VITO OILTESTER Manuale utente

OILTESTER - Deep-frying Oil Tester
Instruction manual

2

1 Contents
3
1 Contents
1 Contents ................................................................................................... 3
2 Safety and the environment .................................................................... 4
2.1. About this document ........................................................................ 4
2.2. Ensure safety ................................................................................... 5
2.3. Protecting the environment .............................................................. 5
3 Specifications .......................................................................................... 6
3.1. Use .................................................................................................. 6
3.2. Technical data ................................................................................. 6
4 Product description ................................................................................. 8
4.1. Overview.......................................................................................... 8
4.2. Basic properties ............................................................................. 10
5 First steps .............................................................................................. 11
5.1. Commissioning .............................................................................. 11
5.2. Getting to know the product ........................................................... 13
5.2.1. Switching the instrument on/off ...................................................................... 13
5.2.2. Description of important functions and displays ............................................. 14
5.2.2.1. Alarm indicator ................................................................................ 14
5.2.2.2. Setting the TPM limit values ............................................................ 14
5.2.2.3. Hold function ................................................................................... 15
5.2.2.4. Auto-Hold function ........................................................................... 15
5.2.2.5. Auto-off function .............................................................................. 15
5.2.2.6. Battery capacity ............................................................................... 16
5.2.3. Configuring the instrument ............................................................................. 16
5.2.4. Locking / unlocking configurations ................................................................. 18
6 Using the product .................................................................................. 20
6.1. General measurement information ................................................ 20
6.2. Carrying out measurements .......................................................... 21
6.3. Function test .................................................................................. 23
7 Maintaining the product ........................................................................ 24
7.1. Changing batteries ........................................................................ 24
7.2. Cleaning the sensor ....................................................................... 24
7.3. Cleaning the housing ..................................................................... 25
7.4. Cleaning the plastic case ............................................................... 25
7.5. Calibrating/adjusting the instrument .............................................. 25
8 Tips and assistance ............................................................................... 26
8.1. Questions and answers ................................................................. 26

2 Safety and the environment
4
2 Safety and the environment
2.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Warnings
Always pay attention to information that is marked by the following
warnings with warning pictograms. Implement the specified
precautionary measures.
Representation Explanation
WARNING Indicates potential serious injuries
NOTICE indicates circumstances that may lead to
damage to the products
Symbols and writing standards
Represen-
tation
Explanation
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu Elements of the instrument, the instrument displays
or the program interface.

2 Safety and the environment
5
[OK] Control keys of the instrument or buttons of the
program interface.
... | ... Functions/paths within a menu.
“...” Example entries
2.2. Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your
area when performing the measurements.
> Temperatures given on probes/sensors relate only to the
measuring range of the sensors. Do not expose handles and
feed lines to any temperatures in excess of 70 °C unless they
are expressly permitted for higher temperatures.
> Do not perform contact measurements on non-insulated, live
parts.
> Transport and store the instrument exclusively in the aluminium
case provided in order to avoid damage to the sensor.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Vito AG.
2.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Vito AG for disposal.

3 Specifications
6
3 Specifications
3.1. Use
The OILTESTER is a handy measuring instrument for the fast
inspection of deep-frying oils and is designed in equal measure for
left and right handers.
The TPM value (total polar materials) enables a statement on the
ageing of deep-frying oils due to the effects of heat.
The following measuring tasks can be performed:
• Display temperature of the deep-frying oil:
Indicator for correct setting of the deep-fryer, inspection of
integrated temperature displays.
• Display TPM value:
Indicator for the ageing of the deep-frying oil.
The sensor works on a capacitive basis and determines the total
amount of polar materials in % as the reading.
The free fatty acids that are determined above all for the evaluation
of unloaded oils (rancidity) cannot be detected.
The temperature of the deep-frying oil to be measured must
be at least 40 °C. The maximum operating temperature is
200 °C.
The sensor and the probe tube are designed to come into
contact with oil used in deep fat fryers for the typical
duration of a spot check measurement. The materials used
in these components meet the relevant requirements in the
Regulation (EC)1935/2004.
3.2. Technical data
Feature Values
Measuring range Temperature: 40.0 to 200.0 °C /
104.0 to 392.0°F
TPM: 0 to 40%
Accuracy Temperature: ± 1.5 °C
TPM1: ±2% (40.0 to 190.0 °C / 104 to 374 °F)
1 typical, referred to internal reference, at an ambient temperature of 25 °C.

3 Specifications
7
Feature Values
Resolution Temperature: 0.1 °C / 0.1 °F
TPM: 0.5%
Power supply Batteries: 2x micro (type AAA)
Battery (micro AAA) Designation according to IEC: LR03
Chem. composition: Zn-MnO2 (alkaline)
Battery life at 20 °C Approx. 25 hrs continuous operation
(corresponds to 500 measurements) without
display illumination
Temperature
sensor
PTC
TPM sensor Capacitive sensor
Operating
temperature
0 to 50 °C / 32 to 122 °F
Ambient humidity 0 to 90%RH
Storage/transport
temperature
-20 to 70 °C / -4 to 158 °F
Display LCD, 2-line, display illumination
Weight 255 g
Housing material Top part: ABS
Lower part: ABS-PC glass fibre 10%
Dimensions Approx. 50 mm x 170 mm x 300 mm
(WxHxD)
TPM response time Approx. 30 secs
Protection class IP65
Warranty 24 months
EC Directive 2014/30/EC

4 Product description
8
4 Product description
4.1. Overview
1 Display
2 Control keys
3 Battery compartment
4 Probe shaft
5 Oil quality (%TPM) and temperature sensor
6 Min. immersion depth
7 Max. immersion depth

4 Product description
9
Items shown on the display:
Items shown Function/property
200 (flashing
temperature value
> 200 °C)
Temperature measuring range exceeded
40 (flashing
temperature value
< 40 °C)
Temperature measuring range undershot
Alarm Alarm indicator activated
PIN Configuration mode locked
Battery capacity 100 %
Battery capacity 66 %
Battery capacity 33 %
Battery capacity < 10 %
Alarm Upper TPM limit value exceeded
Alarm Lower TPM limit value exceeded
Hold Readings held (manually)
Auto-Hold Readings held (automatically)
°C / °F Temperature in °C or °F
Important display messages
Item shown on the
display
Explanation
000 lights up Instrument is ready to measure, sensor is not
in oil.
Reading >190
flashes
Measured temperature is above 190 °C
(374 °F). The reading flashes in the range
from 190.1 °C (374 °F) to 200 °C (392 °F).

4 Product description
10
Control keys
Buttons Function/property
[] • Switch instrument on/off
• Configure instrument
[Hold] • Hold measuring values manually
• Switch to the measuring mode
• Configure instrument
[] • on/off; yes/no - Configure instrument
[] • on/off; yes/no - Configure instrument
Battery compartment
4.2. Basic properties
Power supply
The power is supplied to the instrument via two micro batteries
(type AAA). The batteries are included in the delivery.
Indice
Altri manuali VITO Apparecchiature di prova
Manuali Apparecchiature di prova popolari di altre marche

SMART
SMART KANAAD SBT XTREME 3G Series Manuale utente

Agilent Technologies
Agilent Technologies BERT Serial Manuale utente

Agilent Technologies
Agilent Technologies N3280A Manuale utente

Vernier
Vernier Go Direct Voltage Manuale utente

Lifeloc
Lifeloc R.A.D.A.R. Manuale utente

Fluke
Fluke T5-600 Istruzioni per l'uso e l'installazione