Waters ACQUITY RDa Manuale operativo

ACQUITY RDa Detector
Overview and Maintenance Guide
715005003
Version 04 (Previously released as Revision C)
Copyright © Waters Corporation 2021
All rights reserved

General information
Copyright notice
© 2021 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IN
IRELAND. ALL RIGHTS RESERVED. THIS DOCUMENT OR PARTS THEREOF MAY NOT BE
REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHER.
The information in this document is subject to change without notice and should not be construed
as a commitment by Waters Corporation. Waters Corporation assumes no responsibility for any
errors that may appear in this document. This document is believed to be complete and accurate
at the time of publication. In no event shall Waters Corporation be liable for incidental or
consequential damages in connection with, or arising from, its use. For the most recent revision
of this document, consult the Waters website (www.waters.com).
Trademarks
ACQUITY™ is a trademark of Waters Corporation.
RDa™ is a trademark of Waters Corporation.
THE SCIENCE OF WHAT'S POSSIBLE™ is a trademark of Waters Corporation.
UNIFI™ is a trademark of Waters Corporation.
Victrex PEEK™ is a trademark of Victrex PLC.
Viton® is a registered trademark of DuPont Performance Elastomers LLC.
Waters™ is a trademark of Waters Corporation.
waters_connect™ is a trademark of Waters Corporation.
Waters Quality Parts™ is a trademark of Waters Corporation.
All other trademarks are property of their respective owners.
Customer comments
Waters’ Customer Experience and Knowledge Management organization invites you to report
any errors that you encounter in this document or to suggest ideas for otherwise improving it.
Help us better understand what you expect from our documentation so that we can continuously
improve its accuracy and usability.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page ii

We seriously consider every customer comment we receive. You can reach us at
Contacting Waters
Contact Waters with enhancement requests or technical questions regarding the use,
transportation, removal, or disposal of any Waters product. You can reach us via the Internet,
telephone, fax, or conventional mail.
Contact method Information
www.waters.com The Waters website includes contact
information for Waters locations worldwide.
iRequest Request technical support for your Waters
instruments and software. iRequest is a free
and secure Web service that allows you to
obtain quick answers from our online
Knowledge Base, request support and service
for Waters instruments and software, or
schedule a planned service activity.
Note: In areas managed by authorized
distributors, iRequest may not be available.
Contact your local distributor for more
information.
Telephone and fax From the USA or Canada, phone
800-252-4752 or fax 508-872-1990.
For other locations worldwide, phone and fax
numbers appear in the Waters Local Offices
information.
Conventional mail Waters Corporation
Global Support Services
34 Maple Street
Milford, MA 01757
USA
Legal manufacturer
Waters Corporation
34 Maple Street
Milford, MA 01757
USA
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page iii

Safety considerations
Some reagents and samples used with Waters instruments and devices can pose chemical,
biological, or radiological hazards (or any combination thereof). You must know the potentially
hazardous effects of all substances you work with. Always follow Good Laboratory Practice
(GLP), and consult your organization’s standard operating procedures as well as your local
requirements for safety.
Safety hazard symbol notice
The symbol indicates a potential hazard. Consult the documentation for important
information about the hazard and the appropriate measures to prevent and control the hazard.
Considerations specific to the device
Power cord replacement hazard
Warning: To avoid electric shock, observe these precautions:
• The power cord functions as the safety disconnect device. Position the equipment
so that you can reach the power cord easily.
• Use SVT-type power cords in the United States and HAR-type power cords, or
better, in Europe. For requirements elsewhere, contact your local Waters distributor.
• Do not replace power cords with inadequately rated power cords.
• Inspect the power cords for damage and replace them if necessary.
• Connect each module of a system to a common ground.
• Power-off and unplug a system module or stand-alone device before performing
any maintenance operation on it.
Flammable solvents hazard
Warning: To prevent the ignition of flammable solvent vapors in the enclosed space of
a mass spectrometer’s ion source, ensure that these conditions are met:
• Nitrogen flows continuously through the source.
• A gas-fail device is installed to interrupt the flow of LC solvent should the nitrogen
supply fail.
• The nitrogen supply pressure does not fall below 400 kPa (4 bar, 58 psi) during an
analysis requiring the use of flammable solvents.
The gas-fail device for the RDa Detector is the Stop Flow cable connected to the LC. If the
nitrogen pressure falls below 400 kPa (4 bar, 58 psi), the instrument sends a stop-flow signal to
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page iv

