OfiTE 170-00 Manuale utente

HTHP Filter Press for Drilling Fluid Testing
with 175-mL, Single-Capped Test Cell and CO2Pressuring Assemblies
#170-00: 115 Volt
#170-01: 230 Volt
Instruction Manual
Updated 6/14/2019
Ver. 4
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ote.com
©Copyright OFITE 2014

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 1
Intro..................................................................................................2
Specication ...................................................................................3
Components ...................................................................................4
Safety...............................................................................................8
Diagram .........................................................................................15
Quick Start ....................................................................................16
Operation.......................................................................................19
Data................................................................................................26
Maintenance..................................................................................27
Warranty and Return Policy ........................................................31
Table of
Contents

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 2
The OFI Testing Equipment (OFITE) High Temperature High Pressure
(HTHP) Filter Press is designed to evaluate the ltration characteristics of
drilling uids, cement slurries, fracturing uids, and completion uids under
elevated temperatures and pressures. Evaluating uids under HTHP condi-
tions similar to the downhole environment is of paramount importance. Fluid
properties must be monitored while under high temperatures and pressures
as ltration behavior and wall cake building characteristics of permeable
formations change with changing environments. These characteristics are
affected by the shape, type, and quantities of solids present in the uid and
their physical and electro-chemical interactions, all of which are affected by
changing temperatures and pressures.
OFITE manufactures and provides HTHP ltration units in two basic sizes,
175 mL and 500 mL capacities. Both are used extensively throughout the
world and in all environments, but in general the 175 mL units are designed
for eld portability, while the larger 500 mL units are designed for laboratory
usage at higher temperatures and pressures. All OFITE Filtration devices fully
conform to American Petroleum Institute (API) specications.
A complete HTHP Filter Press consists of a controlled pressure source, usu-
ally Nitrogen pressurization or Carbon Dioxide bulbs for the 175 mL units.
Top and bottom pressure manifolds are provided to simulate the differential
pressures found in a down-hole environment, and to prevent evaporation
of the base uid if exceeding the boiling point of that uid. The test cells are
provided in a variety of assemblies, depending upon the type of uid tested,
the lter media, and the temperatures and pressures desired. The test cells
are encased inside a heating jacket, which is adjustable.
A variety of lter media are available, the most common being standard API
lter paper, cement screens, and ceramic lters. The ceramic lters may be
obtained to match the pore throat or permeability of the formation. Natural for-
mation lters or cores may also be used of differing pore throat / permeability
sizes. Slotted disks of varying sizes are frequently used for lost circulation
materials studies.
The 175 mL has a maximum temperature of 350°F (177°C) and the 500 mL
heating jackets are capable of reaching 400°F (204°C). However, lower uid
volumes due to uid expansion at higher temperatures, limit the 175 mL units
to a useful working temperature of 300°F (149°C). Anyone running tests
above 350°F (177°C) must substitute a complete set of o-rings after each and
every test.
Intro

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 3
Size: 7.5" × 11" × 23.5" (19.1 × 27.9 × 59.7 cm)
Weight: 27 lb (12.3 kg)
Shipping Size: 20" × 13" × 13" (51 × 33 × 33 cm)
Shipping Weight: 33 lb (15 kg)
Maximum Temperature
(Heating Jacket): 350°F (177°C)
Maximum Pressure (Cell): 2,000 psi (13.8 MPa)
Maximum Pressure (Receiver): 750 psi (5.1 MPa)
Pressure Source: CO2 Bulbs
Test Cell Capacity: 175 mL
Receiver Volume: 15 mL
Heater: 400 Watt
Power Requirement: 115 VAC, 5 Amps, 50/60 Hz
230 VAC, 3 Amps, 50/60 Hz
Specication

