Temperature Compensation
In order to get accurate measuring results, we recommend that using a temperature probe for the calibration and measurements.
Automatic temperature compensation
Connect the temperature probe to the meter (Refer to page 4 “Connecting the Sensors”). The ATC icon immediately appears on the display, the meter is
now switched to the automatic temperature compensation mode.
Manual temperature compensation
If the meter does not detect a temperature probe, the oC icon will show on the display indicating that the meter is switched to the manual temperature
compensation mode. To set the temperature value of sample, follow the steps below.
1. Press and hold the oCkey for 3 seconds to enter the temperature setting mode.
2. Press the or key to modify the temperature value.
3. Press the Enter key to confirm.
During the setting process, press the or key once, the setting value will increase or decrease by 0.1. Press and hold the or key, the
setting value will increase or decrease by 1.