the LC. You must connect the Stop Flow cable to the LC correctly so that if the gas supply fails,
solvent does not flood the mass spectrometer's source and damage the device.
High temperature hazard
Warning: To avoid burn injuries, exercise care when handling the components of the
source enclosure heated to high temperatures. Wait until the hot components have
sufficiently cooled before you handle them.
Mass detector high temperature hazard
Source components
Source enclosure
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page v

High-voltage hazard
Warning: To avoid electric shock, observe these precautions:
• Do not remove the mass spectrometer’s protective panels. The components they
cover are not user-serviceable.
• When the instrument is in Operate mode, avoid touching the areas marked with the
high voltage warning symbol. To touch external areas marked with the symbol, first
put the instrument in Power Save mode.
Hazards associated with removing an instrument from service
Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
When you remove the instrument from use to repair or dispose of it, you must decontaminate all
of its vacuum areas. These are the areas in which you can expect to encounter the highest levels
of contamination:
• Source interior
• Waste tubing
• Exhaust system
• Rotary pump oil (where applicable)
The need to decontaminate other vacuum areas of the instrument depends on the kinds of
samples the instrument analyzed and their levels of concentration. Do not dispose of the
instrument or return it to Waters for repair until the authority responsible for approving its removal
from the premises specifies the extent of decontamination required and the level of residual
contamination permissible. That authority must also prescribe the method of decontamination to
be used and the appropriate protection for personnel undertaking the decontamination process.
To avoid contamination by carcinogens, toxic substances, or biohazards, you must wear
chemical-resistant gloves when handling or disposing of used oil.
Bottle placement prohibition
Warning: To avoid injury from electrical shock or fire, and damage to the equipment, follow
these guidelines:
• Do not expose the workstation or ancillary equipment to dripping or splashing liquids.
• Do not place objects filled with liquid, such as solvent bottles, on top of the workstation or
ancillary equipment.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page vi

Electrical power safety notice
Do not position the device so that it is difficult to disconnect the power cord.
Equipment misuse notice
If equipment is used in a manner not specified by its manufacturer, the protection provided by the
equipment may be impaired.
Safety advisories
Consult the "Safety advisories" appendix in this publication for a comprehensive list of warning
advisories and notices.
Operating the device
When operating the device, follow standard quality-control (QC) procedures and the guidelines
presented in this section.
Applicable symbols
The following symbols can be present on the device, system, or packaging.
Symbol Definition
Manufacturer
Date of manufacture
Confirms that a manufactured product complies
with all applicable European Community
directives
UK Conformity Assessed marking confirms that
a manufactured product is in conformity with
the applicable requirements for products sold
within Great Britain
Australia EMC compliant
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page vii

Symbol Definition
Confirms that a manufactured product complies
with all applicable United States and Canadian
safety requirements
Environmentally friendly use period (China
RoHS): indicates the number of years from the
date of manufacture until the product, or
components within the product, are likely to be
discarded or degrade into the environment
Consult instructions for use
Alternating current
Electrical and electronic equipment with this
symbol may contain hazardous substances and
should not be disposed of as general waste
For compliance with the Waste Electrical and
Electronic Equipment Directive (WEEE)
2012/19/EU, contact Waters Corporation for the
correct disposal and recycling instructions
For indoor use only
No pushing
10kg
max
Indicates the maximum load you can place on
that item (for example, 10kg)
Serial number
REF
Part number, catalog number
Audience and purpose
This guide is for novice users and assumes no knowledge of liquid chromatography or mass
spectrometry principles. It provides an overview of the instrument and explains how to install it,
prepare it for operation, and maintain it.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page viii

