The Advanced® Bilirubin
Stat-Analy er™ Photometer
Models BR2 and BR2E for Neonatal samples
Service Manual
BR25SM Rev8 082416


3The Advanced®Model BR2 Service Manual
1. Introduction
Instrument pecifications . . . . . . . . . . . .7
General Description and Purpose . . . . . .8
The ervice Manual . . . . . . . . . . . . . . . .8
ervice Assistance . . . . . . . . . . . . . . . . .8
afety . . . . . . . . . . . . . . . . . . . . . . . . . . .8
afety/Regulatory Information . . . . . . . .9
ymbol Conventions . . . . . . . . . . . . . . .9
General Cautions . . . . . . . . . . . . . . . . . .9
ervice & Maintenance Cautions . . . . .10
Hazardous Material Cautions . . . . . . . .11
Regulatory Notices . . . . . . . . . . . . . . . .11
ymbol Conventions . . . . . . . . . . . . . .11
2. General Overview
Removing the Instrument Cover . . . . .15
Calibration . . . . . . . . . . . . . . . . . . . . . .16
Circuit Description . . . . . . . . . . . . . . . .17
BR2100 Main Board Description . . . .17
BR2950 Display Board Description . .19
Design Changes . . . . . . . . . . . . . . . . . .20
Replacement Parts . . . . . . . . . . . . . . . .21
upplies & Accessories . . . . . . . . . . . .21
3. Maintenance
Optical Lamp . . . . . . . . . . . . . . . . . . . .25
Optical Filters . . . . . . . . . . . . . . . . . . . .25
4. Troubleshooting
Internal elf-Diagnosis Voltage Checks .29
Troubleshooting Table . . . . . . . . . . . . .31
5. Ad ustments
Lamp Adjustment . . . . . . . . . . . . . . . . .37
Measuring Circuit Adjustment . . . . . . .39
Direct Zero Adjustment . . . . . . . . . . . .40
6. Replacement Instructions
Interference Filter et (BR2650) .B2P650
Lamp (BR2625) . . . . . . . . . . . . . .B2P625
Voltage election & Fuse
(70027/70028) . . . . . . . . . . . .B2P70011
7. Schematics
BR2003 chematic
BR2100 Main Board Assembly Drawing
BR2950 Display Board Assembly Drawing
APPENDICES
A: ymbol Definitions
B: Product Disposal and Recycling
C: ervice Log
Index
Table of Contents

4
he Advanced®Model BR2 Service Manual
Notes:

1. Introduction


7The Advanced®Model BR2 Service Manual
Model BR2 Specifications
Electrical:
Voltage: 100 to 120 VAC (1/4 A) or 200 to 240
VAC 1/8 A)
Frequency: 50 or 60 Hz
Fuses: 1/4-Amp time delay for 100-120 VAC;
1/8-Amp time delay for 200-240 VAC
Power: 15 Watts
STAT Mode: Power left on continuously
Test Ca ability: Total and Direct Bilirubin, both on
the same serum orplasma sample
Sam le Volume: 30 µL
Sam le Ca acity: Single sample, disposable cuvettes
Readout: Three-digit, seven-segment LED display
Units: mg/dL or µmol/L (S. . units for export
only; 1 mg/dL bilirubin = 17.1 µmol/L)
Range: 0 to 30 mg/dL
Resolution: ±0.1 mg/dL
Performance at Reference Conditions1
Linearity: Within ±0.2 mg/dL from zero to 30 mg/dL
(straight-line method)
Re eatability: Within ± 0.2 mg/dL over entire range
Drift:
Zero: Less than 0.2 mg/dL/hr after 30-minute
warm-up
Calibration: Less than 0.2 mg/dL/24 hr after 30-minute
warm-up
Performance Over O erating Conditions
Hemolysis: No interference from HbO2levels up to
600 mg/dL
Turbidity: Negligible interference from turbidity
associated with triglyceride (TGL) levels
up to 500 mg/dL
O erating Conditions
Tem erature: 15°C to 32°C (59°F to 90°F)
Room Humidity: 5 to 80% relative humidity;
(non-condensing)
Storage Tem erature: -40°C to +70°C
(-40°F to +160°F)
Test Time: nstantaneous for TOTAL Bilirubin, 2
minutes for D RECT Bilirubin
Dimensions inches centimeters
Width: 11 28
Depth: 13 33
Height: 7 18
Weight pounds kilograms
Net: 9 4
Shipping: 12 5.5
Warranty: One-year limited warranty on work-
manship and all parts except glass,
plastic and parts warranted by their
makers.
Certification:
Refer to Regulatory Notices (see page 11) for appli-
cable standards.
*CE marking only available on units delivered prior
to July 22, 2016.
1Reference Conditions: 20 to 25°C (68 to 77°F); 40 to 60% Relative Humidity; tolerances of reference or calibration
solutions excluded.
*

8
he Advanced®Model BR2 Service Manual
General Description and Purpose
The Advanced®Models BR2 and BR2E
Bilirubin Stat-Analyzer™ Photometers are
dual-arm, dual-wavelength, narrow-bandpass
photometers that measure Bilirubin (TBR)
and Azobilirubin (DBR) absorbance quantita-
tively. The digital readout is calibrated to
read bilirubin in milligrams per deciliter
(mg/dL) for the BR2, or micromoles per liter
(µmol/L) for the BR2E of serum or plasma.
The BR2 wavelengths are fixed and will not
shift during the life of the instrument.
he Service Manual
The BR2 Service Manual contains:
• instructions for performing repairs;
• guides for troubleshooting operational and
mechanical problems;
• circuit system illustrations; and
• information about replacement parts and
service repairs.
This manual is designed to assist
service tec nicians, and does not
imply a license to perform repairs
without proper qualifications.
Service Assistance
To contact Advanced Instruments Product
Service:
• Call (800) 225-4034 (toll-free within the
USA and Canada); after normal business
hours, dial extension 2191
• Call +US (781) 320-9000 (elsewhere)
• Or fax (781) 320-3669
When calling Advanced Instruments, have the
model and serial number from the label on
the back of the instrument, and a description
of the problem. Use a telephone as close to
the instrument as possible in order to facili-
tate making recommended diagnostic checks.
A service technician may assist in making
minor repairs over the phone, providing you
with recommended parts (or part numbers),
or may issue an authorization (RMA) to ship
the instrument for factory repair.
The RMA procedure is as follows:
1. Contact Advanced Instruments to obtain
an RMA.
2. Pack and ship the instrument in its origi-
nal carton. Do not send instrument sup-
ply items.
3. Prepay shipment to the factory.
Advanced Instruments cannot accept col-
lect shipments. Insure the shipment or
accept the damage risk.
Safety
Removing the instrument cover
exposes dangerous 200-250 and/or
100-130 AC voltages. Before remov-
ing the cover, make sure the instru-
ment is turned off and unplugged.

