The figure opposite shows a typical pulse observed at the output of the id100-
MMF100 terminated with a 50fi load. A2V amplitude and 10ns width is available.
The recommended trigger level of the measurement device is 1V.
For "pure" counting applications, the trigger slope can be negative or positive.
For timing applications, the trigger slope must be positive in order to take full
advantage of the id100-MMF100 timing resolution. 1ons
,ê
MAINTENANCE
Maintain a temperature range within specifications when transporting or storing the unit. Transportation damage can occur
from improper handling. The following steps are recommended to minimize the possibility of damage:
- pack the unit in the original packing material when shipping.
- store unit at room temperature in a clean and dry area.
- avoid high humidity or large temperature fluctuations.
- keep the unit out of direct sunlight.
- avoid unnecessary shock and vibration.
The following safety precautions must be observed during the operation and servicing of the unit. Failure to comply with these
precautions or with specific indications elsewhere in this manual violates safety standards of intended use of the unit. lD
Quantique assumes no liabilityforthe user's failure to comply with these requirements.
- This unit is intended for indoor use only.
- Unit covers cannot be removed during operation.
- Any adjustments, maintenance, and repair of opened unit under voltage should be avoided and carried out only by
skilled personnel aware of the hazards involved. Do not attempt internal service or adjustment unless another person qualified
in first aid is present. Do not replace any components while power cable is connected.
- Operation of any electrical instrument around flammable gases orfumes constitutes a major safety hazard.
- Installation of replacement parts or modification of the unit should be carried out by authorized personnel only.
There are no user-serviceable components in the id100-MMF100. The id100-MMF100 has been designed to require minimum
maintenance and to provide reliable operation for many years to come.
To help ensure long, trouble-free operation,
- Keep the id100-MMF100 free of dust
- Do not spill liquids on or into the unit. lf the unit does get wet, turn off the power immediately and let the unit dry
completely.
To ensure optimum performance, the optical port should be kept as clean as possible. lt is recommended to connect before the
first use of the detector an optical fibre patchcord to the id100-MMF100 optical input port. This patchcord should then be used
for all connections of the module to a device under test" ln case of inappropriate insertion losses, the end connector of this
patchcord can be easily cleaned by the user.
Nevertheless, if the optical port needs to be cleaned,
1 .Remove the protective cap
2.lnsert a lint-free stick dipped in isopropyl alcohol and gently wipe the receptacle
3.Dry using a dry lint-free stick.
The patchcord fibre ends connected to the id100-MMF100 optical input port must be kept clean at alltimes to ensure minimum
loss and to reduce reflection.
LGently wipe the fibre end with a lint-free swab dipped in isopropyl alcohol (98% pure or better).
2.Dry using clean compressed air.