• The Fiber height Pass/Fail checking has been repaired so that it now properly
uses the IEC or Telcordia curve (allowable fiber recess vs ROC) when the user
has selected the IEC or Telcordia criteria (See Setup tab section in Chapter 6 in
the manual).
• In the Report Generate dialog, once a user clicks on the Connector ID or Job ID
field, the attention (mouse cursor) now stays with that field until the user clicks or
tabs elsewhere in the dialog, regardless of number of characters typed.
• The display of measurement results has been modified by initializing all APC type
measurement fields to 0.0 if the PC scan mode has been selected.
Changes for NCC v3.00:
• Added file export capability to the Report Generation feature. Two types of file
export are supported; PDF and HTML. PDF export under this implementation no
longer requires having Adobe PDF Distiller installed, and unlike the previous PDF
export that used Adobe Distiller, the graphics included in each report always
appear in the document. This new file export capability is activated directly from
the main Report Generation dialog.
• Added a new “JobID” field to data output collection (Scan History screen list, Text
File export, new XL file created from Text File, Database, and generated Reports).
This is an alphanumeric value up to 50 characters long optionally entered by the
User prior to activating a scan.
• Redesigned and expanded filtering options in Report Generation dialog to allow
scan selection on the basis of JobID (see above “Added a new JobID field” note),
overall scan PASS status, scan date, and ConnectorID.
• NCC reports now print in ascending order of time (oldest scan first) when batch
previewing or printing.
• There is now no limit to the amount of scan records that can be batch printed with
the NCC report generation feature (although Print Preview is still limited to 50 at a
time).
• Improved ease of use and intuitiveness of Company Info configuration for
Reports.
• Modified Illumination setup tool and made it available to all levels of NCC users.
It’s purpose is to check for illumination saturation and provide guidance to users
when called upon to adjust Illumination Gain at the back of the unit. Added it as
“BAT” (Brightness Adjust Tool) option under Video Control menu.
• Repaired [Open in XL] function so that it no longer is “unable to find file” if Text
Export full file path is not showing on Export page title.
• Connector ID is no longer cutoff on Report’s label section if it is less than 25 digits
long.
• Fixed incorrect application of Key Error limits greater than +/-0.05 degree.
• Changes made to prevent NCC crash when measuring connectors having high
apex offset (as much as 400 microns).
• Changes made to insure NCC reports always shows correct Pass/Fail
representation (previously, if a parameter showed FAIL in an earlier report, that
same parameter might continue to show fail in a later report though it actually
passed in the corresponding scan).