ACS contsys Flowcont UN Manuale utente

03.14
Operating Instructions
Main features
Flow sensor for conductive and non-conductive liquids
No moving parts, compact design
Process temperature up to 80 ° C, process pressure up to
10 bar
High chemical resistance due to seal-free sensor design
Large display with keypad
Integrated empty pipe detection
Easy to clean, hygienic version available (EHEDG-approved
and FDA compliant), CIP
Maintenance free ow sensor, saving maintenance costs
Adjustable ranges, reduces variants
Can be used for conducting and non-conducting liquids,
reduces variants and storage costs
Straight measuring tube reduces pressure loss, thus
saving on energy costs
Seal-free sensor increases process reliability and
availability
Flexible measuring system for all industries
Flowcont UN
Flow meter
Ultrasonic ow meter Flowcont UN
Non-contact ow measurement

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
2
You have purchased a high-grade and modern measuring device of ACS-CONTROL-SYSTEM GmbH.
We want to give thanks for your purchase and for your condence to us.
The actual technical manual includes instructions for installation, electrical connection and inauguration,
as well as the technical data of the device.
Modications, that answer the purpose of the technical progress, are reserved by
ACS-CONTROL-SYSTEM GmbH without prior notice.
If a question occurs, that can‘t be answered by the listed informations, please call on our technicians
team in Eggenfelden Tel: +49 8721/ 9668-0 or [email protected]
All rights reserved

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 3
Index
Principle of operation4
Safety notes 4
Applications 4
Commissioning5
Installation notes 5
Mounting 6
Electrical connection 8
Operation 11
Commissioning procedure 11
Display and menu 11
Overview of default settings 18
General instructions 18
Dimensional drawing 19
Technical data 22
Return 29
Disposal 29

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
4
Principle of operation
Swimming against the stream requires greater physical strength than swimming with the stream. This
is the simple fact on which ultrasonic ow measurement according to the phase difference process is
based. Two sensors positioned opposite one another send and receive ultrasonic signals in alternation.
When static media are used, both sensors receive the sent ultrasonic signal in phase, meaning without
phase difference. If owing media are used, there is a phase shift. The phase shift is different when
measured in the direction of current than when measured against the direction of current. The phase
difference is directly proportional to the speed of the ow. The ow volume is determined from the
speed of the ow and the known diameter of the pipeline.
Safety notes
• Read the instruction manual before commissioning.
• Connection, mounting, and setting only to be done by trained specialists.
• The Flowcont UN is not a safety module according to the EU Machinery Directive.
• Observe the national safety and work safety regulations.
• Repairs may only be carried out by the manufacturer. Changing or altering the device is not
permitted.
• Wiring work and opening and closing of electrical connections are only to be carried out in a de-
energized state.
• Incorrect handling or improper use can lead to malfunctions in your application.
• Flowcont UN can be used by the Pressure Equipment Directive (PED) of group 2 only for „non-
hazardous liquids.“
Applications
The Flowcont UN ow meter is particularly suited for measuring highly dynamic processes in pipelines.
The device measures liquids. It can be used in the following applications, to name but a few:
• Chemical supply for controlling, logistics and monitoring
• Filling processes in the food industry
• Production machines for controlling and monitoring formulae
• Valve control for dosing liquid volumes
• DI water supply
• Dynamic processes
The following featured comprise its performance characteristics:
• no moving parts, and thus no wear
• high reproducibility
• ease of cleaning
• tamper-proof
• compact design
• integrated empty pipe detection
• chemical resistance
• integrated display with membrane keyboard

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 5
Commissioning
Installation notes
An arrow on the measuring channel of the Flowcont UN indicates the ow measurement direction.
The measuring device must be installed such that the ow passes through the measuring tube in the
direction of the arrow.
TOP -> Outlet
Bottom -> Inlet
Optimal FLowcont UN mounting orientation
To detect gases as quickly as possible, the pipeline between the storage container and the FFU
should be as short as possible. An error-free measurement can only be ensured when the pipeline is
completely full and when no liquid is being released as gas. In the case of dosing applications, it may
be useful to position the FFU as close to the dosing valve as possible, because tubing tends to expand
in diameter depending on system pressure. This can lead to deviations.
To avoid cavitations, depending on the measured liquid it can be helpful to have sufcient
backpressure on the outlet Flowcont UN
If any solid particles are present, this can lead to measurement errors.
When pumps are used, the ow sensor must be mounted in the ow direction behind the pump (on
the delivery side) in order to ensure that adequate pressure will be available. The maximum pressure
rating of the FFU must be observed. In order to ensure correct volume ow measurement using the
FFU, in inlet and output paths should be straight and free of obstacles. The minimum inlet and output
paths are dependant on the nominal size:
DN10 DN15 DN20 DN25
Inlet zone 10 cm 30 cm 50 cm 80 cm
Outlet zone 0 cm 5 cm 10 cm 20 cm
Always make sure that the maximum torque of the nuts for the hydraulic connections is not overstep.
Please use the supplied seals. If you seal with teon tape that can result in to excessive tightening
torques. We commend a maximum torque depending on the diameter of:
DN10 DN15 DN20 DN25
Torque G 2 Nm 3 Nm 4 Nm 6 Nm
Caused by the material characteristics of PSU the Flowcont UN has a limited resistance
against UV rays Remind this when mounting Flowmax outdoor

