APPA 97 Manuale utente

97
User Manual / 使用說明書 / 使用说明书
ユーザーマニュアル
Руководство пользователя
Rugged Industrial Multimeter
堅固型工業用萬用表
坚固型工业用万用表
堅牢型産業用マルチメーター
Прочный промышленный мультиметр
RU
EN
TC
SC
JP
LIMITE D
WARRAN TY
YEAR S

97 EN
WARNING
THESE SERVICING INSTRUCTIONS ARE FOR
USE BY QUALIFIED PERSONNEL ONLY. TO AVOID
ELECTRIC SHOCK, DO NOT PERFORM ANY
SERVICING OTHER THAN THAT CONTAINED IN THE
OPERATING INSTRUCTIONS UNLESS YOU ARE
QUALIFIED TO DO SO.
Also the complete duty cycle for the 20A measurement
should be stated as on for 30 seconds max., and o for
3 minutes min.
Upon removing your new Digital Multimeter from its
packing, you should have the following items:
1. Digital Multimeter.
2. Test lead set (one black, one red).
3. Operators manual.
4. Protective holster.
Terms as Marked on Equipment.
ATTENTION - Refer to manual.
DOUBLE INSULATION - Protection Class I.
DANGER - Risk of electric shock.
Symbols in this Manual:
This symbol indicates where cautionary or other
information is found in the manual.
FUSE
Battery
INTRODUCTION
1-1 Unpacking and Inspection
1-2 Meter Safety
1

97 EN
Refer to Figure 1 and to the following numbered steps to
familiarize yourself with the meter's front panel controls
and connectors.
1. Digital Display - The digital display has a 3400
counts LCD readout with 65 segments analog bar
graph, auto polarity, decimal point, 「」
AC, LOΩ, RANGE, HOLD and unit annunciators.
2. Rotary Switch - Select the Function and Range
desired.
3. COM Input Terminal - Ground input connector.
4. V Ω Input Terminal - Positive input connector for
Volts, Ohms and Diode.
5. HA mA Input Termina - Positive input connector for
Amp measurements (up to 320mA).
6. A Input Terminal - Positive input connector for Amp
measurements (up to 20A).
7. Range Switch, (Manual Range) - "Range" switch is
pushed to select manual ranging and to change
ranges. When "Range" switch is pushed once
"RANGE" anunciator on the LCD is appeared.Push
"RANGE" switch to select appropriate range to be
used. Push "RANGE" switch and hold 2 seconds to
return to Autoranging.
8. AC / DC / / Selection Switch - Push the switch
to measure AC Current or DC Current in the current
mode or to measure continuity or diode in /
mode.
9. Hold Switch - This switch is used to hold measured
value for all functions, and then "HOLD" annunciator
is displayed. Conversions are made but the display
is not updated.
10. DELAY Switch - This switch is used to delay hold
the measured value. The "HOLD" annuciator is
displayed after push the「DELAY 」switch about
6 seconds. Push "HOLD" switch to cancel display
hold or push「DELAY 」switch after 6 seconds
the display hold is canceled. Conversions are made
1-2 Front Panel
2

97 EN
Display : The Liquid Crystal Display (LCD) with a
maximum reading of 3200 and 65 segments
bar graph.
Polarity Indication : Automatic, positive implied,
negative indicated.
Overrange Indication : "OL" or "-OL".
Low Battery Indication :「」is displayed when the
battery voltage drops below
operating voltage.
Sampling : 2 times/sec for digit. 12 times/sec for analog
bargraph.
Auto Power O : Approx 10 minutes.
Operating Ambient : 0°C to 500°C, O to 80% R.H.
Storage Temperature : -200°C to 60°C , 0 to 80% R.H.
when battery removed from meter.
Temperature Coecient : 0.15 x (Specied accuracy)
/ °C, <18°C or > 28°C .
SPECIFICATIONS
2-1 General Specications
but the display is not updated. The beep will sound
when push the HOLD Switch or after about 6
seconds the beep will sound when push the「DELAY
」Switch.
3

