AEMC 5233 Manuale utente

5233
ENGLISH User Manual
DIGITAL MULTIMETER
with Non-Contact Detection

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2125.65
Model #: 5233
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

Digital Multimeter Model 5233
1
Table of Contents
1. INTRODUCTION............................................................................... 3
1.1 International Electrical Symbols................................................4
1.2 DenitionofMeasurementCategories .....................................4
1.3 Receiving Your Shipment..........................................................5
1.4 OrderingInformation.................................................................5
1.4.1 Accessories and Replacement Parts............................5
2. PRODUCT FEATURES ...................................................................... 6
2.1 Description................................................................................6
2.2 Control Features .......................................................................7
2.3 Display Features.......................................................................8
2.4 Button Functions.......................................................................9
2.5 Rotary Functions.......................................................................9
3. OPERATION .................................................................................. 10
3.1 Turning the Multimeter ON......................................................10
3.2 Turning the Multimeter OFF....................................................10
3.3 Activating/Deactivating Auto-OFF...........................................10
3.4 Auto and Manual Range Selection .........................................10
3.5 Non-Contact Voltage (NCV)....................................................11
3.6 Voltage Measurement.............................................................11
3.7 Resistance Measurement .......................................................12
3.8 Continuity Test ........................................................................13
3.9 Diode Test...............................................................................14
3.10 Capacitance Test ....................................................................15
3.11 Temperature Test ....................................................................16
3.12 Current Measurement.............................................................17
4. MAINTENANCE ............................................................................. 18
4.1 Warning ..................................................................................18
4.2 Battery Replacement ..............................................................18
4.3 Fuse Replacement..................................................................19
4.4 Cleaning..................................................................................19

2
Digital Multimeter Model 5233
5. SPECIFICATIONS........................................................................... 20
Repair and Calibration...........................................................................23
Technical and Sales Assistance ............................................................23
Limited Warranty ...................................................................................24
Warranty Repairs...................................................................................24

Digital Multimeter Model 5233
3
CHAPTER 1
INTRODUCTION
Warning
This device complies with safety standard IEC-61010-1 (Ed
2–2001)forvoltagesupto1000VCATIIIor600VCATIV,atan
altitudebelow 2000m,indoors,with apollutionlevel ofnotmore
than 2.
Failure to observe the safety instructions may cause an electric
shock,re,explosion,ordestructionoftheinstrumentandofthe
installations.
• Donotusetheinstrumentinanexplosiveatmosphereorin
thepresenceofammablegasesorfumes.
• Donotusetheinstrumentonnetworksofwhichthevoltage
orcategoryexceedsthosementioned.
• Donotexceedtheratedmaximumvoltagesandcurrents
between terminals or with respect to earth/ground.
• Donotusetheinstrumentifitappearstobedamaged,
incomplete,ornotproperlyclosed.
• Beforeeachuse,checktheconditionoftheinsulationonthe
leads,housing,andaccessories.Anyelementofwhichthe
insulationisdeteriorated(evenpartially)mustbesetasidefor
repair or scrapped.
• Useleadsandaccessoriesratedforvoltagesandcategories
atleastequaltothoseoftheinstrument.
• Observetheenvironmentalconditionsofuse.
• Donotmodifytheinstrumentanddonotreplacecomponents
with “equivalents”. Repairs and adjustments must be done by
approvedqualiedpersonnel.
• Replace the battery as soon as the symbol appears
onthedisplayunit.Disconnectallleadsbeforeopeningthe
battery compartment cover.
• Use personal protective equipment when conditions require.
• Keepyourhandsawayfromunusedterminalsoftheinstrument.
• Whenhandlingprobesorcontacttips,keepyourngers
behind the guards.

4
Digital Multimeter Model 5233
1.1 International Electrical Symbols
Signifies that the instrument is protected by double or reinforced insulation.
This symbol on the instrument indicates a WARNING that the operator must
refer to the user manual for instructions before operating the instrument. In
this manual, the symbol preceding instructions indicates that if the instructions
are not followed, bodily injury, installation/sample and/or product damage may
result.
Compliance with the Low Voltage & Electromagnetic Compatibility European
directives (73/23/CEE & 89/336/CEE)
AC – Alternating current
AC or DC – Alternating or direct current
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
Important instructions to read and understand completely.
Important information to acknowledge.
Ground/Earth symbol
In conformity with WEEE 2002/96/EC
1.2 Definition of Measurement Categories
CAT I: For measurements on circuits not directly connected to the AC
supplywalloutletsuchasprotectedsecondaries,signallevel,and
limited energy circuits.
CAT II: For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements on
household appliances or portable tools.
CAT III:
Formeasurementsperformedinthebuildinginstallationatthedistri-
butionlevelsuchasonhardwiredequipmentinxedinstallationand
circuit breakers.
CAT IV: For measurements performed at the primary electrical supply
(<1000V)suchasonprimaryovercurrentprotectiondevices,ripple
controlunits,ormeters.

Digital Multimeter Model 5233
5
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistent
withthepackinglist.Notifyyourdistributorofanymissingitems.Iftheequip-
mentappearstobedamaged,leaclaimimmediatelywiththecarrierand
notifyyourdistributoratonce,givingadetaileddescriptionofanydamage.
Save the damaged packing container to substantiate your claim
.
1.4 Ordering Information
Multimeter Model 5233......................................................Cat. #2125.65
Includes set of two 5 ft color-coded leads (red/black) with needle tip {1000V CAT IV
15A), thermocouple Type K with adapter, set of three fuses 10A, 600V, 50kA, (fast
blow) 5x32mm, soft carrying case and a user manual.
1.4.1 Replacement Parts
Fuse-Setof10,10A,600V,50kA,(FastBlow),5x32mm
.... Cat. #2118.62
SoftCarryingCase ..............................................................Cat. #2121.54
Adapter - Banana (Male) to Mini (Female)
with K-Type Thermocouple.................................................. Cat. #2125.83
Thermocouple-Flexible(1m)K-Type58°to480°F ............ Cat. #2126.47
Setof2,Color-coded5ftleadswithsafetyneedletips(2mm)
{1000V CAT IV 15A) ..........................................................Cat. #2140.68
MultixMountingSystem.....................................................Cat. #5000.44
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com for availability

6
Digital Multimeter Model 5233
CHAPTER 2
PRODUCT FEATURES
2.1 Description
TheModel5233isaTRMSdigitalmultimeter,speciallydesignedtocom-
binethevarious functions and measurements of the following electrical
quantities:
• Non-contactdetectionofpresenceofnetworkvoltage
(NCVfunction)
• AC voltmeter with low input impedance (voltage measurements
forelectricityandelectricalengineering)
• AC/DC voltmeter with high input impedance
(voltagemeasurementsforelectronics)
• Frequency and duty cycle measurements
• Ohmmeter
• Continuity test with buzzer
• Diode test
• Ammeter
• Capacitance meter
• Thermometerin°Cor°Fbymeasurementandlinearizationof
thevoltageacrosstheterminalsofaK-typethermocouple

Digital Multimeter Model 5233
7
2.2 Control Features
0600
100 500
200 400300
AUTO
AC DC MIN MAX
HOLD
P
°C°F
mVA
MkΩ
µnF
Hz%
REL
∆
2
1
3
4
5
6
Figure 2-1
1. NCV detection sensor (see § 3.5)
2. Analog and digital display (see § 2.3)
3. Function buttons (see § 2.4)
4. Rotary switch (see § 2.5)
5. Current measurement 10A terminal (see § 3.12)
6. Positive (Red) input and COM (Black) input

8
Digital Multimeter Model 5233
2.3 Display Features
Icon Function
AC Alternating Current
DC Direct Current
AUTO Automatic range selection (see § 3.4)
HOLD Freezes the display of the measurement
MAX Maximum RMS value
MIN Minimum RMS value
REL Relative value
The Over Load symbol is displayed when the signal measured
exceeds the range of the device
VVoltage
Hz Hertz
%Duty Cycle
FFarad
°C Degrees Celsius
°F Degrees Fahrenheit
AAmpere
WOhm
nPrefix “nano”
µPrefix “micro”
mPrefix “milli”
kPrefix “kilo”
MPrefix “Mega”
Continuity Beeper Enabled
Diode Test
Low Battery
P
Auto Power OFF function activated
Indice
Altri manuali AEMC Multimetro

AEMC
AEMC SL206 Manuale utente

AEMC
AEMC 701 Manuale utente

AEMC
AEMC PEL 102 Manuale utente

AEMC
AEMC MTX 3282B Manuale di istruzioni

AEMC
AEMC F01 Manuale utente

AEMC
AEMC 5231 Manuale utente

AEMC
AEMC C.A 5220 Manuale utente

AEMC
AEMC 505 Manuale utente

AEMC
AEMC 605 Manuale utente

AEMC
AEMC Megohmmeters 6522 Manuale utente
Manuali Multimetro popolari di altre marche

Gossen MetraWatt
Gossen MetraWatt METRAmax 6 Manuale utente

PeakTech
PeakTech 4000 Manuale per l'uso e la cura

YOKOGAWA
YOKOGAWA 90050B Manuale utente

Gossen MetraWatt
Gossen MetraWatt METRALINE DMM16 Manuale utente

Fluke
Fluke 8846A Manuale operativo e di manutenzione

Tempo Communications
Tempo Communications MM200 Manuale utente