Accurate Technology ProPanel-HD Manuale di istruzioni

ProPanel-HD shown with optional transmitter and steel brace.
SERIAL # _______________
ProPanel-HD
(Systems with Firmware version 4)

ProPanel-HD (FW4), Revised August 2020 Page 2 of 24
WARRANTY
Accurate Technology, Inc. warrants the ProScale Systems against defective
parts and workmanship for 3 years commencing from the date of original
purchase. Upon notification of a defect, Accurate Technology, Inc., shall have
the option to repair or replace any defective part. Such services shall be the
customer's sole and exclusive remedy. Expenses incidental to repair,
maintenance, or replacement under warranty, including those for labor and
material, shall be borne by Accurate Technology, Inc. (Including freight or
transportation charges during the first 30 days).
Except as expressly provided in this warranty, Accurate Technology, Inc. does
not make any warranties with respect to the product, either expressed or implied,
including implied warranties of merchantability or fitness for a particular purpose,
except as expressly provided in this agreement.
Accurate Technology, Inc. shall not be liable for any special, incidental, or
consequential damages or for loss, damage or expense directly or indirectly
arising from the customer's use of or inability to use the equipment either
separately or in combination with other equipment, or for personal injury or loss
or destruction of other property, or from any other cause.
To request repair work, (either warranty qualified parts or not) contact Accurate
Technology, Inc. directly by phone, fax, or e-mail. A Returned Merchandise
Authorization (RMA) number is required before returning a product for repair.

ProPanel-HD (FW4), Revised August 2020 Page 3 of 24
READOUT SERIAL # _______________
DATE OF MANUFACTURE _______________
FACTORY CONFIGURED PARAMETERS WHICH ARE
SPECIFIC TO THIS GAGE:
Programming control
Value
PARAMETER 2 (Reading direction)
1 (positive)
PARAMETER 11 (Available display units)
2 (no fractions)
PARAMETER 13 (Linear Correction Value)
PARAMETER 14 (ProScale Technology)
0 (Inductive)
PARAMETER 30 (Number of offsets used)
PARAMETER 31 (Offset Value #1)
PARAMETER 32 (Offset Value #2)
PARAMETER 33 (Offset Value #3)
PARAMETER 39 (Non-Linear correction used?)
1 (yes)
PARAMETER 40 (Non-Linear correction Interval)
5.0000 inches
PARAMETER 41 (Temperature correction enabled?)
1 (yes)
See www.proscale.com/700-1600-246 for more details about these
settings.

ProPanel-HD (FW4), Revised August 2020 Page 4 of 24
Table of Contents
INTRODUCTION ..........................................................................................................5
SPECIFICATIONS........................................................................................................6
CALIBRATION ..........................................................................................................6
INITIAL SET-UP ..........................................................................................................7
OUTSIDE MEASURING ...............................................................................................8
INSIDE MEASURING ...................................................................................................9
HOLE-TO-HOLE MEASUREMENTS .........................................................................10
EDGE TO HOLE MEASUREMENTS..........................................................................11
COMPARING DIAGONALS .......................................................................................12
GENERAL MAINTENANCE.......................................................................................13
BATTERY REPLACEMENT.......................................................................................13
PRIMARY KEY FUNCTIONS .....................................................................................14
SUPPLEMENTAL KEY FUNCTIONS.........................................................................15
WHAT DO ALL THE SYMBOLS MEAN? ..................................................................16
SPC PORT, SPECIAL MODES, AUTOMATIC ON/OFF ............................................18
PROGRAMMING MODE ............................................................................................19
COMMON ERROR MESSAGES................................................................................20
ACCESSORIES ........................................................................................................21
FREQUENTLY ASKED QUESTIONS ........................................................................22

ProPanel-HD (FW4), Revised August 2020 Page 5 of 24
Introduction
This manual includes operation and use information for ProPanel-HD systems
that use a ProScale General Purpose Enhanced Readout with operating
firmware (F/W) of 4.000C and higher.
(The firmware version is displayed on power-up, i.e. P4.020C.)
NOTE: Some pictures used in this manual may not represent your actual
product’s colors. The terms ProPanel, ProPanel-HD and ProPanel-HD-E
are used interchangeably.
ProPanel™ is a general-purpose portable measuring tool. It is ideal for making
inside, outside, and corner-corner measurements. In addition, edge-to-hole
and hole-to-hole measurements are possible with the optional Cone Set.
ProPanel-HD has been designed using high quality anodized aluminum
extrusion and machined parts to provide the best accuracy and repeatability
over a wide variety of environments.
The measurement system used for ProPanel-HD is based on the ProScale®
Model 590 Digital Measuring System.
Cylinders are included; these are
used to make Corner-Corner
comparisons
Optional Cone Set is used to make
Hole-Hole and Edge-Hole measurements

ProPanel-HD (FW4), Revised August 2020 Page 6 of 24
Specifications
Measuring Range: Stock sizes to 195 inches range
Accuracy: ± .005in (±.13mm)
(When beam is properly supported)
Resolution Inches: 1, 2, 3, or 4 decimal places
Millimeters: 1 or 2 decimal places
Centimeters: 1, 2, or 3 decimal places
Fractions: 1/16ths, 1/32nds, or 1/64ths
Repeatability .001in or .01mm or .001cm
Operating Temp: 32 to 120
F (0 to 51
C)
Max. Slew Rate: 80 inches/sec. (2000mm/sec)
Power: 1 CR123 (3V Lithium battery)
Output Format: Mitutoyo Digimatic® SPC
Calibration
Each ProPanel-HD system is calibrated at factory when initially built. Systems
should be checked at least once per year for accuracy, and serviced as
needed. If you see a NoCo message, the readout is trying to show a dimension
that is outside the range of calibration. Close jaws and press the DATUM key. If
NoCo message does not clear, consult factory for assistance.

ProPanel-HD (FW4), Revised August 2020 Page 7 of 24
Initial Set-up
In most instances, the initial parameters such as a reference point, measurement units,
and resolution will be factory set. (See page 2 for these.)
UNITS: To change the measurement units, press the UNITS Key.
The readout will cycle thru:
•Decimal Inches
•Fractions –16ths of inch
•Fractions –32nd of inch
•Fractions –64th of inch
•Millimeters (or centimeters if enable)
DATUM: Close the jaws together and verify the readout shows zero. Press the DATUM
key if needed. The keypad can be locked if desired; this prevents accidental presses of
the PLUS, MINUS, or DATUM keys from affecting accuracy.
CUSTOMIZE: Many operations/features of the ProPanel system can be configured to
suit best way you will use the gage. See Programming Mode section for details about
the numerous controls/settings available.

ProPanel-HD (FW4), Revised August 2020 Page 8 of 24
Outside Measuring
Use the ProPanel-HD like a handheld caliper or T-square. Close the jaws
against the edges of the item being measured, keeping the jaws as square to
the edges as possible, and read the result on the readout.
Tips:
•Ensure the beam is not sagging during measurements.
•Try to use the full width of the jaws wherever possible.

ProPanel-HD (FW4), Revised August 2020 Page 9 of 24
Inside Measuring
1. Press the F1 key on the keypad. A small number 1will appear just above
the dimension.
2. The readout is now showing the Inside measurement.
Note: This offset value is factory programmed (see page 2), but can be
changed if needed
3. Using the upper or lower tips of the jaws, measure the inside dimension of
interest, and read the result on the readout.
Note: Videos showing how these measurements are made are posted at web
site: www.proscale.com/propanel/

ProPanel-HD (FW4), Revised August 2020 Page 10 of 24
Hole-to-Hole Measurements
(Using optional Cone Kit parts)
1. Close the jaws together and momentarily push
the DATUM key.
2. Install both full cones as shown using the supplied
3/16” wrench.
3. Press the F1 key until the #2 superscript appears
on the screen.
Notes:
•The tips of the cones are even with the edges of the jaws, so the value
displayed is the same as for outside measuring.
•Hole-to-hole measurements less than 1.5” are not possible.
4. Open the jaws and place the cones into the holes in the part. A slight
“wiggling” of the ProPanel will ensure the cones self-center into the holes.
5. Read the Hole-to-Hole distance on the readout.
6. If you need to recalibrate with the cones installed:
a. Press the DATUM key to clear the readout.
b. Close the moving carriage so the edge of the cones touch.
c. Press the F1 key until the #2 superscript appears.
d. The reading should equal 1.500 inches.
Note: Videos showing how these measurements are made are posted at web
site: www.proscale.com/propanel/
Indice
Altri manuali Accurate Technology Strumento di misura

Accurate Technology
Accurate Technology Digi Scale Model 10 Manuale utente

Accurate Technology
Accurate Technology ProScale 950 Series Manuale utente

Accurate Technology
Accurate Technology ProPanel Classic Manuale utente

Accurate Technology
Accurate Technology Pro-60 Manuale utente

Accurate Technology
Accurate Technology ProScale 19 Manuale di istruzioni

Accurate Technology
Accurate Technology DigiStop-4 Guida

Accurate Technology
Accurate Technology Digi-Stop Manuale utente