Arometrix Extraction Finder Manuale utente

Arometrix EXTRACTION FINDER
USER MANUAL
YOU MUST READ THIS MANUAL BEFORE USE | WARNINGS ON PAGE 3

2
TABLE OF CONTENTS
Section 1: Warnings ⚠3
Section 2: Overview 7
Section 3: Fundamentals 7
Section 4: Unpacking & Inspecting 8
Section 5: Installation & Operation 8
Section 6: Understanding the Interface 12
Section 7: Application Note - Ethanol Extraction - Solvent Saturation 16
Section 8: Maintenance 17
Section 9: Issues 17
Section 10: Software Update Instructions 18
Section 11: Terms and Conditions 19
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

3
Section 1: Warnings ⚠
The complete contents of this manual must be reviewed to ensure that the product is used safely. When
integrating the EXTRACTION FINDER, the total extraction system design must be considered to ensure safe,
trouble-free performance. Function, material compatibility, adequate ratings, proper installation,
operation, and maintenance are the responsibilities of the system designer and user. Below are Arometrix’s
warnings about safe operating conditions for the EXTRACTION FINDER:
1. Ensure the system and installation conform to all applicable laws and ordinances, including Fire
Marshall code certifications.
2. Ensure all supporting electronics are installed in a compliant matter.
3. Ensure all connections are properly installed and specifications are appropriate.
4. Ensure extraction equipment is equipped with appropriate warning labels for solvent.
5. If you have any indications of malfunction, return to the manufacturer immediately by emailing
bra[email protected]om.
6. Tampering with this equipment is prohibited, and will result in a void of warranty.
7. Never look directly into the light source.
8. Proper care should be taken when setting up and operating this equipment, as MISUSE MAY
CAUSE SEVERE INJURY OR DEATH.
9. Never install valves or shut off devices on either side of the EXTRACTION FINDER as this MAY
CAUSE SEVERE INJURY OR DEATH.
Figure 1 : Sample Propane Warning Label (Right to Know Label) Figure 2 : Sample Propane NFPA 704 Hazard Rating Placard
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

4
Figure 3 : Sample Pentane Warning Label (Right to Know Label) Figure 4 : Sample Pentane NFPA 704 Hazard Rating Placard
Figure 5 : Sample Methanol Warning Label (Right to Know Label) Figure 6 : Sample Methanol NFPA 704 Hazard Rating Placard
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

5
Figure 7 : Sample Isopropyl Alcohol Warning Label Figure 8 : Sample Isopropyl Alcohol NFPA 704 Hazard Rating Placard
Figure 9 : Sample Hexane Warning Label (Right to Know Label) Figure 10 : Sample Hexane NFPA 704 Hazard Rating Placard
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

6
Figure 11 : Sample Heptane Warning Label (Right to Know Label) Figure 12 : Sample Heptane NFPA 704 Hazard Rating Placard
Figure 13 : Sample Ethanol Warning Label (Right to Know Label) Figure 14 : Sample Ethanol NFPA 704 Hazard Rating Placard
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

7
Figure 15 : Sample Butane Warning Label (Right to Know Label) Figure 16 : Sample Butane NFPA 704 Hazard Rating Placard
Section 2: Overview
What the EXTRACTION FINDER is:
The EXTRACTION FINDER is Arometrix’s real-time molecular monitoring system for botanical extraction.
The technology uses induced fluorescence to detect the presence and track the levels of molecules, such
as Cannabinoids, Chlorophyll, and Lipids in-line. This helps operators optimize extraction process efficiency.
What the EXTRACTION FINDER is not:
The EXTRACTION FINDER is not a quantitative measure. It provides qualitative process information that
directly tracks the relative concentration of Cannabinoids. It is Arometrix’s goal to amass enough spectral
data from the EXTRACTION FINDER to eventually be able to determine quantitative purity in the future.
The EXTRACTION FINDER cannot replace good laboratory practice and experience. The EXTRACTION
FINDER’s data, in combination with good lab practice, will help technicians further perfect their craft.
Section 3: Fundamentals
The Arometrix EXTRACTION FINDER is primarily composed of the Extraction Finder Monitor and the
Extraction Finder Sensor.
Arometrix Digital Monitor:The monitor is the “brains” of the system. This 10” LCD TFT monitor creates a
visualization for the user. The monitor has a stand.
Arometrix Sensor:The ultra-sensitive sensor, which can be installed in-line, is the “eyes” of the system. It
connects and transmits data to the monitor.
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

8
Design Notes: The sensor electronics are bathed in an inert nitrogen environment to further insulate them
from any risk of exposure, even though the sensor is intrinsically safe. Solvent will not touch the electronics.
The 30' sensor cable and 10' power cord allows the monitor to be easily placed away from the C1D1 area.
Section 4: Unpacking & Inspecting
After the instrument is received, it should be carefully unpacked and inspected for damage during
shipment and to confirm that all components are present.
Each EXTRACTION FINDER package comes with:
●Extraction Finder Sensor
●Arometrix Monitor
●Sensor Cable, 30’
●Power Supply, 10’
Each EXTRACTION FINDER 2022 and newer package comes with:
●Extraction Finder Sensor
●Arometrix Monitor
●Sensor Cable, 30’
●Power Supply, 10’
●Adapter (optional)
Section 5: Installation & Operation
Warning: Installing a valve or any device that could cause isolation in the line violates NFPA 58. Piping
that can contain liquid LP-Gas and that can be isolated by valving requires a hydrostatic relief valve, as
specified under Section 6.15.
Installation of the Extraction Finder Sensor involves the following steps (also depicted in our Quick Start
Guide):
1. First, mount the Extraction Finder Sensor to the frame of the equipment. An example is shown.
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

9
2. Next, connect the hoses.
3. Then, ensure all connections are properly installed.
4. Finally, ensure the system and installation conform to all applicable laws and ordinances.
Installation of the Extraction Finder Monitor involves the following steps:
1. Install the Extraction Finder monitor outside of the C1D1-designated area. Use the supplied 10’
power cord to power your monitor.
2. Connect the sensor cable to the monitor. The monitor will take a few seconds to turn. You will see
splash screens, followed by the interface screen. It will boot up in 5 minutes or less.
3. In the Settings section on the monitor’s UI, ensure the following:
a. Set Auto Integration by checking the AID (Auto) checkbox
Arometrix, Inc. sales@arometrix.com arometrix.com/resources

10
b. Set Scans to Average: 5
c. Ensure that the Device Status and Server Status indicators on the bottom left-hand side of
the UI are green – If not, make sure the cords are properly connected.
d. Ensure that the “Light On/Light Off” toggle button is turned on. The light above it should
be green.
e. To set ranges, go to Settings, select “Wavelength Settings” to track specific molecules. For
extraction, we recommend setting:
i. Region 1: Delta-9 (or CBD Indicator)
ii. Region 2: Lipids
iii. Region 3: Chlorophyll
iv. Pro-tip: Once you return back to the homescreen, check each region’s checkbox for
easier tracking during the process.
4. Record on (optional) at the start of the run: Toggle the Record button to On. Ensure the Record
indicator is green.
5. Once the Extraction Finder system is safely in operation, prepare extraction as you typically would.
Arometrix, Inc. sales@arometrix.com arometrix.com/resources
Altri manuali per Extraction Finder
2
Indice