506/506B EN
Read First
Safety Information
WARNING
Understand and follow operating instructions carefully.
Use the meter only as specied in this manual; otherwise,the
protection provided by the meter may be impaired.
This identies hazardous conditions and actions that could
cause BODILY HARM or DEATH. To avoid possible danger,
follow below guidelines.
• Use the meter only as specied in this manual or the
protection by the meter might be impaired.
• Never operate the meter with the cover removed or the
case open.
• To avoid false readings that can lead to electric shock and
injury, replace battery as soon as low battery indicator.
• Use caution with voltages above 30VAC rms, 42VAC peak,
or ±30VDC. These voltages pose a shock hazard.
• When using test leads or probes, keep your ngers behind
the nger guards.
• Remove test lead from meter before opening the battery
door or meter case.
• Always use proper terminals, switch position, and range for
measurements.
• Do not apply more than the rated voltage, as marked on
meter, between terminals or between any terminal and
earth ground.
• Do not use the High Frequency Rejection (Low Pass
Filter) option to verify the presence of hazardous voltages.
Voltages greater than what is indicated may be present.
First, make a voltage measurement without the lter to
detect the possible presence of hazardous voltage. Then
select the lter function.
• To avoid possible electric shock or personal injury, never
attempt an in-circuit current measurement where the open
circuit potential to earth is greater than 1000V.
• Replace the fuse as soon as the indicator (FUSE) appears.
• Only replace the blown fuse with the proper rating as
specied in this manual.
• Do not use the meter around explosive gas, vapor or dust.
• To reduce the risk of re or electric shock do not expose
this product to rain or moisture.
1

506/506B EN
• Do not attempt a current measurement when the open
voltage is above the fuse protection rating. Suspected
open voltage can be checked with voltage function.
• Never attempt a voltage measurement with the test lead
inserted into the A input terminal.
• Disconnect the test leads from the test points before
changing the position of the function rotary switch.
• Disconnect circuit power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
• Always use proper terminals, switch position, and range for
measurements.
• Do not use the LoZ mode to measure voltages in circuits
that could be damaged by this mode’s low impedance.
• Replace the fuse as soon as the indicator (FUSE) appears.
• Never connect a source of voltage with the function rotary
switch in resistance, diode, continuity, and capacitance
position.
• Never set the meter in current function to measure the
voltage of a power supply circuit in equipment that could
result in damage the meter and the equipment under test.
This identies conditions and actions that could DAMAGE
the meter or equipment under test. To avoid possible
damage, follow below guidelines.
CAUTION
2

506/506B EN
Hazardous voltage
Risk of danger. Important information. See manual
AC (Alternating Current)
DC (Direct Current)
Bluetooth
Double insulated
Low battery
Fuse
Earth ground
Conforms to European Union directives
Do not discard this product or throw away
IEC Overvoltage Category
CAT III equipment is designed to protect against transients
in equipment in xed equipment installations, such as
distribution panels, feeders and short branch circuits, and
lighting systems in large buildings.
CAT IV equipment is designed to protect against transients
from the primary supply level, such as an electricity meter or
an overhead or underground utility service
Electrical Symbols
CAT III
CAT IV
Fuse broke. Replace fuse as soon.
Meter error. Have meter serviced.AC (Alternating Current)
Test probe alert. Displayed when the test probes are in the A
or mA terminal and the selected rotary switch position does
not correspond to the terminal being used.
Error Message
ProbE
FUSE
Er
3

506/506B EN
The Meter Description
Front panel illustration :
1. Auto backlight sense point
2. 40,000 count dual display
3. Push buttons
4. Rotary switch
5. Input terminal for voltage, frequency, resistance,
continuity,diode, capacitance and temperature measurements.
6. Return terminal for all measurements.
7. Input terminal for 0 to 400mA current measurements.
8. Input terminal for 0 to 10A current measurements.
TRUE RMS
600V
CAT.III
1000V
MAX
FUSED
400mA
MAX
10A
FUSED
CAT.IV
COM
mA V
A
Hz
AC+DC
A
AC+DC
V
AC+DC
mV
Hz
mV
LoZ
Auto V
V
Hz
OFF
P HOLDCANCEL
°C/°F
A HOLDENTERHFRRANGE
Hz
1
2
3
4
56
7
8
4

506/506B EN
Power On Options
Making Basic Measurements
When turn the power on, press the function button to
execute the below options.
• When connecting the test leads to the DUT (Device Under
Test), connect the common test lead before connecting the
live lead. When removing the test leads, remove the test
live lead before removing the common test lead.
The following sections describe how to take measurements
with the meter.
Push Buttons
Function
(Yellow)
RANGE
HFR
A-HOLD
ENTER
CANCEL
°C / °F
P-HOLD
Bluetooth
Select measurement function.
Select measurement range. Press > 1 sec to enter auto
range mode.
Enable/Disable the High Frequency Reject mode in the AC
measurements.
Enable the Peak-Hold mode in the AC or DC measurements.
In this mode, press button to select Peak-Hold MAX or MIN.
Press > 1 sec to disable the Peak-Hold mode.
Enable/Disable the Bluetooth mode.
Enable/Disable the Auto-Hold mode.
Enter menu function in pointer positon.
Cancel current menu function.
Select degrees Celsius or degrees Fahrenheit.
Display the rmware version.
Display all LCD segments.
Power On Options
ENTER
A HOLD
WARNING
5

506/506B EN
Measuring AC and DC Voltage
Make dB measurement
This meter have true rms readings, which are accurate
for distorted sine waves and other waveforms (with no dc
oset) such as square waves, triangle waves, and staircase
waves.
The ranges of measuring voltage are 40mV, 400mV, 4V,
40V, 400V and 1000V. To select the mV range, turn the
rotary switch to mV position.
The meter is capable of displaying voltage as a dB value, either
relative to 1 milliwatt (dBm), a reference voltage of 1 volt (dB).
A dBm measurement must use a reference impedance (600Ω)
to calculate a dB value based on 1 milliwatt. A dB measurement
uses a 1 volt reference voltage to compare the present
measurement against. Dene as below :
For best accuracy when measuring the DCmV, touch the
probe tips together and read the DC oset. If necessary, you
can use the relative (Δ) mode to automatically subtract this
value.
AC Voltage DC Voltage
dBm = 20log ( )
Vrms
dBm = 20log ( )
Vrms
1V
6

506/506B EN
Measuring Voltage in LoZ Mode
To use dB or dBm function, turn the rotary switch to ACV or
ACmV position. Then move the blink cursor of menu to
dB or dBm position, and press the ENTER button to enter
function.
Press the CANCEL button to exit function.
Caution
• Do not use the LoZ mode to measure voltages in circuits
that could be damaged by this mode’s low impedance.
To eliminate ghost voltages, the meter’s LoZ mode presents
a low impedance across the leads to obtain a more accurate
measurement. The ranges of measuring LoZ voltage
are 400V and 1000V. In this mode, meter will automatic
measure input signal which is AC or DC and determine
range.
To use the LoZ mode, turn the rotary switch to LoZ position.
Measuring AC and DC Current
• Always use proper terminals, switch position, and range for
measurements.
• To avoid possible electric shock or personal injury, never
attempt an in-circuit current measurement where the open
circuit potential to earth is greater than 1000V.
• Only replace the blown fuse with the proper rating as
specied in this manual.
WARNING
7

506/506B EN
Caution
• Replace the fuse as soon as the indicator (FUSE) appears.
To measure current, you must break the circuit under test,
thenplace the meter in series with the circuit.
The ranges of measuring current are 40mA, 400mA, 4A and
10A. AC current is displayed as an rms value. Insert the
black lead into the COM terminal. For currents less than 400
mA, insert the red lead into the mA terminal. For currents
above 400 mA, insert the red lead into the A terminal.
You can press the function (yellow) button to select current
measurement function.
Measuring Frequency
The meter measures the frequency of a voltage or current
signal by counting the number of times the signal crosses a
threshold level each second.
This function only can be operated in AC voltage and current
measurements. The ranges of measuring frequency are
400Hz,4kHz, 40kHz and 100kHz.
If a reading shows as 0 Hz or is unstable, the input signal
may be below or near the trigger level. The detail of
frequency trigger level refer to the electrical specications.
To use the frequency function, press function (yellow) button
to select measurement function.
8

506/506B EN
Make Peak-Hold Measurement
The Peak-Hold mode records wave peak maximum and
minimum input values. Response time of Peak-Hold is 10us.
To use the Peak-Hold mode, press the P-HOLD button to
enable the Peak-Hold mode in the AC or DC measurements.
In this mode, press P-HOLD button to select peak MAX or
MIN value. Press > 1 sec to disable the Peak-Hold mode.
For square wave, use the peak-hold mode in the DC
measurements.
Make High Frequency Rejection Measurement
• Do not use the High Frequency Rejection (Low Pass Filter)
option to verify the presence of hazardous voltages.
Voltages greater than what is indicated may be present.
First, make a voltage measurement without the lter to
detect the possible presence of hazardous voltage. Then
select the lter function.
The High Frequency Rejection mode equip a low pass lter
in the AC measurements. The cut-o frequency (-3dB point)
of low pass lter is 800Hz.
To use the HFR mode, press the HFR button to equip a low
pass lter in the AC measurements.
WARNING
9
Questo manuale è adatto per i seguenti modelli
1
Indice
Lingue:
Altri manuali APPA Multimetro
Manuali Multimetro popolari di altre marche

Gossen MetraWatt
Gossen MetraWatt METRAmax 6 Manuale utente

PeakTech
PeakTech 4000 Manuale per l'uso e la cura

YOKOGAWA
YOKOGAWA 90050B Manuale utente

Gossen MetraWatt
Gossen MetraWatt METRALINE DMM16 Manuale utente

Fluke
Fluke 8846A Manuale operativo e di manutenzione

Tempo Communications
Tempo Communications MM200 Manuale utente