Lauda LRZ 930 Manuale utente

Interface moduleLRZ930
Ethernet module Advanced
Operation manual
V3R14 Read this manual prior to performing any task!

replaces issue V1R12
Translation of the original operation manual
Manufacturer:
LAUDA DR. R. WOBSER GMBH & CO. KG
Laudaplatz 1
97922 Lauda-Königshofen
Germany
Telephone: +49 (0)9343 503-0
E-mail: [email protected]
Internet: https://www.lauda.de
Q4DA-E_13-024, 3, en_US 09/29/2023©LAUDA 2022
V3Interface module LRZ 9302 / 44

Table of contents
1 General................................................................................................................................................................................... 5
1.1 Intended use................................................................................................................................................................... 5
1.2 Compatibility.................................................................................................................................................................. 6
1.3 Technical changes.......................................................................................................................................................... 6
1.4 Warranty conditions...................................................................................................................................................... 6
1.5 Copyright....................................................................................................................................................................... 6
1.6 License texts.................................................................................................................................................................. 6
1.7 Contact LAUDA........................................................................................................................................................... 7
2 Safety..................................................................................................................................................................................... 8
2.1 General safety information and warnings................................................................................................................... 8
2.2 Information about the interface module.................................................................................................................... 9
2.3 Personnel qualification.................................................................................................................................................. 9
3 Unpacking............................................................................................................................................................................ 10
4 Device description................................................................................................................................................................ 11
4.1 Purpose.......................................................................................................................................................................... 11
4.2 Structure........................................................................................................................................................................ 11
5 Before starting up................................................................................................................................................................ 12
5.1 Installing the interface module................................................................................................................................... 12
5.2 Using the module box.................................................................................................................................................. 14
6 Commissioning..................................................................................................................................................................... 15
6.1 Contact assignment Ethernet interface................................................................................................................... 15
6.2 Software update........................................................................................................................................................... 15
6.3 Module generation....................................................................................................................................................... 15
7 Operation.............................................................................................................................................................................. 17
7.1 Menu structure............................................................................................................................................................ 18
7.2 Interface functions....................................................................................................................................................... 18
7.2.1 General Information................................................................................................................................. 19
7.2.2 Read commands...................................................................................................................................... 20
7.2.3 Write commands...................................................................................................................................... 23
7.2.4 Error messages......................................................................................................................................... 25
7.3 Web server LAUDA Command................................................................................................................................ 26
7.4 Establishing a network connection............................................................................................................................ 29
7.4.1 Configuring network settings................................................................................................................. 30
7.4.2 Checking the network connection.......................................................................................................... 31
7.5 Setting up the process interface................................................................................................................................ 31
7.5.1 Activating the process interface (Process SST on/o)........................................................................ 31
7.5.2 Setting up the virtual COM port........................................................................................................... 32
V3 Interface module LRZ 930 3 / 44

7.5.3 Checking the connection monitoring function.................................................................................... 33
8 Maintenance........................................................................................................................................................................ 34
9 Faults.................................................................................................................................................................................... 35
9.1 Error.............................................................................................................................................................................. 35
9.2 Warning........................................................................................................................................................................ 36
10 Decommissioning................................................................................................................................................................ 37
11 Disposal................................................................................................................................................................................ 38
12 Technical data...................................................................................................................................................................... 39
13 Declaration of Conformity................................................................................................................................................ 40
14 Glossary................................................................................................................................................................................ 41
15 Index..................................................................................................................................................................................... 43
V3
Interface module LRZ 9304 / 44

1 General
Many types of LAUDA constant temperature equipment have vacant
module slots for installing additional interfaces. The number, size and
arrangement of the module slots vary depending on the device and are
described in the operating manual accompanying the constant temperature
equipment. Two additional module slots available as accessories can be fitted
to a LiBus module box, which is then connected as an external casing to the
LiBus interface on the constant temperature equipment.
This operating manual describes how to install and configure the Ethernet
interface module (catalog no. LRZ 930).
Constant temperature equipment can be connected to a PC or network
via the Ethernet interface and controlled from there using the LAUDA
command set. The interface functions provided for this purpose are
described in chapters ÄChapter 7.2.2 “Read commands” on page20
and ÄChapter 7.2.3 “Write commands” on page23.
The USB interface is intended for future expansion and does not currently
have any function.
1.1 Intended use
The interface module can only be operated as intended and under the
conditions specified in this operating manual.
The interface module may only be used in the following areas:
Production, quality assurance, research and development in an industrial
environment
The interface module is an accessory item that is used to control and monitor
the LAUDA constant temperature equipment. The interface module is built
into the device and connected to the 24 volt supply. It may only be installed
in constant temperature equipment that supports the interface provided.
Refer to the chapter "Compatibility” in this operating manual for a list of
compatible product lines.
Operation of the interface module is also permitted in combination with the
LiBus module box (LAUDA catalog no. LCZ 9727). This operating manual
also contains a description of how to install and connect up the module box.
Operation on a non-compatible device
Outdoor operation
Operation in a potentially explosive area
Operation after incomplete assembly
Operation using defective cables or connections or those that do not
confirm to standards
Operation under medical conditions in accordance with
DIN EN 60601-1 or IEC 601-1
Reasonably foreseeable improper use
V3 Interface module LRZ 930 5 / 44

1.2 Compatibility
The interface module is available as an accessory for the following LAUDA
product lines:
Variocool, not compatible with Variocool NRTL (Ethernet interface
fitted ex works)
No operation of interfaces of the same type
Depending on the constant temperature equipment, only one
Ethernet interface can be used. This applies irrespective of the
interface type.
1.3 Technical changes
All technical modifications are prohibited without the written consent of the
manufacturer. Damage resulting from a failure to observe this condition will
void all warranty claims.
However, LAUDA reserves the right to make general technical modifica-
tions.
1.4 Warranty conditions
LAUDA grants a standard warranty of one year.
1.5 Copyright
This operating manual was written in German, checked and approved. If
the content of other language editions deviates from the German edition,
the information in the German edition shall take precedence. If you notice
any discrepancies in the content, please contact LAUDA Service, see
ÄChapter 1.7 “Contact LAUDA” on page7.
Company and product names mentioned in the operating manual are usually
registered trademarks of the respective companies and are therefore subject
to brand and patent protection. Some of the images used may also show
accessories that are not included in the delivery.
All rights reserved, including those relating to technical modifications and
translations. This operating manual or parts thereof may not be modified,
translated or used in any other capacity without the written consent of
LAUDA. Violation of this may obligate the violator to the payment of dam-
ages. Other claims reserved.
1.6 License texts
You can obtain the license texts for the software used in the device from the
English version of our LAUDA website.
1. Enter the following into your browser's address line: https://
www.lauda.de/en/ and confirm your entry.
2. Click Services Download center.
3. In the Download center, chose the [English] option in the [Language]
drop-down list.
A long list of files with multiple formats appears.
V3
Interface module LRZ 9306 / 44

4. Now also filter the [Software] entry in the [Document type] drop-
down list.
The list is now more clear.
5. Search for the following software files:
Lizenztexte_µTZ.txt
Lizenztexte_Qt.txt
GUF-Yocto-31.2-r8616-0.tar.zip
Click on the relevant file line to start the download.
The file starts downloading.
1.7 Contact LAUDA
Contact the LAUDA Service department in the following cases:
Troubleshooting
Technical questions
Ordering accessories and spare parts
Please contact our sales department for questions relating to your specific
application.
Contact information
LAUDA Service
Phone: +49 (0)9343 503-350
Email: [email protected]
V3 Interface module LRZ 930 7 / 44

2 Safety
2.1 General safety information and warnings
Read this operating manual carefully before use.
Keep the operating manual in a place within easy reach of the interface
module.
This operating manual is part of the interface module. If the interface
module is passed on, the operating manual must be kept with it.
This operating manual is applicable in combination with the operating
manual of the constant temperature equipment in which the interface
module is installed.
Manuals for LAUDA products are available for download on the
LAUDA website: https://www.lauda.de
The warnings and safety instructions in this operating manual must be
observed without fail.
There are also certain requirements for personnel, see ÄChapter 2.3
“Personnel qualification” on page9.
Warning signs Type of danger
Warning – danger zone.
Signal word Meaning
WARNING! This combination of symbol and
signal word indicates a potentially
dangerous situation that can result
in death or serious injury if it is not
avoided.
NOTICE! This combination of symbol and
signal word indicates a poten-
tially dangerous situation that can
result in material and environmental
damage if it is not avoided.
Structure of warnings
V3Interface module LRZ 9308 / 44

2.2 Information about the interface module
Always disconnect the constant temperature equipment from the power
supply before installing the interface module or connecting interfaces.
Always take the recommended safety measures against electrostatic
discharge before handling interface modules.
Avoid touching the circuit board with metallic tools.
Do not start up the constant temperature equipment before installation
of the interface module is complete.
Store any unused interface modules in their packaging in accordance
with the specified ambient conditions.
Use only suitable cables of sucient length for cable connections.
Make sure that the protective screen on the cables and connec-
tors complies with EMC regulations. LAUDA recommends using pre-
assembled cables.
Always lay cables correctly so that they do not pose a tripping hazard.
Secure the laid cables and make sure that they cannot be damaged
during operation.
Check the condition of the cables and interfaces prior to each opera-
tion.
Immediately clean any soiled parts, in particular unused interfaces.
Make sure that the signals transmitted via the interface correspond to
the permitted operating parameters of the interface module.
2.3 Personnel qualification
Only specialized personnel are permitted to install interfaces modules. Spe-
cialized personnel are personnel whose education, knowledge, and experi-
ence qualify them to assess the function and risks associated with the device
and its use.
Specialized personnel
V3 Interface module LRZ 930 9 / 44

3 Unpacking
NOTICE!
Transport damage
Device damage
Closely inspect the device for transport damage prior to
starting up.
Never operate a device that has sustained transport
damage!
NOTICE!
Electrostatic discharge
Material damage
Always observe safety measures against electrostatic dis-
charge.
Please observe the following installation sequence:
1. Remove the interface module from its packaging.
2. If you want to store the interface module at the installation location,
use the outer packaging. This packaging is protected against static
charging.
3. After installing the equipment, dispose of the packaging materials in
line with environmental regulations, see Ä“Packaging” on page38.
If you discover any damage on the interface module, contact
LAUDA Service immediately, see ÄChapter 1.7 “Contact
LAUDA” on page7.
V3Interface module LRZ 93010 / 44
Altri manuali per LRZ 930
1
Indice
Altri manuali Lauda Unità di controllo
Manuali Unità di controllo popolari di altre marche

Festo
Festo Compact Performance CP-FB6-E Manuale elenco delle parti

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manuale utente

JS Automation
JS Automation MPC3034A Manuale utente

JAUDT
JAUDT SW GII 6406 Series Guida rapida

Spektrum
Spektrum Air Module System Manuale utente

BOC Edwards
BOC Edwards Q Series Manuale utente