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
6
Mounting
The measuring device is mounted in the pipeline using a process connection. In order to ensure
optimal gas venting, we recommend mounting the Flowcont UN vertically in the pipe. The device
should not be mounted behind an output valve, or it might run idle. To prevent outgassing and bubble
formation in the media during the ow measurement, the Flowcont UN should always be mounted on
the pressure side of the system pump.
If the device cannot be mounted vertically, the pipe in which the measuring device is located should
always be full. Preventing the formation of gas bubbles in the Flowcont UN measuring channel will aid
optimal measurements.
For applications with a „Clean Design“ when complete drain of the measuring tube is needed, we
recommend to mount the ow meter in vertical position. When ow meter is mounted horizontal
caused by the internal geometry of the device there might be a rest of liquid inside.
Attention: The process connections from the FFU are offset by 5 mm (see dimensional drawing on
page 17-19).
Mounted Flowcont UN must alway be mechanically stress free. Strong tension in the pipe system can
lead to damage to the devise. Shocks or mechanical loads can affect the measurement accuracy of
the measuring device. If the FFU must be additionally secured due to the presence of vibrations or
mechanical
motion, two mounting clamps can be mounted to the inlet and outlet of the measuring device.
Flowcont UN with mounting clamps

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 7
Application with hot uids
In applications where the Flowcont UN is subject to high heat (t < 60 °C), we recommend mounting
the measuring device and electronic housing beneath the measurement area. This allows the heat to
radiate upwards and reduces its effects on the device electronics.

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
8
Electrical connection
The Flowcont UN is connected via a ready-made cable socket with M12x1 plug, with 5 pins or 8
pins. Plug in the cable socket without power onto the sensor and screw it tight. Connect the cable in
accordance with its function. After application of the supply voltage, the sensor carries out a self-test
– in an installed state, the sensor is ready for operation after the self-test has been concluded (ca. 4
s) – the display shows the current measured value.
Connection diagram 5-pin variant
Pin Wire colour Function Description
1brown L+ Supply voltage
18 V DC … 30 V DC
2white Digital output Q1
Functions:
1. Pulse
2. Empty pipe detection
3. Dosing output
4. Threshold monitoring
5. Negative ow
Digital output Q1
Can be freely set in the
range of 0.1 to 3000 ml/
pulse
in stages of 0.1 ml/pulse,
npn or pnp transistor,
max. load 18 ... 30 V /
100mA.
Programmable output of 0 V
or 24 V when measuring
tube is empty.
Programmable output of 0 V
or 24 V when dosing
volume is reached.
Programmable output of 0 V
or 24 V when a preset
threshold is exceeded or not
reached.
Programmable output of 0 V
or 24 V during negative
ow.
3blue GND Supply ground: 0 V
4black Communication Communication interface

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected] 9
5grey Analog output QA
Functions:
1. Flow
2. Temperature
4 ... 20 mA; 0 ... 20 mA
For example: 0 l/min => 4
mA
36 l/min => 20 mA
Alarm => 3,5 mA
(dependent on set measuring
area)
Connection diagram 8-pin variant
Pin Wire colour Function Description
1brown L+ Supply voltage
18 V DC … 30 V DC
2white Digital output Q1
Functions:
1. Pulse
2. Empty pipe detection
3. Dosing output
4. Threshold monitoring
5. Negative ow
Digital output Q1
Can be freely set in the range of
0.1 to 3000 ml/
pulse in stages of 0.1 ml/pulse,
npn or pnp
transistor,
max. load 18 ... 30 V / 100 mA.
Programmable output of 0 V or
24 V when
measuring tube is empty.
Programmable output of 0 V or
24 V when a preset
threshold is exceeded or not
reached.
Programmable output of 0 V or
24 V during
negative ow.
3blue GND Supply ground: 0 V

ACS-CONTROL-SYSTEM GmbH l Lauterbachstr. 57 l D-84307 Eggenfelden l www.acs-controlsystem.de l [email protected]
10
4black Digital output Q2
Functions:
1. Empty pipe detection
2. Pulse
3. Threshold monitoring
4. Negative ow 5.Negativ
Fluss
Digital output Q2
Can be freely set in the range of
0.1 to 3000 ml/
pulse in stages of 0.1 ml/pulse,
npn or pnp transistor,
max. load 18 ... 30 V / 100 mA.
Programmable output of 0 V or
24 V when
measuring tube is empty.
Programmable output of 0 V or
24 V when a preset
threshold is exceeded or not
reached.
Programmable output of 0 V or
24 V during
negative ow.
5grey Analog output QA
Functions:
1. Flow
2. Temperature
4 ... 20 mA; 0 ... 20 mA
For example: 0 l/min => 4 mA
36 l/min => 20 mA
Alarm => 3,5 mA
(dependent on set measuring
area)
6pink Communication Communication interface
7purple Digital input IN1
1. Dosing input
2. Offsett compensation
3. Reset counter
4. Creeping ow
Digital input IN1
Starts the dosage by a rising
edge of 24V.
The Offset is set by a rising
edge of 24V.
Reset of the counter by a rising
edge of 24V.
Creeping suppression is
deactivated as long as
there are 24V at the input
8orange No function
Indice
Altri manuali ACS contsys Strumento di misura

ACS contsys
ACS contsys DAL-311-2 Series Manuale utente
ACS contsys
ACS contsys Sonicont USN4SP Manuale utente

ACS contsys
ACS contsys DAL-311x3x0S Manuale utente

ACS contsys
ACS contsys RCE 300 Manuale utente

ACS contsys
ACS contsys DAL-311 5 0S Series Manuale utente

ACS contsys
ACS contsys DAP-311 0 0S Series Manuale utente

ACS contsys
ACS contsys DAL-311x0x0S Manuale utente

ACS contsys
ACS contsys SAT Manuale utente