Monarch Nova-Strobe dax Manuale utente

Nova-Strobe dax
and
Nova-Strobe dbx
Portable Deluxe Stroboscopes
MONARCH INSTRUMENT
Instruction Manual
15 Columbia Drive
Amherst, NH 03031 USA
Phone: (603) 883-3390
Fax: (603) 886-3300
E-mail: [email protected]
Website: www.monarchinstrument.com
(Nova-Strobe dbx shown)
Printed in the U.S.A.
Copyright 2007 Monarch Instrument, all rights reserved
1071-4203-111R

Safeguards and Precautions
1. Read and follow all instructions in this manual carefully, and
retain this manual for future reference.
2. Do not use this instrument in any manner inconsistent with
these operating instructions or under any conditions that
exceed the environmental specifications stated.
3. Use of this product may induce an epileptic seizure in persons
prone to this type of attack.
4. Objects viewed with this product may appear to be stationary
when in fact they are moving at high speeds. Always keep a
safe distance from moving machinery and do no touch the target.
5. There are lethal voltages present inside this product. Refer to
the section on Lamp Replacement before attempting to open
this product.
AC Stroboscopes that have three wire mains cable must
have the earth wire connected to a suitable Earth point.
6. Do not allow liquids or metallic objects to enter the ventilation
holes on the stroboscope as this may cause permanent
damage and void the warranty.
7. Do not allow cables extending from unit to come into contact
with rotating machinery, as serious damage to the equipment,
or severe personal injury or death may occur as a result.
8. This instrument may not be safe for use in certain hazardous
environments, and serious personal injury or death could occur
as a result of improper use. Please refer to your facility’s
safety program for proper precautions.
CE DECLARATION OF CONFORMITY
As Manufacturer:
Monarch Instrument
Division of Monarch International Inc.
15 Columbia Drive, Amherst NH 03031 USA
declares under Monarch’s sole responsibility that the product:
to which this declaration relates is in conformity with the following standards:
and therefore conforms with the requirements of Council Directive 89/336/EEC
relating to electromagnetic compatibility and 93/68/EEC relating to the low voltage
directive with amendments, when operated in accordance with the user guide.
EMC testing of this product was performed by Retlif Testing Laboratories, NH, in
February of 2007 (File R-4702N-5). Safety testing per Technical Construction File
NSXPB-0207.
Name: Nova-Strobe X Series
Models: pbx, dbx, bbx
14th February, 2007
Manufacturer (Amherst,NH) Alan Woolfson, VP Engineering (Authorized Signature)
EN61326:1997 EMC /A1:1998/A2:2001/A3:2003 Class A
Specifically CISPR 16-1:2003/CISPR 16-2:2003
EN55011:1998/A1:1999/A2:2002 EN61000-4-2 EN61000-4-3
EN61010-1:2001-2 Safety Regulation

9. The Nova-Strobe dbx contains Nickel Metal Hydride batteries
which must be disposed of in accordance with Federal, State,
& Local Regulations. Do not incinerate. Batteries should be
shipped to a reclamation facility for recovery of the metal and
plastic components as the proper method of waste
management. Contact distributor for appropriate product
return procedures.
10. This instrument is not user serviceable. For technical
assistance, contact the sales organization from which you
purchased the product or Monarch Instrument directly.
In order to comply with EU Directive 2002/96/EC on Waste
Electrical and Electronic Equipment (WEEE): This product
may contain material which could be hazardous to human health
and the environment. DO NOT DISPOSE of this product as un-
sorted municipal waste. This product needs to be RECYCLED in
accordance with local regulations, contact your local authorities
for more information. This product may be returnable to your distributor for
recycling - contact the distributor for details.
Monarch Instrument’s Limited Warranty applies. See
www.monarchinstrument.com for details.
Warranty Registration and Extended Warranty coverage available
online at www.monarchinstrument.com.
CE DECLARATION OF CONFORMITY
As Manufacturer:
Monarch Instrument
Division of Monarch International Inc.
15 Columbia Drive, Amherst NH 03031 USA
declares under Monarch’s sole responsibility that the product:
to which this declaration relates is in conformity with the following standards:
and therefore conforms with the requirements of Council Directive 89/336/EEC
relating to electromagnetic compatibility and 93/68/EEC relating to the low voltage
directive with amendments, when operated in accordance with the user guide.
EMC testing of this product was performed by Retlif Testing Laboratories, NH (File
R-4702N-4). Safety testing per Technical Construction File NSXDA-0207.
Name: Nova-Strobe X Series
Models: dax, bax
1st July, 2007
Manufacturer (Amherst,NH) Alan Woolfson, VP Engineering (Authorized Signature)
EN61326:1997 EMC /A1:1998/A2:2001/A3:2003 Class A
Specifically CISPR 16-1:2003/CISPR 16-2:2003
EN55011:1998/A1:1999/A2:2002 EN61000-4-2 EN61000-4-3
EN61000-4-4 EN61000-4-5 EN6100-4-6 EN6100-4-3
EN61010-1:2001-2 Safety Regulation