he Advanced®Model BR2 Service Manual
To reduce the risk of bodily injury,
electric shock, fire, and damage to
your instrument, please read and
observe the following precautions.
• If the product is used in a manner not in
accordance with the equipment design,
operating instructions or manufacturer's
recommendations, the operation of the
product may be impaired to the extent that
a safety hazard is created.
• Do not attempt to perform electrical work
if you are not fully qualified. This manual
is not a substitute for electrical training.
Symbol Conventions
The exclamation point within an
equilateral triangle is intended to
alert the user to the presence of
important operating and maintenance
(servicing) instructions in the litera-
ture accompanying this product.
The lightning flash with arrowhead
symbol within an equilateral triangle
is intended to alert the user to the
presence of uninsulated dangerous
voltage within the product's enclo-
sure that may be of sufficient magni-
tude to constitute risk of electric
shock to persons.
The static symbol within an equilat-
eral triangle is intended to alert the
user to the presence of internal com-
ponents that could be damaged by
static electricity.
This static symbol is intended to
alert the user to the presence of a
specific component that could be
damaged by static electricity.
This symbol indicates the presence
of alternating current (AC).
This symbol indicates the presence
of a fuse.
This symbol indicates the presence
of protective earth ground.
This symbol indicates the power is
ON.
This symbol indicates the power is
O .
NO E Additional symbol definitions are
available in the Symbols section.
General Cautions
• This product should be operated only with
the type of power source indicated on the
product’s electrical ratings label. Refer to
the installation instructions included with
the product.
• If the power cord provided is replaced for
any reason or if an alternate cord is used,
the cord must be approved for use in the
local country. The power cord must be
approved for the product’s listed operating
voltage and be rated at least 20% greater
than the ampere ratings marked on the
product’s electrical ratings label. The cord
end that connects to the product must have
an IEC 60320 connector.
• Plug the product into an approved ground-
ed electrical outlet.
• Do not disable the power cord’s grounding
plug.
• If an extension cord or power strip is used,
make sure that the cord or strip is rated for
the product, and that the total ampere rat-
ings of all products plugged into the exten-
sion cord or strip do not exceed 80% of the
cord’s or strip’s rating limit.
Safety / Regulatory Information
9

he Advanced®Model BR2 Service Manual 10
• Route power cords so that they will not be
walked on, tripped on, or pinched by items
placed upon or against them. Pay particu-
lar attention to the plug, electrical outlet,
and the point where the cord exits the
product.
• Do not pull on cords and cables. When
unplugging cords or cables, grasp the cor-
responding connector.
• Do not install or use this product in any
area subject to extreme short-term temper-
ature variations, or locations that exceed
the specified operating environment tem-
peratures.
• Never use this product in a wet area.
• To avoid injury or fire hazard, do not oper-
ate this product in an explosive atmos-
phere.
• Do not install or use the product on an
unstable, non-level work surface.
• Do not operate this product with the covers
removed or unsecured.
Service & Maintenance Cautions
•Unplug the power cord prior to opening or
removing covers, or else you may be
exposed to electric shock, excessive tem-
peratures, or mechanical hazards.
•Performing service or maintenance not
detailed in the User’s Guide, with or with-
out this Service Manual, should only be
done by a qualified service technician.
•Never restrict airflow into or out of the
product. Occasionally, check the air vents
for blockage.
•Wipe the exterior of the product with a
soft, damp cloth as needed. Using cleaning
products other than those specified, may
discolor or damage the finish.
•If the product requires service for any of
the following reasons, unplug the product
from the electrical outlet and refer service
to a qualified service technician.
-The power cord, extension cord, power
strip or power input module is dam-
aged.
-Liquid has been spilled into the interior
of the product.
-A foreign object has fallen into the
product.
-The product has been dropped or dam-
aged by a falling object.
-There are noticeable signs of overheat-
ing or a burning odor.
-The product does not operate normally
when you follow the operating proce-
dures.
-The main supply fuse(s) or any internal
fuse(s) continually fail.
•A discharge of static electricity from con-
tact with the human body or other conduc-
tor may damage system boards or static
sensitive devices. Never perform internal
maintenance without following recom-
mended static protection procedures.
• The product is equipped with operator
accessible fuses. If a fuse blows, it may be
due to a power surge or failure of a com-
ponent. Replace the fuse only once. If the
fuse blows a second time, it is probably
caused by failure of a component part. If
this occurs, refer service to qualified serv-
ice personnel. Always replace the fuse
with one of the same rating, voltage, and
type. Never replace the fuse with one of a
higher current rating.
• When servicing the product, use only fac-
tory-specified parts.
• WARNING: When returning this product
for service, or shipping this product to a
second location, remove all hazardous
specimens and decontaminate the product
before packaging for shipment. If the
product cannot be decontaminated, consult
with your shipping agent on appropriate
packaging and marking.