Intended use of the ACQUITY RDa Detector
The ACQUITY RDa Detector is a time-of-flight mass detector with an ESI source, and is intended
for use as an analytical tool. It provides accurate mass data with minimal interaction from the
user. The LC/MS ESI instrument is not intended for use in diagnostic applications.
Calibrating
When calibrating mass spectrometers, consult the instrument’s online Help system for calibration
instructions.
Quality control
Routinely run three QC samples that represent subnormal, normal, and above-normal levels of a
compound. If sample trays are the same or very similar, vary the location of the QC samples in
the trays. Ensure that QC sample results fall within an acceptable range, and evaluate precision
from day to day and run to run. Data collected when QC samples are out of range might not be
valid. Do not report these data until you are certain that the instrument performs satisfactorily.
EMC considerations
FCC radiation emissions notice
Changes or modifications not expressly approved by the party responsible for compliance, could
void the user's authority to operate the equipment. This device complies with Part 15 of the FCC
Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
Canada spectrum management emissions notice
This class A digital product apparatus complies with Canadian ICES-001.
Cet appareil numérique de la classe A est conforme à la norme NMB-001.
ISM classification: ISM group 1 class A
This classification was assigned in accordance with CISPR 11 Industrial Scientific and Medical
(ISM) instruments requirements.
Group 1 products apply to intentionally generated and/or used conductively coupled radio-
frequency energy that is necessary for the internal functioning of the equipment.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page ix

Class A products are suitable for use in all establishments other than residential locations and
those directly connected to a low-voltage power supply network supplying a building for domestic
purposes.
There may be potential difficulties in ensuring electromagnetic compatibility in other environments
due to conducted and radiated disturbances.
This equipment complies with the emission and immunity requirements described in the relevant
parts of IEC/EN 61326: Electrical equipment for measurement, control, and laboratory use —
EMC requirements.
EMC emissions
Do not use the equipment in close proximity to sources of strong electromagnetic radiation (for
example, unshielded intentional RF sources). The radiation can interfere with the equipment’s
proper operation.
Safe disposal
Contact [email protected] with any questions or concerns regarding the proper handling or
disposal.
Dispose of Waters instrumentation products in accordance with applicable requirements and best
practices as described below.
• Follow appropriate procedures for flushing the instrument’s fluid paths of any hazardous
samples or solvents.
• Waters instruments are subject to European Union’s Waste Electrical and Electronic
Equipment (WEEE) and Restriction of Hazardous Substances (RoHS) Directives. According
to these directives, do not dispose of instruments in the general waste stream. Similar “e-
waste” laws also apply in other jurisdictions. In all cases, ensure that a certified electronics
recycler processes end-of-life instruments.
• Some Waters instruments use batteries, mercury-containing lamps, or other replaceable
components during the life span of the instrument. Handle such materials in accordance with
local laws governing their processing and safe disposal.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page x
Indice
Altri manuali Waters Sensore di sicurezza

Waters
Waters 2487 Manuale utente

Waters
Waters 432 Manuale

Waters
Waters ACQUITY QDa Manuale operativo

Waters
Waters 2487 Dual A Manuale

Waters
Waters 996 Manuale

Waters
Waters ACQUITY ELS Manuale utente

Waters
Waters 2998 Manuale

Waters
Waters TQ Detector Manuale utente

Waters
Waters 2465 Manuale utente

Waters
Waters 3100 Manuale

Waters
Waters 2475 Manuale

Waters
Waters ACQUITY Manuale operativo

Waters
Waters 2489 Manuale

Waters
Waters 2489 Manuale operativo

Waters
Waters 2996 Manuale

Waters
Waters 2414 Manuale operativo

Waters
Waters ACQUITY UPLC Manuale operativo

Waters
Waters Micromass ZQ 2000 Manuale

Waters
Waters 2489 Manuale utente

Waters
Waters ACQUITY UPC2 Manuale operativo