TABLE OF CONTENTS
TABLE OF CONTENTS
1.0 OVERVIEW ............................................................................. 1
1.1 Display Panel / Definition of Buttons ........................... 1
2.0 PREPARATION FOR USE ........................................................ 3
2.1 Power ............................................................................ 3
2.2 Input / Output Connections .......................................... 3
3.0 MENU ...................................................................................... 5
4.0 OPERATION ............................................................................ 6
4.1 Internal Mode - Standard Strobe Operation ............... 6
4.2 Internal Mode - TACH Frequency Generator ............... 9
4.3 External Input Mode ..................................................... 9
4.4 Tachometer Mode - External Input Required ............ 10
4.5 Power Up Features .................................................... 10
5.0 USING THE STROBOSCOPE TO MEASURE RPM ................ 11
6.0 LAMP AND FUSE REPLACEMENT ....................................... 13
6.1 Lamp Replacement .................................................... 13
6.2 Fuse Replacement ..................................................... 15
7.0 BATTERY POWERED MODELS ONLY .................................. 16
7.1 Low Battery Indication ............................................... 16
7.2 Charging the Batteries ............................................... 17
7.3 Battery Disposal ......................................................... 18
8.0 SPECIFICATIONS .................................................................. 18
9.0 OPTIONS AND ACCESSORIES ............................................ 21
9.0 OPTIONS AND ACCESSORIES
C-4027 Set of mating 1/8 inch [3.5 mm] stereo phone plugs (to provide
TTL signal and sensor power)
CA-4044-6 6 foot [1.8 m] Input / Output cable, 1/8 inch [3.5 mm] male
phone plug to male BNC connector
CA-4045-6 6 foot [1.8 m] Input / Output cable, 1/8 inch [3.5 mm] male
phone plug to 1/8 inch [3.5 mm] male phone plug for daisy
chaining strobes together
CC-7 Latching carrying case for Strobe with provision for
accessories
L-1903 Digital Stroboscope replacement lamp
PSC-2 Recharger, 115 Vac - 50/60 Hz for battery operated Nova-
Strobes
PSC-2U Universal Recharger, 115/230 Vac with USA, U.K., AUS,
Euro Adapter Plugs for battery operated Nova-Strobes
ROS-P Remote Optical Sensor with 8 foot [2.5 m] cable for triggering
strobe
T-5 Reflective tape - 5 foot [1.5 m] roll, 0.5 inch [12.7 mm] wide
MT-190P Magnetic Trigger Sensor/Amplifier with 8 foot [2.5 m] cable
for triggering strobe
IRS-P Infrared Sensor for use without reflective target at 0.5 inch
[12 mm] gap with 8 foot [2.5 m] cable for triggering strobe
SPC-1 Splash proof Protective Cover for Battery Powered Strobe
ONLY
21