97 EN
Indoor use.
Pollution Degree : 2
Maximum Altitude : 2000 Meter.
Installation Category : IEC 1010, 1000V Cat. II,
600V Cat. IlI.
Accuracy is ±(% reading + number of digits) at 23°C +
5°C at less than 80% R.H.
Input Impedance : 10MΩ
(1) DC Volts
2-2 Environmental Conditions
2-2 Electrical Specications
CAT Application eld
The circuits not connected to mains.
The circuits directly connected to Low-voltage installation.
The building installation.
The source of the Low-voltage installation.
I
II
III
IV
Range
3V
300mV
1 mV
100 µV
30V 10mV
300V
1000V
100 mV
±(0.5% +
2dgt)
1000V rms
1V
Resolution Accuracy Over voltage
protection
Power Requirements : IEC LR03, AM4 or AAA size
1.5V x 2
Battery Life : Alkaline 700 hours.
Dimensions (W x H x D) :
84mm x 175mm x 31mm without holster.
95mm x 192mm x 50mm with holster.
Accessories : Protective Holster, battery, operators
manual.
4

97 EN
* Frequency Response : 40Hz ~ 300Hz for 3V Range.
AC Conversion Type : Average Sensing rms indication.
Input Impedance : 10 MΩ // <100PF。
(2) AC Volts
Range
30V
3V
10 mV
1mV
300V 100mV
750V 1V
1000V rms
±(1.3%rdg+
5dgt)
40Hz to 500Hz
±(1.3%rdg+
5dgt)*
Resolution Accuracy Over voltage
protection
(3) DC Current
Range
3mA
300µA
1µA
0.1µA ±(1%rdg +
2dgt)
±(1.2%rdg +
2dgt)
±(1%rdg +
2dgt)
±(1.2%rdg+
2dgt)
±(2%rdg +
3dgt)
200mV max.
2V max.
200mV max.
2V max.
2V max.
30mA 10µA
300mA
*20A
0.1mA
10mA
Resolution Accuracy Over voltage
protection
* 20A Range : 30 seconds maximum above 10A input.
Overload Protection : 1A (500V) fast blow fuse for
µAmA input.
16A (500V) fast for A input.
5

97 EN
(4) AC Current
* Frequency Response : 40Hz~ 500Hz.
* 20A Range : 30 seconds maximum above 10A.
Overload Protection : 1A (500V) fast blow fuse t for
µAmA input.
16A (500V) fast for A input.
* AC Conversion Type : Average Sensing rms indication.
Open circuit Voltage : 1.3V approx.
(5) Resistance
3KΩ
300Ω
1Ω
0.1Ω
30KΩ 10Ω
300KΩ
3MΩ
30MΩ
100Ω
±(0.8%rdg +
2dgt)
±(1.0%rdg +
4dgt)
±(1.2%rdg+
3dgt)
±(2.5%rdg +
5dgt)
600V rms
1KΩ
10KΩ
3mA
300µA
1µA
0.1µA
±(1.5%rdg+
3dgt)
±(2%rdg+
3dgt)
±(2.5%rdg +
5dgt)
200mV max.
200mV max.
2V max.
2V max.
2V max.
30mA 10µA
300mA
*20A
0.1mA
10mA
Range Resolution Accuracy Voltage Burden
Range Resolution Accuracy Over voltage
protection
6

97 EN
* for 0.4 to 0.8V
Overload protection : 600Vrms max.
Continuity : Built in-buzzer sound when resistance is
less than 30Ω
(6) Diode Check and Continuity
Range
1mV ±(1.5%rdg +
5dgt)* 1.5mA 3.3V
Resolution Accuracy Max.Open
Circuit Voltage
Max.Test
Current
The meter will automatically shut itself o after
approximately 10 minutes after power on.
The meter can be turned back on by pushing "Power
Reset" key switch.
The beep will sound if the test lead is connected to
the µAmA (10A) input terminal but the rotary function
selector is not in the pAmA (10A) positions.
(7) Auto Power O
(8) Beep Guard
This instrument has been designed and tested in
accordance with IEC Publication 1010, Safety Require-
ments for Electronic Measuring Apparatus and has been
supplied in a safe condition. This instruction
manual contains some information and warnings which
have to be followed by the user to ensure safe
operation and to retain the instrumnent in safe condition.
1, Before measurement, warm up for at least 60
seconds.
2. When the rotary function selector is changed during
measurement , be sure do so only after removing
the test leads from the equipment.
OPERATION
3-1 Preparation and Caution before Measurement
7

