBM
HBM RTN/M2 Series Manuale utente

HBM
HBM FIT 5A Manuale utente

HBM
HBM PW15PH Series Manuale utente

HBM
HBM PW25 Manuale utente

HBM
HBM C16i series Manuale utente

HBM
HBM T10FM Manuale utente

HBM
HBM PW29P Series Manuale utente

HBM
HBM FS64 Manuale utente

HBM
HBM T40FH Manuale utente

HBM
HBM PW15AH Manuale utente

HBM
HBM T40B Manuale utente

HBM
HBM PW27A Manuale utente

HBM
HBM precix 6 PW6C Series Manuale utente

HBM
HBM PW29 Series Manuale utente

HBM
HBM FS63 series Manuale utente

HBM
HBM C16i series Manuale utente

HBM
HBM HLC Series Manuale utente