1
1.0 OVERVIEW
All descriptions in this manual apply to both the battery powered (dbx)
and AC mains powered (dax) digital stroboscopes except where noted.
1.1 Display Panel / Definition of Buttons
The display panel consists of
a backlighted liquid crystal
display with six numeric digits
on top and five alphanumeric
digits on the bottom, which
indicate modes, flash rates, etc.
(see Figure 1).
Additional information
displayed include:
TIME Unused
AUTO Unused
ALT. Indicates
alternate function of each button (lower section) and
knob will be used
TACH Tachometer Mode active (strobe won’t flash)
LOCK Unused
EXT External Input Mode active
On Target Indicator for Tachometer Mode and
Remote Sensor in External Mode
- - - - - Indicates input frequency exceeds the limit of the
stroboscope
(Battery Powered Model Only) Battery indication,
see section 7.0
Figure 1 Display Panel
20
dbx Specific:
Trigger to Flash Delay 13 µsec typical
Input Power Battery powered: Internal Rechargeable
Batteries 6 Vdc, ExternalAC recharger (100 Vac
to 240 Vac, 50/60 Hz)
Light Output Average: 13 Watts typical > 4000 FPM
Instantaneous (per flash): 230 mJoule typical
to 4000 FPM
Run Time 2 hours typical at 1800 FPM, and over 1 hour at
6000 FPM with fully charged batteries
Charge Time 4-5 hours typical with PSC-2U
Weight 1.875 lbs [0.8505 kg] including batteries
dax Specific:
Trigger to Flash Delay 16 µsec typical
Input Power AC powered: 115 Vac OR 230 Vac, 35VA (as
ordered)
Light Output Average Power: 15.5 Watts typical > 4000 FPM
Instantaneous (per flash): 230 mJoule typical
to 4000 FPM
Run Time Continuous within temperature limitations.
Vents must not be restricted. Unit must be
horizontal.
Weight 1.55 lbs [0.70 kg]

2
Below the display are six membrane buttons which control the operation
of the Stroboscope. They are:
Multiplies flash rate by 2 times
ALT Function - Starts Menu (See section 3.0)
Hold when powering up to show all segments, then
Rev # and display test
Divides flash rate by 2
Hold when powering up to reset factory defaults
Toggles display between RPM and RPS
ALT Function - Toggles Tach Mode (flashing) on/off
Manually toggles between Internal and External Modes
ALT Function - Reads and Stores 9 preset flash rates
Activates Alternate Function for buttons (lower section)
and knob
In Internal Mode, toggles between normal flash rate
adjust and “phase” delay adjust
The ALT FUNCTION button toggles ALT. in the display. When ALT. is
displayed, the buttons will perform their secondary function listed in
the lower section of each button. It also changes how the tuning knob
works.
19
External Input TTL Compatible (24V pk max), 500 nanosec
min pulse width,
Positive or Negative edge triggered (menu
selectable)
Output Pulse 40 µsec positive/negative pulse (menu
selectable), 3.3 Vdc typical
Note: dax – Input/Output connectors isolated
from AC power
Time Base Ultra Stable Crystal Oscillator
Display LCD display with 6 numeric 0.506 inch [12.85
mm] high digits and
5 alphanumeric 0.282 inch [7.17 mm] high digits
Indicators Battery level, On Target, ALT, TACH, and EXT
icons
Knob Adjustment Digital Rotary switch with 36 detents per
revolution; velocity sensitive
Memory Last setting before power down is remembered
and restored on next power up. 9 user settable
flash rates.
Flash Duration 10-25 microseconds (auto adjust with flash rate)
Flash Tube (Lamp) Life 100 million flashes
This product is designed to be safe for indoor use under the following
conditions (per IEC61010-1).
Operating Temperature 32-104 ºF [0-40 °C]
NOTE: Safety thermal feature will set unit into TACH Mode (stops
flashing) in the event of internal overheating.
Humidity Maximum relative humidity 80% for temperature
up to 88 °F [31 °C] decreasing linearly to 50%
relative humidity at 104 °F [40 °C]