97 EN
1. Connect the red test lead to "µAmA" terminal and
the other (black) test lead to "COM" terminal, or use
the "A" and "COM" terminal in the 20A range.
2. Set function selector rotary switch to "µA" or "mA"
or "A".
3. Measurement of AC current can be performed by
pushing the "AC/DC" switch.
4. Connect the test leads to the circuit to be measured.
3-3 Current Measurement
1. Connect the red test lead to the " V Ω " input
terminal and the other (black) test lead to the "COM"
terminal.
2. Set the rotary function to the ACV or DCV position.
3. Connect the test leads to the device to be measured.
3-2 Voltage Measurements
WARNING
TO AVOID ELECTRICAL SHOCK , HAZARD OR
DAMAGE TO METER , DO NOT A ATTEMPT TO
MEASURE VOLTAGE THAT MIGHT EXCEED 1000
VOLT DC OR 750 VOLT AC . DO NOT APPLY MORE
THAN 1000V DC OR AC RMS BETWEEN THE
COMMON INPUT TERMINAL AND EARTH GROUND.
NOTICE : UNSTABLE DISPLAY MAY OCCUR
ESPECIALLY AT 300mV RANGE , EVEN THOUGH YOU
DON'TPUT TESTED LEADS INTO INPUT TERMINALS.
IN THIS CASE , IF AN ERRONEOUS READING IS
SUSPECTED, SHORT THE " V Ω " TERMINAL AND
THE "COM" TERMINAL, AND MAKE SURE
THE ZERO DISPLAY.
3. If the equipment is used near noise generating
equipment, be aware that display may become
unstable or indicate large errors.
4. Maximum rated voltage to earth for voltage and
current measurements terminals is 1000V AC / DC
CAT. II .
8

97 EN
1. Connect the red test lead to the " V Ω " terminal
and the other (black) test lead to the "COM" terminal.
2. Set the rotary function selector to " " position.
3. Connect the test leads to the circuit to be measureu .
The biszzer will sound if the resistance of the circuit
measured is lower than 30g.
1. Set the rotary switch at " " position.
2. Connect black test lead to "COM" terminal and red
lead to " V Ω " input terminal.
3. Connect test leads to the diode normally the forward
voltage drop of good silicon diode is shown between
400V to 900V. If the diode under test is defective .
"000" (short circuit) or "OL" (non-conductance) is
displayed. Reverse check of diode if the diode under
test is defective ''000" or other value are to be
displayed.
3-5 Continuity Check by Buzzer
3-6 Diode Check
TO AVOID ELECTRICAL SHOCK REMOVE TEST LEAD
BEFORE OPENING THE COVER.
MANTENANCE
WARNING
1. Connect the red test lead to the " V Ω " terminal and
the other (black) test lead to the "COM" terminal.
2. Set the rotary function selector to "Ω" position to
measure the resistance.
3. For correct reading , ensure that the device being
tested contains no voltage.
4. Connect the test leads across the resistor to be
measured. In order to ensure the best accuracy in
measurement of low resistance, short the test leads
before measurement and memory the test probe
resistance in mind. This necessary to subtract for the
resistance of the test leads themselves.
3-4 Resistance Measurement
9
Altri manuali per 97
1
Indice
Lingue:
Altri manuali APPA Multimetro
Manuali Multimetro popolari di altre marche

Gossen MetraWatt
Gossen MetraWatt METRAmax 6 Manuale utente

PeakTech
PeakTech 4000 Manuale per l'uso e la cura

YOKOGAWA
YOKOGAWA 90050B Manuale utente

Gossen MetraWatt
Gossen MetraWatt METRALINE DMM16 Manuale utente

Fluke
Fluke 8846A Manuale operativo e di manutenzione

Tempo Communications
Tempo Communications MM200 Manuale utente