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 4
Components Supplies:
#153-14 Graduated Cylinder, 50 mL × 1 mL
#154-10 Dual-Scale Thermometer with Dial, 5" Stem, 50° - 500°F (0° -
250°C)
#170-19 Filter Paper, 2½" (6.35 cm), Specially Hardened for Filter Presses,
100/Box
#170-35 Adjustable Wrench, 6"
Assemblies:
#170-04 CO2Pressurize Unit:
#143-02-10 CO2 Puncture Head Assembly
#143-02-12 Puncture Pin
#143-02-13 O-ring
#143-02-14 O-ring
#143-03 Barrel for CO2Cartridge
#170-08 Regulator
#170-20 Manifold Block
#170-32 Needle Valve, Male, ⅛" × ⅛"
#171-23-1 Safety Pin with Lanyard
#171-34 Gauge, 1,500 psi, 2" Face, ¼" NPT Bottom
#170-06 Back Pressure Receiver, 15-mL Stainless Steel Tube for CO2
#143-00 Regulator
#143-01 Gauge, 200 psi, ⅛" Bottom Connection
#143-02-10 CO2 Puncture Head Assembly
#143-02-12 Puncture Pin
#143-02-13 O-ring
#143-02-14 O-ring
#143-03 Barrel for CO2Cartridge
#143-06 Safety Bleeder Valve
#143-11 Elbow
#170-07 O-ring
#170-28 Receiver Body
#170-32 Needle Valve, Male, ⅛" × ⅛"
#171-23-1 Safety Pin with Lanyard
#170-12-2 Cell Assembly, 175 mL, Single Cap, 2,000 psi
#170-13-3 O-ring for Test Cell, Viton®/Fluorocarbon (FKM)
#170-16 Valve Stem
#170-17 O-ring for Valve Stem, Viton®/Fluorocarbon (FKM)
#170-26-1 Locking Screw, Hardened Alloy Steel
#170-27 Allen Wrench, 5/3 2 "
#170-45-2 Cell Body, 316 Stainless Steel
#171-21 Cell Cap, 2000 psi, with 60 Mesh Screen
#170-00-1 Heating Jacket, 115 Volt:
#170-01-1 Heating Jacket, 230 Volt:
#164-32 Male Connector for Power Cable (230 Volt)
#170-05 Thermostat for HTHP Filter Press 50-500
#170-10 Thermostat Pilot Light
#170-11 Heating Element, 200W, Qty: 2
#170-15 Base

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 5
#170-21 Stand
#170-30 Stainless Steel Thermostat Cover
#170-44 Rubber Foot ½", Qty: 4
#171-32 Midget Knob
#171-82 Power Cord with Male Plug, 8' (115 Volt)
Optional:
#143-05 CO2 Bulbs, 8-Gram, Package of 10, UN 1013
#152-00 Hamilton Beach Mixer, With Container
#152-01 Armature For Model 936 H.B. Mixer, 115 Volt
#155-20 Timer; 60 Min. Interval
#170-03 Carrying Case, Stainless Steel
#170-13 O-ring for Test Cell, NBR/Nitrile (Buna N), For temperatures up
to 250°F (121°C)
#170-13-4 O-ring for Test Cell, Peruorocarbon (FFKM), For temperatures
up to 500°F (260°C)
#170-13-5 O-ring for Test Cell, Ethylene propylene (EPM/EPDM), For tem-
peratures up to 400°F (204.4°C), Water-based uids only
#170-33 HTHP Cell Cap Puller
#170-40 Test Cell Removal and Carrying Tool
#170-91 HTHP Pressure Relief Tool
#170-92 Safety Clamp for HTHP Fluid Loss Cells
#170-00-SP Spare Parts for 170-00 Series Filter Press:
Part Number Description Qty.
#140-60-01 O-ring for Bleeder Valve 2
#143-00-1 Diaphragm for Airco Regulator 1
#143-01 Gauge, 200 psi, ⅛" Bottom Connection 1
#143-02-13 O-ring for Puncture Pin Holder 2
#143-02-14 O-ring for Puncture Pin Holder, Rear 2
#143-05 CO2 Bulbs, 8 Gram, Package of 10, UN 1013 60
#143-07 Repair Kit for Regulator (#143-00) 1
#153-14 Graduated Cylinder, 50 mL × 1 mL 2
#154-10 Dial Thermometer, 5" Stem, 50° - 500°F and 0 - 250°C 1
#170-13-3 O-ring for Test Cell, Viton®/Fluorocarbon (FKM) 50
#170-16 Valve Stem 4
#170-17 Valve Stem O-ring 48
#170-19 Filter Paper, 2½" (6.35 cm), Hardened for Filter Press 5
#170-26-1 Locking Screw, Hardened Alloy Steel 12
#170-27 Allen Wrench, 5/3 2 "1
#171-23-1 Safety Pin with Lanyard 1
Spare parts listings are intended to be used as a reference for future purchas-
es. Everyone’s consumable requirements will be different, and replacement
quantities needed will depend upon the number of test performed on a daily
and/or weekly basis.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 6
Optional Items for HTHP Filtration Testing:
The items listed below are not included in the HTHP Filter Press, but they are
items that will enable the technician to perform a more uniform and reproduc-
ible test while maintaining a high degree of safety. As optional items, the us-
age is not compulsory, but consideration should be given to these items when
running tests at elevated temperatures and pressures. Some of the items will
be used only on cell assemblies using set screws as fasteners, while others
should be implemented when operating all ltration equipment.
Interval Timer, 60 minute
(#155-20)
Cell Cap Removal Tool
(#170-33)
(Set Screw Cell Assemblies Only)
Safety Clamp (#170-92)
(Set Screw Cell Assemblies Only)
HTHP Pressure Relief Tool
(#170-91)
(To release trapped pressure)
Safety Shield (#171-06)
Cell Carrying Tool (#170-40)