3
2.0 PREPARATION FOR USE
The Strobe may be hand held or mounted on a tripod or other user supplied
bracket using the ¼-20 UNC bushing at the base of the handle.
2.1 Power
The AC powered strobe must have its power cord plugged into an AC
outlet (115Vac or 230Vac).
The battery powered strobe has internal rechargeable batteries. The
unit should be charged before use (see section 7.0). The actual operating
time of the stroboscope depends on the flash rate and duty cycle of
operation. Slower flash rates (below 4,000 FPM) increase the operating
time. Note that the strobe will not operate from the recharger supplied
with the unit.
2.2 Input / Output Connections
The strobe has input and output jacks on the left side of the stroboscope.
These can be used for external triggering or synchronization (daisy
chaining two or more strobes). These jacks accept 1/8" (3.5mm) phone
plugs (input - stereo, output - mono). The input and output are TTL
compatible. See Figures 2 and 3 for connector connection detail.
Signal Input
+6V Out to
Sensor
Common
(GND)
Common
(GND)
+6V Out to
Sensor
Signal Input
Figure 2 Input Connector Detail (Stereo plug)
18
7.3 Battery Disposal
Prior to disposing of the battery-powered strobe, the user must remove
the Nickel Metal Hydride batteries. To do this, remove the lens,
reflector and lamp as detailed in the Lamp Replacement section. This
will expose four (4) screws that must be removed so the reflector
housing can be dismantled. There are four (4) additional screws in the
case half opposite the input and output jacks that must be removed.
The case halves can now be separated, exposing the batteries. Remove
the cables from the batteries and place tape over the battery terminals
to prevent them from shorting. The batteries should be sent to a
recycling center or returned to the factory. The rest of the parts may
now be disposed of.
8.0 SPECIFICATIONS
Internal Mode:
Flash Range 30 - 20,000 FPM (Flashes Per Minute)
Flash Rate Accuracy 0.002% of setting or ±last digit
Flash Rate Resolution 0.01 to 1 FPM (menu selectable), 0.1 FPM
max resolution above 9,999.99 FPM
Display Update Rate Instantaneous
External Modes:
Flash Range and Display same as internal mode - External flash rates
to 0 are acceptable
Tachometer Measurements 5 to 250,000 RPM
Accuracy ±0.001% of reading or ± last digit
Display Update Rate 0.5 second typical

4
The input jack (Spointing into socket) enables an external signal to
trigger the strobe. Inserting a plug into the input jack will automatically
put the strobe into the External Input Mode. When the plug is removed,
the strobe will be put back into the Internal Mode. The Internal Mode
can be forced by pressing the INPUT button. The polarity of the
input pulse can be set in the MENU options.
With no external input, the output jack (Tpointing away from socket)
provides a TTL compatible pulse from the strobe’s internal oscillator.
If an external input is applied, the output pulse is in sync with the
input pulse. This output pulse may be used to trigger a second
stroboscope synchronously to illuminate larger areas. Many strobes
can be “daisy chained”. The output jack of one strobe is connected to
the input jack of the next strobe causing all the strobes to flash together
and be controlled by the first strobe in the chain. The polarity of the
output pulse can be set in the MENU options.
Signal Input
Common
(GND)
Common
(GND)
Signal Input
Figure 3 Output Connector Detail (Mono plug)
17
The strobe has a protection feature that prevents the strobe from
operating if the battery voltage is too low. This condition is indicated
by no flash and the display shows “LO BAT”. At this time the
batteries must be recharged. Remember to release the trigger switch.
7.2 Charging the Batteries
The unit may be recharged at any time. You do not need to wait until
the low battery condition is indicated.
To charge the battery powered strobe with the recharger:
1. Release the trigger so the strobe is off.
2. Plug the recharger cable into the recharger socket (located below
the display panel behind the handle).
3. Plug the recharger into an AC mains wall outlet.
CAUTION: Use of rechargers other than the one supplied
(PSC-2 or PSC-2U) will damage the stroboscope
and void the warranty.
The strobe will display the current battery charge percent and the
recharger will fast charge the batteries for about 4-5 hours. Once the
fast charge is completed, the recharger will trickle charge the batteries.
Allow the recharger to charge the batteries until the display shows
DONE for peak battery life performance. If the batteries are not
charged to 100% regularly, the batteries will lose capacity.
When the recharger plug is inserted into the recharger jack, the strobe
will go into the Charging Mode. Make sure the trigger switch is not
depressed. The strobe will not do anything else when charging (e.g. it
will not flash and the buttons have no function).

5
3.0 MENU
The strobe has a Menu which allows the user to select settings such as
number of decimal places, backlight on or off, and positive or negative edge
for input and output signal.
To enter the MENU:
1. Press the ALT FUNCTION button and then the MENU button.
2. SetUP and the menu option will be displayed.
3. Turn the tuning knob to cycle through the main menu selections.
4. Once the desired menu option is displayed, press the MENU button to
select/save it. Press any other button to cancel.
5. Turn the tuning knob to edit the menu option.
6. Press the MENU button to save your changes. Press any other button
to cancel.
7. Press any button other than MENU to exit the Main Menu.
8. DONE will be displayed.
Below is a list of the menu items:
DECPTDECPT
DECPTDECPT
DECPT- Decimal Point (none, 1 or 2)
BLITEBLITE
BLITEBLITE
BLITE- Backlight (Yes=On or No=Off)
INPUTINPUT
INPUTINPUT
INPUT- Positive (pos) or Negative (neg) Edge for Input Signal
OUTPTOUTPT
OUTPTOUTPT
OUTPT- Positive (pos) or Negative (neg) Edge for Output Signal
16
7.0 BATTERY POWERED MODELS ONLY
The Nova-Strobe dbx is fitted with rechargeable NiMH (Nickel Metal
Hydride) batteries. These batteries contain fewer toxic metals than NiCd
(Nickel Cadmium) and are currently classified “environmentally friendly”.
They also have 30% more capacity than NiCd batteries of the same size.
Like NiCds, NiMH batteries are prone to self-discharge - 10 to 15% of
charge is lost in the first 24 hours then continues at a rate of 0.5 to 1% per
day. For maximum performance, charge the batteries just prior to use.
When not in use, the batteries should be charged at least every three months,
otherwise the battery capacity will be reduced or the batteries may become
unusable.
Charge the batteries before use and allow 3-5 cycles of charging and
discharging for batteries to reach full capacity.
The enclosure contains control electronics to properly and safely charge the
batteries. Never remove the batteries from the enclosure and attempt to
charge externally. Always use the charger supplied (PSC-2 or PSC-2U).
7.1 Low Battery Indication
The Low Battery icon will be displayed differently depending on the
battery voltage. If the Low Battery icon is not displayed, the batteries
are full. Otherwise, the Low Battery icon will indicate that the battery
voltage is decreasing as follows:
Half solid ¾ Full
Half with block blinking ½ Full
Outline only solid ¼ Full
Outline blinking Low battery (very little time left)

4.0 OPERATION
To turn on the stroboscope, depress and hold the trigger. The trigger may be
locked in position using the side locking button. To lock the stroboscope
on, depress the trigger as far as it will go and then press the locking button.
Once the locking button is set you may release the trigger and the trigger will
be held in place. To unlock the stroboscope, simply depress the trigger and
then release.
There are three major operating modes for the Strobe. These are Internal,
External Input and Charging (Battery powered model only). In the
Internal Mode, the knob adjusts the flash rate. In the External Input
Mode, an external signal is used to trigger the flash and the knob has no
effect. The Charging Mode (Battery powered model only) is when the
strobe has the battery recharger plugged into it. The strobe will continuously
display the state of the battery charge while being recharged.
4.1 Internal Mode - Standard Strobe Operation
In the Internal Mode the stroboscope generates it’s own signals and
functions like a tunable stroboscope. The strobe is in the Internal
Mode when nothing is plugged into the input jack or when manually
set using the INPUT button.
To change the flash rate:
With the power on, turn the knob counter clockwise to increase the flash
rate and clockwise to decrease it. The knob is velocity sensitive. Turn the
knob slowly to have each “click” is equal to 0.01 FPM. Turning the knob
more quickly will adjust the FPM by larger steps. When adjusting flash
rate, quickly turn the knob (or use the x2 or ÷2 buttons) to coarsely
change the FPM. Then slowly turn the knob for fine adjustments.
615
6.2 Fuse Replacement
Under normal operating conditions, the fuse within the stroboscope
should never blow. Examples of abnormal operating conditions would
be foreign materials entering the strobe, such as water, pulp, ink, etc.
The AC Powered stroboscope has a replaceable fuse inside the unit,
which may be accessed by removing the lens and reflector - refer to
Figure 4. If the fuse needs to be replaced, replace only with a fuse of
the same type and value: Fast Blow - 750mA, 2AG.
WARNING: Before attempting to replace the fuse, make
sure the stroboscope is turned off and any
mains cord is removed from the AC outlet. Allow
the lamp to cool waiting at least 5 minutes.
The Battery Powered stroboscope has a resettable fuse, which will
reset once conditions are normal again.
Questo manuale è adatto per i seguenti modelli
1
Indice
Altri manuali Monarch Apparecchiature di prova
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione