innovAg DairyTest Lite Manuale utente

© 2021 innovAg 1
DairyTest
Lite
Dairy Vacuum
& Pulsation Tester
USER MANUAL
Part No. DR51-0060-13
For software v2.4.1

2 © 2021 innovAg

© 2021 innovAg 3
Contents
1. What is DairyTest Lite? 6
2. Safety Precautions 6
3. Maintenance 6
4. Calibration & Vacuum System Check 7
5. The Basics 8
The Button 8
First start up 9
Setting up 10
Parlour Presets 11
Pulsation alarm limits 12
System Information 13
Force Zero 14
Starting a session 15
Vacuum measurement 17
Displaying vacuum measurement 18
Vacuum Recorder 19
Flushing 19
Displaying vacuum recordings 20
Pulsation Measurement 21
Regulation Characteristics 24
Displaying Regulation Characteristics 26
6. Generating a report 27
7. GeneratingaLogimatle(Frenchversiononly) 28
8. Screen capture 28
9. Leak Test 28
10. Response Test 29
11. FactorySettings 30
12. FeatureCodes 31
13. Updatingsoftware 31
14. Replacing the batteries 32
15. Filtertubeassembly 32
16. ExternalFilters 33
17. Repairs 33
18. Troubleshooting your DairyTest Lite 33
19. Specications 34
20. Warranty 36

4 © 2021 innovAg
FCCComplianceStatement
This device complies with Part 15 of the FCC rules. Operation is subject
to the following two conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including
interference that may cause undesired operation.
Declaration of Conformity
StandardstowhichConformityisDeclared:
• EN 61000-4-2
• EN 61000-4-3
• EN 61000-6-1
• EN 61000-6-2
• EN 61000-6-4
• IEC 61326:2013 (IEC 61326-1:2012 Ed2)
• EU Directive 2011/65/EU
• EU Directive 2014/30/EU
• 47 CFR Part 15B
• FCC 15.109
• ANSI c63.4:2014
• China RoHS (See following page)
Manufacturer’sName: innovAg Pty. Ltd.
Manufacturer’sAddress: 37/328 Reserve Road, Cheltenham
Victoria 3192, Australia
TypeofEquipment:Pulsation Tester/ Vacuum Gauge
BrandName: DairyTest Lite
ModelNumber: DR40-0177

© 2021 innovAg 5
I, the undersigned, hereby declare that the equipment specied above
conforms to the above Directives and Standards.
Braham Basser Date: 18 March 2018
Director
InnovAg Pty. Ltd
Enclosure
LCD
Silicone tubing
Metal ttings
Front panel label
Rating labels

6 © 2021 innovAg
1. What is DairyTest Lite?
DairyTest Lite is an easy-to-use update of the industry-standard
DairyTest Professional. New features include:
- Colour TFT touch screen.
-Built-in report, viewed and printed via PC / Mac / tablet without
software installation.
- Lightweight, rubberised, IP53 enclosure.
- Non-volatile data storage.
-
Valve purge system to protect the sensors from uid ingress.
- Designed, built and serviced in Australia.
WithinAustralia wehaveaFreecall helpline: 1800061 167.International
users should call their local distributor or contact us via email at
2. Safety Precautions
DairyTest Lite is intended to be used in the milking stall – always allow
for the unpredictable nature of animals. During the measurements,
sharp needles may be used – take care they do not hurt people or
animals. Always ask the dairy farmer about the behaviour of the
animals and milking method used.
If using DairyTest Lite in wet tests, always keep the unit above the test
point to prevent uid entry. Damage to the unit from uid entering the
sensors is not covered under the warranty. Note that the use of external
lters in DairyTest Lite measurement tubes is NOT RECOMMENDED.
See the ExternalFilters section.
3. Maintenance
Clean DairyTest Lite with a moist cloth. Do not use cleaning agents.
DairyTest Lite performs self-tests when turned on and during use. If
an error message is displayed or you suspect the unit is damaged,
contact innovAg or the local distributor for further information.

© 2021 innovAg 7
4. Calibration & Vacuum System Check
DairyTest Lite automatically performs a zero calibration when turned on.
Do not turn on your DairyTest Lite with vacuum connected. If the unit
reads other than zero on start-up, make sure vacuum is disconnected
and turn o and on again. If it still does not read zero a manual ‘Force-
zero’ can be used. Refer to the section ‘Troubleshooting your
DairyTest Lite’.
A complete calibration should be performed with an approved and
certied calibration test instrument every 12 months which can be
organised by innovAg or the local distributor. When calibration is due a
‘Needs calibration’ warning will appear on the startup screen.
The warning can be temporarily overridden by performing a Vacuum
System Check which requires a successful Response Test and Leak
Test to be performed on both ports.
Enter the Factory Settings screen and the Vacuum System Check
button will show the status of Leak Tests (L1 & L2) and Response Tests
(R1 & R2). The labels are grey if the test has not been performed or not
passed.
If all labels are green the tests have passed and pressing the button will
disable the ‘Needs calibration’ warning.

8 © 2021 innovAg
5. The Basics
The Button
There is only one button on the DairyTest Lite and it has three functions.
1. Press to turn the unit on.
2. From the sessions screen press and hold to turn o.
3. Press to exit the current screen.

© 2021 innovAg 9
Setup
Firststartup
When you rst turn on the DairyTest Lite the ‘needs calibration’ icon may
be shown indicating the date/time have not been set. Continue to the
Setup section to set the clock.

10 © 2021 innovAg
Backlight brightness
Time to half brightness
Time to sleep
Time to auto power o
Set date/format
Parlour Presets *
System information *
Vacuum units (kPa or inHg)
Pulsation units (% or ms)
Language
Setting up
* refer to following section
Indice
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione