Trace2O HydroCheck HC1000 Manuale utente

HydroCheck®
HC1000
Multi-Parameter
Electrochemical
Water Tester
Instruction Manual

2
Contents
1.1 Introduction.........................................................................................................4
1.2 Unpacking the Meter...........................................................................................4
1.3 List of Accessories..............................................................................................4
2.1 Display................................................................................................................5
2.2 Keypad................................................................................................................6
2.3 Connectors .........................................................................................................7
2.4 Connecting the Sensors......................................................................................7
2.5 Inserting/Replacing the Batteries........................................................................9
2.6 Refilling Electrolyte Solution for DO Probe .......................................................10
3.1 Power On/Off.................................................................................................... 11
3.2 pH Measurement ..............................................................................................11
3.3 ORP Measurement...........................................................................................11
3.4 Conductivity/TDS/Resistivity Measurement......................................................12
3.5 Salinity Measurement .......................................................................................12
3.6 Dissolved Oxygen Measurement......................................................................12
3.6 Hold Function....................................................................................................14
3.7 Storing and Recalling Data from Memory.........................................................14
4.1 Setup Menu ......................................................................................................15
4.2 Adjusting the settings:.......................................................................................20
4.3 Exiting setup mode: ..........................................................................................20
4.4 Setting the Date and Time ................................................................................21
4.5 Temperature Compensation..............................................................................22
4.6 Automatic Temperature Compensation:............................................................22
4.7 Manual Temperature Compensation:................................................................22
5.1 pH Calibration...................................................................................................23
5.2 Single Point Calibration:....................................................................................23
5.3 Multi-Point Calibration:......................................................................................24
5.4 pH Calibration with User-Defined Buffers: ........................................................24
5.5 pH Calibration Report .......................................................................................25
6.1 ORP Calibration................................................................................................26
6.2 ORP Calibration Report....................................................................................26

3
7.1 Conductivity Calibration....................................................................................28
7.2 Single Point Calibration:....................................................................................28
7.3 Multi-Point Calibration:......................................................................................29
7.4 Manual Calibration:...........................................................................................30
7.5 Conductivity Calibration Report ........................................................................31
8.1 DO Calibration in % Saturation Mode...............................................................32
8.2 100% saturation calibration:..............................................................................32
8.3 2 Point Calibration: ...........................................................................................32
9.1 DO Calibration in mg/L or ppm Mode................................................................34
9.2 2 Point Calibration: ...........................................................................................34
10.1 Temperature Calibration..................................................................................36
11.1 pH Electrode Care and Maintenance..............................................................37
11.2 ORP Electrode Care and Maintenance...........................................................38
11.3 Conductivity Electrode Care and Maintenance ...............................................39
11.4 DO Probe Care and Maintenance...................................................................39
12.1Troubleshooting...............................................................................................40
13.1 Appendix 1: How to calculate the temperature coefficient...............................41
13.2 Appendix 2: How to calculate the TDS conversion factor................................42
13.3 Appendix 3: Pressure vsAltitude Table...........................................................43
14.1 Warranty .........................................................................................................44

4
1.1 Introduction
Thank you for selecting the Trace2o® HydroCheck HC1000 portable multiparameter water
quality meter. This manual provides instructions to guide the user in normal operation,
diagnostics and maintenance of the instrument. Please read the instructions carefully before
use.
1.2 Unpacking the Meter
Before unpacking, ensure that the current work environment meet following conditions.
● Relative humidity is less than 80%.
● Ambient temperature is greater than 0oC and less than 60oC.
● No potential electromagnetic interference.
The following list describes the standard accessories of the meter.After the unpacking, please
check all accessories are complete. If any are damaged or missing, please contact nearest
distributor.
1.3 List of Accessories
● pH Electrode
● Conductivity Electrode
● Dissolved Oxygen Electrode
● pH Buffer Solutions (pH4.01, 7.00)
● Conductivity Calibration Solution (1413µS/cm)
● Batteries

5
2.1 Display
The Trace2o® HydroCheck® HC1000 is equipped with a backlit LCD display that shows
measured values, as well as mode and status indicator graphics.
INDEX:
Measurement
mode
Battery indicator
Calibration mode
Stable indicator
Setup mode
Hold indicator
Memory mode
Calibration Due Reminder
Electrode slope
indicator
Automatic Temperature
Compensation:

6
2.2 Keypad
The Trace2o® HydroCheck® HC1000 has a tactile membrane keypad. Words and symbols
describe functions of each key. The direction keys are multi-function left/right and up/down
keys.
INDEX:
KEY
DESCRIPTION
● Power the meter ON/OFF.
● Holds the current measured value; press the key again to resume
measuring.
● In the calibration or setting mode, returns to measurement mode.
mode
● Toggles between available measurement modes (pH, mV, ORP,
Ion, Conductivity, TDS, Salinity, Resistivity, DO).
● Press and hold to enter temperature setting mode.
cal
● Press to enter the calibration mode.
● Press and hold to enter the setup menu.
m+ / ▲
● Press to store current measured value.
● Press in setup mode to scroll up through menu.
● Press in temperature setting mode to increase the setting value.
mR / ▼
● Press to view calibration report or stored data.
● Press in setup mode to scroll down through menu.
● Press in temperature setting mode to decrease the setting value.
enter
● Confirms the calibration, setting value or displayed option.
● Press and hold to turn On/Off the backlight.

7
2.3 Connectors
The Trace2o® HydroCheck® HC1000 has three connectors for connecting the various types of
sensors.
CONNECTOR
FUNCTION
Waterproof BNC
Connector
pH electrode (or optional ORP or Ion Selective Electrodes)
Waterproof 6-pin
connector
Dissolved oxygen electrode
Waterproof 3-pin
connector
Conductivity electrode
2.4 Connecting the Sensors
Take out the sensor from the carrying case. Ensure the connectors are clean and dry before
connecting.
pH ELECTRODE
● Insert the BNC connector on the pH electrode into corresponding connector on the
instrument. Once pushed firmly into place, tighten the waterproof locking collar.

8
CONDUCTIVITY ELECTRODE
● Insert the 3-pin connector on the conductivity electrode into the corresponding connector on
the instrument, ensuring that the lugs are aligned. Once pushed firmly into place, tighten the
waterproof locking collar.
DISSOLVED OXYGEN ELECTRODE
● Insert the 6-pin connector on the temperature probe into the corresponding connector on
the instrument, ensuring that the lugs are aligned. Once pushed firmly into place, tighten the
waterproof locking collar.
● Once connected, DO NOT pull on the cables.

9
2.5 Inserting/Replacing the Batteries
The HC1000 is supplied pre-installed with the requisite battery.
The HC1000 requires 1 x 9V (PP3) battery.
To replace the battery, follow the below procedure:
1. Remove the battery cover from the rear of the meter.
2. Disconnect the snap connector and remove the battery.
3. Note the polarity and attach the snap connector to the appropriate terminals on the new
battery.
4. Ensure that the white waterproofing seal is correctly located, and replace the battery cover.
When batteries are depleted, the meter allows you to use the USB cable connected to
computer as a temporarily power supply.

10
2.6 Refilling Electrolyte Solution for DO Probe
The Dissolved Oxygen electrode requires an electrolyte solution which must be kept topped up
during normal use. To refill, follow the below procedure:
1. Unscrew the membrane cap from the bottom of the dissolved oxygen probe.
2. Fill the membrane cap halfway with electrolyte solution.
3. Screw the membrane cap onto the probe, excess electrolyte will drain out.
4. Be sure the cathode of probe makes contact with membrane cap, the electrolyte solution in
membrane cap should be without an air bubble.
Prior to Use
Remove the protective cap from the bottom of the sensor.
…………Cathode
Indice
Altri manuali Trace2O Apparecchiature di prova
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione