ASONIK GL2M Manuale utente

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
1
The meter GL2M type
intended to moisture & relative
permittivity measurements
of the soft and thin layer
The GL2M is very accurate and sensitive metering unit
intended to determine moisture content in the tea leaf.
Every single measurement pertain to measure single
layer (single leaf) – see addendum “ ” of the manual.
dditionally are once taked also measurements of leaf
body thickness and temperature, also air ambient relative
humidity and temperature as well.
Before every measurement is essentionally to run the
calibration procedure.
To protect against potential errors is not possible to
pass over it.
Mentioned procedure removes all ambient temperature and
moisture influences on mechanical parameters of sensitive
thickness sensor.
In case of big mechanical stress, like rapid drop the
unit on the floor or on the desk would be required to
take more detailed calibration doings described in the
attached "thickness sensor calibration" point.
Driving down the external arm equiped with OLED display
bring the meter to work and simultaneously bring also to
taking hold of the sample inserted in to the measuring
aperture. The internal arm touch the leaf body with
controlled force about 1N regardless of the pushing
force applyied on the external arm, defined further as
“ possible lowest force”.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
2
1.Starting the calibration procedures:
Before running the unit, the measuring aperture must
be free of dust and other substances.
To start - PUSH and continue pushing the arm (with
possible lowest force), you hear the 2kHz sound and see
at the display the message as below:
.
If the aperture is clear and empty the GL2M proceed the
thic ness calibration state. If not, the procedure will
restart. For positive check, the procedure takes about
two second duration, at the end you hear short "chirp"
and stop of the 2kHz sound, next see at the display:
B.
The ifo let you to stop push the arm and lift the arm
off itself. Whenever the unit detect that the arm is
lifted up you hear the 2kHz sound.
The GL2M inform that now is in the offset callibration
state. The procedure takes up to one second. duration,
at the end you hear next short "chirp" , stop of the 2kHz
sound and see the display:
C.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
3
Screen "C" inform that the GL2M is ready to measure
the leaf moisture content and also leaf body thickness
leaf temperature, the the ambient air relative humidity
and ambient air temperature as well.
Caution: If at any step of described calibrations
you will stop manipulate the unit - the GL2M will
automatically switch off after 5s. or 13s. depending
on present calibration state.
2.Ta ing the measurement:
The GL2M is designed to measure moisture content in the
pure tea leafes body and the moisture content of crushed
tea leafes as well. How to prepare the crushed leafe
sample on PCV plastic tray and the way of measure is
described in addendum “B” at the end of manual.
Let insert the leaf body in to the sensor aperture.
Please pay attention to place the leaf body on the
mirrored central point of the plastic red stand-off.
In front of this point, at the upper arm is located the
infrared temperature sensor which must "see" the leaf
body to determine the leaf temperature. It is needful
to provide accurate moisture content enumerations.
Next push and continue pushing the arm (with
possible lowest force). The measurement takes about
two second duration, at the end you hear short chirp
and see at the display the data like example below:
D.
If the meter identify that the measured substance is
the leafe body – the screen wil show in the left top
corner the letter “L”, as on figure D. For presence of
crushed tea leafes inserted on plastic tray – the screnn
will show the letter “P”, as on figure C.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
4
Every measurement is automatically stored in one
of 30 memory banks. t the above screen can see that
presently, the working memory bank has the number "20".
The GL2M is programmed to take up to 16 different
measurements with current averaging of all results
in to the destinated memory bank.
The averaged values are: leaf moisture content, the
leaf body thickness and leaf body temperature as well.
t the screen you can see the averaged results and also
the current average number, at above screen it is "01".
The readiness duration between every measurement
is no longer than 13s. fter 16th measurement or after
the measurement duration readiness state greather
than 13s. the G2LM will switch off with increasing
about one the number of working memory bank,
for instance: 1 to 2,,, 29 to 30, and 30 again to 1.
If in course of the measurement duration period
the user will simply push and close the arm without
any leaf inserted in to the sensor aperture, the GL2M
will automatically restart new callibration procedure
preceding next measurement readiness. ll without change
of the working memory bank number. This option protects
again "waste" the memory bank in case of bad or improper
done measurement, for example after measurement realized
with not acurately inserted leafe.
If the user's concern is to take measurement without
averaging, is possible to realize it by closing the arm
after (as described) every separate measurement or
letting the GL2M to switch off automatically after 15s.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
5
3.Charging the battery:
The unit is powered with help of 3pcs. of type
rechargeable Ni-MH batteries. The state of capacity
is focused at the display, like at the above example
57%. Below 10% of capacity the unit will automatically
switch off, therefore to ensure working continuity
is good to keep charged batteries as is only possible.
To charge batteries, is provided the micro USB socket
available at the body of GL2M. The charging process will
begin after connection with any USB plug from the
working PC or external +5V/300m charger.
Caution: During the charging process the unit can't
measue anything except the battery state.
Caution: It can be used any +5V Battery Charger equiped
with micro USB plug, for example cell phone charger.
The maximum charging time is less than 3 hours, the
charging process is fully controlled through GL2M
microcontroller and the charging state is shown at
the display as next below:
E.
To carry out the charge is required to connect the
powered USB in to the GL2M USB socket. If the unit was
in measurement rediness duration state or totally
switched off the charging process will start accordingly,
the display will show as above.
When the batteries achieve the maximum charge, the
display will show 99% and the charging process will
start to work in standby mode.
Therefore is possible to charge the batteries without
paying attention on the process.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
6
4. Removing the batteries:
In the reason of batteries malfunction or potential
possible hardware electronics jamm - when the unit do
not respond on any task described earlier is essential
to remove and replace (if necesary new) bateries set.
To do it, unscrew 2 visible bolts, slide out the
plastic compartment, open it and remove batteries.
Caution: Never use other than rechargable Ni-MH 1.2V
battery types, othewise the unit electronics will damage.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
7
5.Data collect and parameter setup software:
The windows PC software let the user to modify
some of GL2M parameters, download, store and printout
the GL2M memory data banks and also the potential
firmware upgrade as well. fter running the PC software
- "GLM.EXE" the PC screen show the window as below:
G.
Now connect the PC USB plug with GL2M USB socket.
The connection procedure is the same as described by
battery charging proccess. Firstly, after connection
the GL2M display will show the battery charge screen.
Next, after pushing the "Connect" button, just after
come in to contact the GL2M display will show:
F.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
8
For the battery state safety the display shows also
the battery charge state, in course of PC data connection
the batteries are not charged.
Simultaneously the PC screen will (as example) show:
H.
To disconnect - push the button "Disconnect".
To save the shown data table in to the PC files
as "CSV" format - push the button "Save".
To printout on the attached printer or in the PDF
format the shown data table - push the button "Print"
To clear the data table and simultaneously clear all
GL2M data banks - push the button "Clear".
To switch between different GL2M working modes -
push the button "Edit", select the desired option
and push the button "Store"
The constants as "Kt", Ke", "Km" are changable only
in colaboration with SONIK service specialists.
The firmware upgrade is possible only if available
and exclusively in touch with SONIK.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
9
6.Thic ness sensor calibration:
In case of the thickness sensor mechanical deregulation
is neccesary to correct potential measurement errors.
To fix the problem let measure any thin nonconductive
layer with known thickness.
The value should be in range between 0.20mm and 0.50mm.
Next, with help of GLM software, change the
parameter described as "Kt" to obtain as a result the
exact measured value which corresponds the reference
thickness.
nother thickness calibration method is described in
point 1. connected with initial calibration procedures.
In this case you must-have the attached stainless steel
roundel with calibrated thickness – exact 1,000mm.

__________________________________________________________________________
GL2M Instruction Manual ASONIK 12.11.2017.
10
7.Technical Specification:
1. The meter designed for moisture content determination
in tea leafes.
2. Relative permittivity measured range: 0,9 to 99,9
3. Relative permittivity measurement resolution: 0,1
4. Relative permittivity measurement max error < 2% ±0.1
5. Leaf temperature measurement range: 0,1oC to 60,0oC
6. IR Leaf temperature measurement method, resolution 0,1oC
7. Leaf moisture content region of interest: 2,9% to 99,9%
8. Leaf moisture content measurement resolution: 0,1%
9. Leaf moisture content measurement error < 2% ±0.1%
10. Leaf thickness measurement range: 0,010mm up to 2,000mm
11. Leaf thickness measurement resolution: 0,001mm
12. Leaf thickness measurement error < 1% ±0.01mm
13. Press force on measured layer: 1,0N
14. Readout display: OLED 128 x 64 dots
15. Data interface: USB (micro USB socket)
16. Powering: 3 x Ni-MH 1.2V cells("ReadyToUse" or "ReCyko")
17. Battery charge – built in charging controll.
18. External charge power source:
+5V/300m with micro USB power plug source
19. Calibration duration: up to 5 sec.
20. Measurement duration: up to 2 sec.
21. Measurement mode with current averaging up to 16 cycles
22. Built in memory for 30 readouts
23. mbient air temperature and Rh measurement option
24. mbient air temperature range: -40oC to + 60oC
25. mbient temperature measurement accuracy tolerance: ±0,2oC
26. mbient air Rh range: 0 to 99.9%
27. mbient air Rh measurement accuracy: ±2%
28. One button handling – ASONIK created algorithm
29. Total weight: 500g
30. Dimensions: 45mm x 45mm x 125mm
31. Used materials: stainless steel, bronze, nylone, teflon
32. Food Grade approval
33. CE requirements
34. PC oftware available: GL2M v1.09
__________________________________________________________________
ASONIK
http://www.asoni .pl mailto:asoni @asoni .pl
POLAND, PL-62007 Tuczno,str. Słoneczna 6.
Tel./Fax +48 61 8156261, Mobile:+48 509993249