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 7
High Pressure Nitrogen Assy.
(#171-31)
Thermocouple Assembly
(#171-45-1)
(Direct temperature measurement
Of the uid Inside the Cell)
Stand for HTHP Cell Assembly
(#171-190-028)

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 8
Safety Pressure and Temperature Considerations
Do not use or recongure this equipment in a manner not specied in this
manual.
Pressure
There are two reasons for operating at elevated pressures when performing a
ltration analysis.
1. To test uids at temperatures above the boiling point, the vessel must
be pressurized, which in turn elevates the vapor pressure (boiling
point) so that the uid remains liquid and does not turn to steam.
2. If testing for drilling applications, pressurization will more approximate
down-hole conditions, enabling the technician to match both bottom
hole temperatures and pressures.
HTHP Filter Presses are pressurized either with Carbon Dioxide or with
Nitrogen gas. The OFITE 175 mL HTHP Filter Press, 170-00 Series, is small
enough to be portable, so it is usually pressurized with small (65 mm length)
Carbon Dioxide (CO2) bulbs, which contain 10 cm3 of CO2 gas and weigh 8
grams. These bulbs are pressurized to approximately 1,000 psi and they con-
tain plenty of carrier gas to run a complete 30 minute ltration test, if running
the standard API 500 psi differential test, which is usually 600 psi on the top
manifold and 100 psi on the bottom or back pressure.
For temperatures less than 200°F (93°C) a Back Pressure Receiver is not
required as the ltrate will not reach the boiling point. However when operat-
ing above the boiling point of water, a suitable Back Pressure Receiver is
required, otherwise the test uid will turn to steam and the test is invalidated.
The standard receiver tube supplied with the 175 mL HTHP Filter Press holds
15 mL of ltrate, so it is very important that the receiver outlet ball valve is
opened after approximately ten seconds of ltrate collection, as a uid with
poor ltration qualities can easily ll the 15 mL receiver tube almost instan-
taneously. If this happens the ltrate hopefully will all be ejected from the
safety bleeder valve, but if there is too much uid volume, the liquid may end
up inside the regulator rendering it useless which will require servicing by a
knowledgeable technician.
Temperature
Normally when one experiences a loss of pressure it is not due to a failure
of the metal alloy in the cell, but rather is a failure of the o-ring or elastomer,
which provides the seal. These o-ring may deform or melt under elevated
temperatures usually over 400°F (204°C) causing a pressurization failure of
the cell, which is often sudden and catastrophic. For example, if the valve
stem o-ring suddenly fails, then steam at 400°F and under great pressure
may shoot outward horizontally in one or several directions. A safety shield
should always be used when operating any HTHP Filter Press and especially
when one is going to extreme temperatures and pressures. Above 350°F
(176°C) All o-rings must be replaced after each and every test.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 9
Heating Jacket
Turn the thermostat knob on the front of the metal plate to change the tem-
perature of the heating jacket. The thermostat cover is marked from 1 to 10
and each whole number represents a separate temperature. Each lter press
will heat up a little differently so it is a good idea to mark with a pen directly
on the face-plate of the thermostat cover the temperature which corresponds
with each number. The actual temperature of the heating jacket is measured
with a stem thermometer and then the thermometer is placed into the cell
body.
The heating jacket will get very hot during the test so be careful not to touch
the outside of the jacket at any time during the heat-up period or the test. It is
especially easy to get burned when installing or removing the Back Pressure
Receiver as the technician is working in a very conned space. The power
cord is supplied for either 115 Volts or 230 Volts and due to the various types
of plug connectors around the world it may be necessary to change the plug
on the end of the power cord. The power cable is 8 feet in length and the
heating jacket should be placed no farther than 8 feet from the appropriate
electrical outlet. The heat up time will vary from one instrument to the next
and the pilot light will turn on when the set temperature has been reached.
The temperature of the uid inside the cell however will not be at the set tem-
perature, so always allow one hour of heating time for the uid, after the cell
has been fully inserted into the heating jacket.
Questo manuale è adatto per i seguenti modelli
1
Indice
Altri manuali OfiTE Apparecchiature di prova

OfiTE
OfiTE 170-00-2 Manuale utente

OfiTE
OfiTE 120-70 Manuale utente

OfiTE
OfiTE 150-50 Manuale utente

OfiTE
OfiTE PPT Manuale utente

OfiTE
OfiTE 120-700 Manuale utente

OfiTE
OfiTE 170-00-3 Manuale utente

OfiTE
OfiTE 150-50 Manuale utente

OfiTE
OfiTE 112-00 Manuale utente

OfiTE
OfiTE 112-00 Manuale utente

OfiTE
OfiTE 152-95 Manuale utente
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione