S217
Dew Point Transmitter

Dear Customer,
Thank you for choosing our product.
The operating instructions must be read in full and carefully observed
before you start up the device. The manufacturer cannot be held liable
for any damage which occurs as a result of non-observance or non-
compliance with this manual.
Should the device be tampered with in any manner other than a
procedure which is described and specified in the manual, the warranty
is void and the manufacturer is e empt from liability.
The device is destined e clusively for the described application.
SUTO offers no guarantee for the suitability for any other purpose.
SUTO is also not liable for consequential damage resulting from the
delivery, capability or use of this device.
2 S217

Table of contents
1 Safety Instructions......................................................................4
2 Registered Trademarks.................................................................6
3 Application..................................................................................6
4 Features.....................................................................................6
5 Technical Data.............................................................................7
6 Dimensional Drawing...................................................................8
7 Electrical Connection ..................................................................9
8 Modbus Output (S217-8 and S217-9)...........................................10
8.1 Modbus Information.............................................................10
8.2 Modbus Holding Register Table...............................................11
8.3 Interpretation of System Status.............................................12
9 Installation ...............................................................................13
9.1 Installation Requirements......................................................13
9.2 Installation Procedure ..........................................................14
9.2.1 Install with Measuring Chamber A699 3491.......................14
9.2.2 Install with By-pass Measuring Chamber A699 3493...........15
9.2.3 Install without Measuring Chamber...................................17
9.3 Remove the Dew Point Transmitter.........................................18
9.4 Order Information................................................................18
10 Calibration..............................................................................19
11 Maintenance............................................................................19
12 Disposal or Waste....................................................................19
S217 3

1 Safety Instructions
1 Safety Instructions
Please check if this instruction manual matches with the
product type
Please observe all notes and instructions indicated in this
manual. It contains essential information which must be
observed before and during installation, operation and
maintenance. Therefore this instruction manual must be read carefully
by the technician as well as by the responsible user / qualified
personnel.
This instruction manual must be available at the operation site of the
Dew Point Transmitter at any time. In case of any obscurities or
questions, regarding this manual or the product, please contact the
manufacturer.
WARNING!
Compressed air!
Any contact with quickly escaping air or bursting
parts of the compressed air system can lead to
serious injuries or even death!
•Do not e ceed the ma imum permitted pressure range (see
sensors label).
•Only use pressure tight installation material.
•Avoid that persons get hit escaping air or bursting parts of the
instrument.
•The system must be pressureless during maintenance work.
WARNING!
Voltage used for supply!
Any contact with energized parts of the product, may
lead to a electrical shock which can lead to serious
injuries or even death!
•Consider all regulations for electrical installations.
•The system must be disconnected from any power supply during
maintenance work.
•Any electrical work on the system is only allowed by authorized
qualified personal.
4 S217

1 Safety Instructions
ATTENTION!
Permitted operating parameters!
Observe the permitted operating parameters, any
operation exceeding this parameters can lead to
malfunctions and may lead to damage on the
instrument or the system
•Do not e ceed the permitted operating parameters.
•Make sure the product is operated in its permitted limitations.
•Do not e ceed or undercut the permitted storage and operation
temperature and pressure.
General safety instructions
•It is not allowed to use the product in e plosive areas.
•Please observe the national regulations before/during installation
and operation.
Remarks
•It is not allowed to disassemble the product.
•Always use spanner to mount the product properly.
ATTENTION!
Measurement values can be affected by malfunction!
The product must be installed properly and
frequently maintained, otherwise it may lead to
wrong measurement values, which can lead to
wrong results
Storage and transportation
•Make sure that the transportation temperature of the transmitter
is between -30 ... +70°C.
•For transportation it is recommended to use the packaging which
comes with the transmitter.
•Please make sure that the storage temperature of the transmitter
is between -10 ... +50°C.
•Avoid direct UV and solar radiation during storage.
•For the storage the humidity must be <95% rH, no condensation.
S217 5

