Guidance and manufacturer's declaration – electromagnetic
immunity – for all EQUIPMENT and SYSTEMS
Immunity test IEC 60601
test level Compliance level Electromagnetic
environment - guidance
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Guidance and manufacturer´s declaration – electromagnetic immunityThe
The model IRT101B/IRT101 is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
IRT101B/IRT101 should assure that it is used in such an environment.
Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY
– for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunityThe
The model IRT101B/IRT101 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model IRT101B/IRT101 should
assure that it is used in such an environment.
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Radiated
RF
IEC
61000-4-3
Portable and mobile RF communications
equipment should be used no closer to
any part of the model IRT101B/IRT101,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the
compliance level in each frequency range b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the model IRT101B/IRT101
is used exceeds the applicable RF compliance level above, the model IRT101B/ IRT101
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the model
IRT101B/IRT101.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
3 V/m
80 MHz to 2.5
GHz 3 V/m
± 6 kV contact
± 8 kV air ± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or
ceramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at least
30 %.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.