2 Registered Trademarks
2 Registered Trademarks
SUTO® Registered trademark of SUTO iTEC
MODBUS® Registered trademark of the Modbus Organization,
Hopkinton, USA
3 Application
The SUTO Dew Point Transmitter S217-OEM provides reliable and long
term stable dew point monitoring in industrial applications. The newly
developed transmitter features improved signal and stability in
demanding industrial applications.
It’s designed for OEM applications in desiccant and refrigeration dryers.
The measured dew point is output via the loop-powered 4 ... 20 mA
signal. Parameters such as analogue output scaling, physical units, can
be set e factory.
4 Features
•Small size makes it ideal for dryer installations
•Measures dew points down to –60ºC Td
•2-wire loop powered sensor or 3-wire
•Modbus/RTU signal output (model depending)
•IP65 casing provides robust protection in rough industrial
environment
•Very fast response time ensures safe and reliable indication
whenever dew points are out of valid ranges
•Can be installed directly into dryers through G 1/2” thread
•High accuracy of 1 … 2ºC Td dew point
•Withstands condensation
6 S217

5 Technical Data
5 Technical Data
Measuring range
(model depending)
-60 ... +20°C Td
-50 ... +20°C Td
-20 ... +50°C Td
Accuracy * ±1ºC Td [-5 … +50ºC]
±2ºC Td [-60 … -5ºC]
Pressure range -0.1 ... 5.0 MPa
-0.1 ... 35 MPa (optional)
Power supply 12 ... 30 VDC
Measured gas Non-corrosive gases
Ambient temperature -20 ... +50°C
Ambient humidity 0 ... 100% rH
Transport temperature -30 ... +70°C
Response time t90
(@ 4 l/min)
-40°C Td → -20°C Td: 20 seconds
0°C Td → -40°C Td: 120 seconds
Output signal 4 ... 20 mA, 2-wire (S217-0 / S217-3 )
4 ... 20 mA, 3-wire (S217-4 / S217-5)
Modbus/RTU (S217-8 / S217-9)
Cable M8 connector, 1.8 m, open wires, 4 poles
Casing Process connection: stain-less steel
Casing: Al alloy
Cassification IP65
Process connection G ½” thread (ISO 228/1)
Sensor protection Stainless steel sinter filter pore size <30 µm
Approval CE
* Stated accuracy under following conditions:
•Ambient / process temperature 23 ± 3°C
•Ambient humidity <95% rH, no condensation
•Airflow > 2 l/min at sensor tip
S217 7

7 Electrical onnection
7 Electrical Connection
Pin assignment
Pin 1 Pin 2 Pin 3 Pin 4
2-wire SDI -VB +VB N/A
3-wire SDI -VB +VB +I
Modbus/RTU -D -VB +VB +D
Color Brown White Blue Black
Legend for pin assignment
SDI Digital signal (Internal use only)
-VB Negative supply voltage
+VB Positive supply voltage
+I 4 ... 20 mA (3-wire)
+D Positive data Modbus
-D Negative data Modbus
Schemas
S217 9

8 Modbus Output (S217-8 and S217-9)
8 Modbus Output (S217-8 and S217-9)
8 1 Modbus Information
The default settings of the Modbus interface are as follows:
Mode : RTU
Baud rate : 19200
Device address : last two digits of serial number
Framing / parity / stop bit : 8, N, 1
Response time : 1 second
Response delay : 0 ms
Inter-frame spacing : 7 char
Response message that the device returns to the master:
•Function code: 03
The information of the byte order is shown in the table below:
Byte
Order
Sequence Data
Type
1st 2nd 3rd 4th
1-0-3-2 Byte 1
(MMMMMMMM*)
Byte 0
(MMMMMMMM *)
Byte 3
(SEEEEEEE)
Byte 2
(EMMMMMMM *) FLOAT
1-0-3-2 Byte 1 Byte 0
LSB
Byte 3
MSB Byte 2 UINT32
INT32
1-0 Byte 1
MSB
Byte 0
LSB --- --- UINT16
INT16
1-0 Byte 1
XXX *
Byte 0
DATA --- --- UINT8
INT8
* S: Sign, E: E ponent, M: Mantissa, XXX: no value
Explanations of MSB and LSB
MSB MSB refers to Most Significant Byte first, which follows the Big-
Endian byte order.
For e ample, if the main system follows the MSB first order:
When the 4-byte floating number, in the order of Byte1-Byte0-
Byte3-Byte2, is received from the slave (sensor), the master
must change the byte order to Byte3-Byte2-Byte1-Byte0 for the
correct display of the value.